CRE™ PRO M00558680Recalls & reported problems

Boston Scientific Corporation · Esophageal/Pyloric/Colonic/Biliary Wireguided Balloon Dilatation Catheter · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Boston Scientific CRE PRO wireguided balloon dilatation catheter is used to widen narrowed areas of the esophagus, pylorus, colon, or bile ducts. The 55 reports mostly describe packaging that arrives torn or punctured, along with balloons that burst during inflation; no deaths and one injury were reported. One open recall, started in May 2026, covers a possible sterile breach of the packaging pouches.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
60
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 60 adverse event reports that list this exact device (DI 08714729797562), received between Jan 22, 2018 and Sep 29, 2026 — classified as 1 injury, 59 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (23 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 5 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 3 reportsAug 2026: 3 reportsSep 2026: 5 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 21, 2026Open, Classified

    Potential sterile breach of the pouches in which devices are packaged.

    Z-2628-2026 · Root cause: Package design/selection · FDA record ↗

Most reported problems

Device problems

  • Burst Container or Vessel23
  • Tear, Rip or Hole in Device Packaging16
  • Material Puncture/Hole10
  • Leak/Splash6
  • Adverse Event Without Identified Device or Use Problem5
  • Inflation Problem5
  • Use of Device Problem4
  • Packaging Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions41
  • No Consequences Or Impact To Patient14
  • No Patient Involvement3
  • No Known Impact Or Consequence To Patient2
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 60 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Inflation Problem; Material Puncture/Hole

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940192 ↗
  • MalfunctionSep 24, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50876560 ↗
  • MalfunctionSep 23, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50876368 ↗
  • MalfunctionSep 15, 2026Tear, Rip or Hole in Device Packaging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50779382 ↗
  • MalfunctionSep 15, 2026Tear, Rip or Hole in Device Packaging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50779378 ↗
  • MalfunctionAug 11, 2026Tear, Rip or Hole in Device Packaging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50214627 ↗
  • MalfunctionAug 11, 2026Tear, Rip or Hole in Device Packaging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50214624 ↗
  • MalfunctionAug 11, 2026Tear, Rip or Hole in Device Packaging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50214623 ↗
  • MalfunctionJul 16, 2026Tear, Rip or Hole in Device Packaging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25850269 ↗
  • MalfunctionJul 16, 2026Tear, Rip or Hole in Device Packaging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25850259 ↗
  • MalfunctionJul 16, 2026Tear, Rip or Hole in Device Packaging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25850254 ↗
  • MalfunctionMar 13, 2026Tear, Rip or Hole in Device Packaging; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS RECEIVED BY THE CUSTOMER ON AN UNKNOWN DATE. IT WAS REPORTED THAT UPON RECEIVING THE PACKAGING, THE BALLOON PACKAGING WAS PERFORATED/TORN IN THE AREAS WHERE THE SCRATCHES ARE OBSERVED. MEDIA INSPECTION OF THE PRODUCT PACKAGING PROVIDED BY THE COMPLAINANT REVEALED MULTIPLE SMALL PUNCTURES AND INDENTATIONS LOCATED ON THE STERILE POUCH. THE DAMAGE APPEARS AS SEVERAL CLUSTERED PIN SIZED PENETRATIONS AND SURFACE DENTS ON THE FRONT SIDE OF THE STERILE POUCH. THE COMPLAINANT ALSO REPORTED THAT THE OUTER SHIPPING BOX CONTAINING THE STERILE BALLOON POUCHES SHOWED NO SIGNS OF DAMAGE. THE DEVICE WAS NOT USED IN A PROCEDURE. NO PATIENT WAS INVOLVED AT THE TIME OF EVENT, AND NO PATIENT COMPLICATIONS WERE REPORTED IN CONNECTION WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24598977 ↗
  • MalfunctionMar 13, 2026Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED DURING A PROCEDURE TO TREAT STRICTURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, IT WAS REPORTED THAT THE DEVICE WAS INFLATED BUT BURST WITHIN SECONDS WHILE IN THE STRICTURE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24598947 ↗
  • MalfunctionMar 13, 2026Inflation Problem; Leak/Splash

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED DURING A DILATION PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THERE WAS A PRESSURE LOSS, AND THE BALLOON COULD NOT BE INFLATED TO 4.5 ATM. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THE ANATOMY LOCATION OF THE PROCEDURE IS UNKNOWN AND IS BEING REPORTED AS IT MAY HAVE OCCURRED IN THE ESOPHAGUS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO HAVE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24592431 ↗
  • MalfunctionMar 12, 2026Inflation Problem; Leak/Splash

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED DURING A DILATION PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THERE WAS A PRESSURE LOSS, AND THE BALLOON COULD NOT BE INFLATED TO 4.5 ATM. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THE ANATOMY LOCATION OF THE PROCEDURE IS UNKNOWN AND IS BEING REPORTED AS IT MAY HAVE OCCURRED IN THE ESOPHAGUS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO HAVE FULLY RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24587491 ↗
  • MalfunctionMar 12, 2026Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED DURING A PROCEDURE TO TREAT STRICTURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, IT WAS REPORTED THAT THE DEVICE WAS INFLATED BUT BURST WITHIN SECONDS WHILE IN THE STRICTURE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO REPORTED PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24584088 ↗
  • MalfunctionFeb 23, 2026Burst Container or Vessel; Difficult to Remove; Adverse Event Without Identified Device or Use Problem; Failure to Deflate

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN ESOPHAGEAL DILATION TO TREAT ESOPHAGEAL STENOSIS PERFORMED ON (B)(6) 2026. DURING THE PROCEDURE, THE DILATION BALLOON RUPTURED DURING ESOPHAGEAL DILATION. THEY WERE NOT ABLE TO EASILY DEFLATE AND PULL OUT THE BALLOON, SO THEY HAD TO CUT THE CATHETER OF THE BALLOON FOR IT TO BE LOOSE ENOUGH FOR REMOVAL. NO PIECE OF THE DEVICE WAS DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION WAS EXPECTED TO FULLY BE RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24423725 ↗
  • InjuryJan 13, 2026Inflation Problem; Detachment of Device or Device Component

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS USED DURING A DILATION PROCEDURE TO TREAT BENIGN STRICTURE PERFORMED ON AN (B)(6) 2025. DURING THE PROCEDURE, THE BALLOON DETACHED AND FAILED TO INFLATE. THE DEVICE WAS RETRIEVED USING GRASPING FORCEPS. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE FROM DIFFERENT BATCH. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24064115 ↗
  • MalfunctionJul 16, 2025Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE SIGMOID COLON DURING A COLONOSCOPY PROCEDURE TO TREAT POST-RESECTION OBSTRUCTION PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE BALLOON WAS DILATED TO 10 MM AND 11 MM AND WORK FINE. HOWEVER, WHEN DILATING TO 12MM, THE BALLOON BURST. IT WAS REPORTED THAT NO PIECE OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PRO WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22523769 ↗
  • MalfunctionJun 6, 2025Use of Device Problem; Material Integrity Problem; Use of Device Problem; Material Integrity Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE DUODENUM DURING A BILIARY DILATATION PROCEDURE PERFORMED TO TREAT DILATION ASSISTED STONE EXTRACTION (DASE) ON (B)(6) 2025. DURING PREPARATION, THE NURSE NOTICED THAT THE BALLOON WAS NOT MAINTAINING PRESSURE AND AFTER THEY CHECKED THEY FOUND THAT THE BALLOON HAS A DEFECT. ADDITIONAL CLARIFICATION TO DETERMINE THE NATURE OF THE DAMAGE WAS NOT PROVIDED AND THE BALLOON MAY HAVE BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PRO WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: THE INSTRUCTIONS FOR USE (IFU) INDICATE THAT THIS BALLOON SHOULD NOT BE PRE-INFLATED PRIOR TO USE IN THE PROCEDURE. HOWEVER, THE CUSTOMER REPORTED THAT THE BALLOON WAS PRE-INFLATED OUTSIDE OF THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22162021 ↗
  • MalfunctionApr 24, 2025Material Puncture/Hole; Material Puncture/Hole

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE ESOPHAGUS DURING AN ESOPHAGEAL DILATION PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, PINHOLE LEAK OCCURRED DURING DILATION BETWEEN 10MM TO 12MM. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PRO WIREGUIDED DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21902280 ↗
  • MalfunctionMar 21, 2025Leak/Splash; Use of Device Problem; Leak/Splash; Use of Device Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) WITH DILATATION PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE CRE BALLOON LEAK. IT WAS REPORTED THAT NO PIECE OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE PRO WIREGUIDED DILATATION BALLOON. THE ANATOMY LOCATION WAS REPORTED TO BE THE ESOPHAGOGASTRODUODENOSCOPY (EGD) AND IS BEING REPORTED AS IT MAY HAVE OCCURRED IN THE ESOPHAGUS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: THE INSTRUCTIONS FOR USE (IFU) INDICATE THAT THIS BALLOON SHOULD NOT BE PRE-INFLATED PRIOR TO USE IN THE PROCEDURE. HOWEVER, THE CUSTOMER REPORTED THAT THE BALLOON WAS TESTED OUTSIDE THE PATIENT DURING PREPARATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21665476 ↗
  • MalfunctionMay 28, 2024Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem; Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE PAPILLA DURING AN ENDOSCOPIC SPHINCTEROTOMY (EST) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE BALLOON BURST AT 4 ATM. THE CUSTOMER STATED THAT NO PIECES OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER DIFFERENT MODEL DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19410688 ↗
  • MalfunctionMay 16, 2024Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE SPHINCTER OF ODDI DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE BALLOON WAS ABLE TO BE INFLATED AT 10 MM BUT WHEN THE BALLOON WAS ATTEMPTED TO BE INFLATED TO 11 MM THE BALLOON BURST ON THE SHOULDER PROXIMAL TO THE ENDOSCOPE BEFORE GETTING TO THE 5 ATM OF PRESSURE. THE CUSTOMER STATED THAT NO PIECES OF THE BALLOON DETACHED INSIDE THE PATIENT. THE PROCEDURE WAS ABLE TO BE COMPLETED WITH THE ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19336675 ↗
  • MalfunctionApr 12, 2024Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem; Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE PRO WIREGUIDED DILATATION BALLOON WAS ATTEMPTED TO BE USED DURING A PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE BALLOON BURST WHEN IT REACHED MAXIMUM DILATATION. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19101293 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →