Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Charger (49)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729803805), received between May 13, 2019 and Jan 29, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Inflation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Inflation Problem1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 29, 2021Inflation Problem; Inflation Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 21-JAN-2021. IT WAS REPORTED THAT INFLATION FAILURE WERE ENCOUNTERED. THE TARGET LESION WAS LOCATED IN THE RIGHT ILIAC ARTERY AND SUPERFICIAL FEMORAL ARTERY. THE LESION WAS 90% STENOSED WITH MILD CALCIFICATION. A 10.0 X 60, 75CM CHARGER BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, THE BALLOON WOULD NOT HOLD ITS PRESSURE AND WAS NOT RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A BALLOON PINHOLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11251649 ↗
  • MalfunctionMay 13, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT ILIAC. AFTER STENT PLACEMENT POST DILATION WAS PERFORMED WITH A CHARGER BALLOON CATHETER (12X40, 12X60 AND 10X60). EACH CHARGER BALLOON CATHETER RUPTURED. A FOURTH CHARGER (12X60) WAS USED SUCCESSFULLY DILATED AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS FINE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8605254 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →