Charger™ H74939206120670Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Charger (49)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729803874), received between May 13, 2019 and Apr 6, 2021 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Rupture3
- Leak/Splash2
- Deflation Problem1
- Detachment of Device or Device Component1
- Difficult to Remove1
Patient problems
- No Consequences Or Impact To Patient4
- Device Embedded In Tissue or Plaque1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 6, 2021Leak/Splash; Leak/Splash
REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 19-MAR-2021. IT WAS REPORTED THAT BALLOON LEAK OCCURRED. THE MODERATELY STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED LOWER EXTREMITY VEIN. A 12.0 X60, 75CM CHARGER BALLOON CATHETER WAS SELECTED FOR USE BUT THE BOTTOM OF THE BALLOON WAS LEAKING SALINE CONTRAST WHEN FLUSHED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED BALLOON PINHOLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11622083 ↗InjuryJan 12, 2021Leak/Splash; Difficult to Remove; Detachment of Device or Device Component; Leak/Splash; Difficult to Remove; Detachment of Device or Device Component
IT WAS REPORTED THAT THERE WAS A LEAK, DIFFICULTY REMOVING THE DEVICE, AND UNRETRIEVED DEVICE FRAGMENTS LEFT IN THE PATIENT. A 12MM X 60MM X 75CM CHARGER BALLOON CATHETER WAS SELECTED FOR USE IN A LEFT UPPER ARM BALLOON ANGIOPLASTY AND STENTING PROCEDURE. THE LEFT UPPER SUBCLAVIAN VEIN HAD TWO PREVIOUSLY PLACED STENTS. ONE STENT APPEARED TO HAVE BEEN PLACED THROUGH THE STRUTS OF THE OTHER. SOME FRACTURES WERE PRESENT WITH EXPOSED NITINOL STRUTS. UPON INFLATION AND DEFLATION IN THE AREA OF THE FRACTURED STENT, BLOOD WAS NOTED IN THE INFLATION DEVICE, WHICH INDICATED A PIN HOLE OR TEAR IN THE BALLOON MATERIAL. RESISTANCE WAS FELT UPON REMOVAL. SOME BALLOON MATERIAL REMAINED ON THE STENT STRUT. A SECOND WIRE WAS PLACED NEXT TO THE FIRST AND USED TO COMPLETE THE CASE. A 12MM X 60MM VENOUS WALLSTENT WAS PLACED OVER THE SECOND WIRE AND POST DILATED WITH A 10MM X 40MM BALLOON, WHICH OPENED THE IN-STENT RESTENOSIS AND TRAPPED ANY BALLOON MATERIAL BEHIND THE STENT. ANGIOGRAPHIC AND INTRAVASCULAR ULTRASOUND (IVUS) IMAGING PROVED TO BE WIDELY PATENT. BOTH THE FIRST AND SECOND WIRES WERE REMOVED WITHOUT INCIDENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 11155458 ↗MalfunctionJul 24, 2020Deflation Problem; Deflation Problem
IT WAS REPORTED THAT BALLOON DEFLATION FAILURE OCCURRED. A 12.0 X60, 75CM CHARGER BALLOON CATHETER WAS SELECTED FOR USE. HOWEVER, IT WAS NOTED THAT THE BALLOON WOULD NOT DEFLATE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10319860 ↗MalfunctionOct 3, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 12.0 X60, 75CM CHARGER BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, WHEN THE BALLOON WAS INFLATED AT 3 ATMOSPHERES, THE BALLOON RUPTURED. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9151510 ↗MalfunctionJun 9, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED ILIAC VEIN. A 12.0 X60, 75CM CHARGER BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING FIRST INFLATION AT 10 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8681348 ↗MalfunctionMay 13, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT ILIAC. AFTER STENT PLACEMENT POST DILATION WAS PERFORMED WITH A CHARGER BALLOON CATHETER (12X40, 12X60 AND 10X60). EACH CHARGER BALLOON CATHETER RUPTURED. A FOURTH CHARGER (12X60) WAS USED SUCCESSFULLY DILATED AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8605248 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.