Charger™ H74939206041510Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Charger (49)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729804048), received between Aug 4, 2020 and Aug 10, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
- Difficult to Advance1
- Difficult to Remove1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryAug 10, 2021Break; Difficult to Remove; Difficult to Advance; Break; Difficult to Remove; Difficult to Advance
IT WAS REPORTED THAT REMOVAL DIFFICULTIES WERE ENCOUNTERED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY (SFA). PHYSICIAN CROSSED RETROGRADE FROM THE POSTERIOR TIBIAL ARTERY AND GAINED ACCESS FROM THE CONTRALATERAL GROIN. THE PHYSICIAN WAS ABLE TO FLOSS THE GUIDEWIRE THROUGH THE 6F SHEATH OUT THE GROIN ACCESS. A 4.0 X150, 135CM CHARGER BALLOON CATHETER WAS ADVANCED; HOWEVER, IT WAS NOTICED THAT ADVANCING WAS EXTREMELY DIFFICULT. THE PHYSICIAN PERFORMED MULTIPLE INFLATIONS UP AND DOWN THE LENGTH OF SFA. A LOT OF RESISTANCE WAS FELT WHEN THE PHYSICIAN TRIED TO REMOVE THE BALLOON. THE BALLOON STOPPED MOVING BACK AND THE CATHETER STRETCHED AND SNAPPED APPROXIMATELY 5 CM FROM THE PROXIMAL HUB. THE PHYSICIAN USED A 6F NON-BSC GUIDE CATHETER AND ADVANCED IT OVER THE BALLOON CATHETER TO TRY TO CAPTURE THE BALLOON IN THE GUIDE CATHETER. HE WAS ABLE TO ADVANCE THE GUIDE CATHETER TO THE PROXIMAL BALLOON RO MARKER, BUT UNABLE TO ADVANCE THE GUIDE CATHETER OVER THE BALLOON. HE MANIPULATED THE GUIDE CATHETER AND THE CHARGER BALLOON CATHETER TOGETHER PULLING IN SHORT STOKES TO REMOVE THE PORTION OF THE REMAINING CHARGER BALLOON
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12304691 ↗MalfunctionAug 4, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. A 2.00MM PERIPHERAL ROTALINK PLUS BURR WAS SELECTED FOR USE BUT FAILED TO REACH OPTIMUM SPEED DURING INITIAL TESTING OUTSIDE THE PATIENT'S BODY. ANOTHER PERIPHERAL ROTALINK PLUS BURR WAS USED AND WORKED JUST FINE. A 4.0 X150, 135CM CHARGER BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT NOMINAL PRESSURE, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED FROM THE PATIENT'S BODY AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10363181 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page