Skip to content

Charger™ H74939206051510Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Charger (49)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729804185), received between Sep 10, 2018 and Sep 10, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Rupture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Rupture1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2018Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND HEAVILY CALCIFIED LEFT SUPERFICIAL FEMORAL ARTERY. A 5.0 X150, 135CM CHARGER BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION TO A NOMINAL PRESSURE, THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED FROM THE PATIENT'S BODY. A 5.0MMX80MMX135CM (4F) STERLING BALLOON CATHETER WAS THEN ADVANCED FOR DILATATION. HOWEVER, UPON INFLATION TO A NOMINAL PRESSURE, THE BALLOON ALSO RUPTURED. THE DEVICE WAS COMPLETELY REMOVED FROM THE PATIENT'S BODY AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS FINE. THE DEVICE WAS RETURNED FOR ANALYSIS. A VISUAL EXAMINATION IDENTIFIED THAT THE BALLOON WAS NOT FOLDED WHICH INDICATES THAT THE BALLOON WAS SUBJECTED TO POSITIVE PRESSURE. BLOOD WAS IDENTIFIED WITHIN THE BALLOON AND INFLATION LUMEN WHICH IS EVIDENCE OF A DEVICE LEAK. A VISUAL AND MICROSCOPIC EXAMINATION IDENTIFIED A BALLOON LONGITUDINAL TEAR BEGINNING AT THE DISTAL END OF THE DISTAL MARKERBAND AND EXTENDING APPROXIMATELY 68MM PROXIMALLY ACROSS THE BALLOON MATERIAL. AN EXAMINATION

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7860010 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page