Skip to content

Charger™ H74939206081010Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Charger (49)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729804574), received between Sep 25, 2019 and Jul 27, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Break1
  • Difficult to Remove1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 27, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. AN 8.0 X100, 135CM CHARGER BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION AT 6 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER 8X100 CHARGER BALLOON CATHETER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS STABLE. IT WAS FURTHER REPORTED THAT THE BALLOON RUPTURED AT 10 ATMOSPHERES AND THE DEVICE WAS COMPLETELY REMOVED THROUGH SHEATH USING STANDARD EXTRACTING PROCESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12232736 ↗
  • MalfunctionSep 25, 2019Break; Difficult to Remove; Break; Difficult to Remove

    IT WAS REPORTED THAT A SHAFT BREAK OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT FEMORAL VEIN. A 8.0 X100, 135CM CHARGER BALLOON CATHETER WAS ADVANCED FOR DILATATION. AFTER THE LAST INFLATION, THE BALLOON WAS TOTALLY DEFLATED; HOWEVER, WHEN THE DEVICE WAS PULLED OUT, THE BALLOON CATHETER BECAME STUCK ABOUT HALF WAY THROUGH THE INTRODUCER SHEATH. THE CATHETER BROKE IN TWO AT THE CONNECTION BETWEEN THE BALLOON AND THE CATHETER. THE BROKEN PARTS WERE RETRIEVED AS IT CAME OUT WITH THE SHEATH. VENOGRAM WAS TAKEN AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9119523 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page