Skip to content

Sterling™ H74939032501210Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Sterling (153)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729845348), received between Aug 26, 2020 and Nov 13, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Markings/Labelling Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Device Markings/Labelling Problem1
  • Failure to Advance1
  • Inflation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 13, 2025Failure to Advance; Adverse Event Without Identified Device or Use Problem; Failure to Advance; Adverse Event Without Identified Device or Use Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 22OCT2025. IT WAS REPORTED THAT THE BALLOON WAS UNABLE TO CROSS THE LESION OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED VESSEL. A 5.0X120X150 STERLING CATHETER WAS SELECTED FOR RENAL ARTERY PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURES. THE DEVICE DID NOT ADVANCE WELL WITHIN THE BLOOD VESSELS. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATION AS THE RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE BALLOON HAD A RUPTURE AT 7CM FROM THE TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23542543 ↗
  • MalfunctionAug 12, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT LABELING ISSUE OCCURRED. A 6.0MMX80MMX135CM-HYBRID STERLING WAS SELECTED FOR USE. HOWEVER, DURING PREPARATION, IT WAS NOTED THAT THE PRODUCT INNER POUCH DID NOT MATCH THE OUTER SHELF CARTON. THE PROCEDURE WAS COMPLETED USING AN ALTERNATE DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19966685 ↗
  • MalfunctionAug 26, 2020Inflation Problem; Inflation Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 19AUG2020. IT WAS REPORTED THAT BALLOON INFLATION FAILURE OCCURRED. A 5.0X120X150 STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, IT WAS NOTED THAT THE BALLOON COULD NOT BE INFLATED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED AND THE PATIENT STATUS WAS STABLE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A BALLOON PINHOLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10455156 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page