Sterling™ H74939032601210Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Sterling (153)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 08714729845423), received between Nov 12, 2019 and Jan 14, 2025 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture8
  • Adverse Event Without Identified Device or Use Problem1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient3
  • Vascular Dissection1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 14, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TOTALLY STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0X120X150 STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING INFLATION, THE BALLOON BURST BEFORE REACHING THE NORMAL BURST PRESSURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS STABLE POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21147470 ↗
  • MalfunctionAug 12, 2024Leak/Splash; Material Rupture; Leak/Splash; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0X120X150 STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON BREAK BLOOD FLOWS AND WOULD NOT HOLD BLOOD BACK INTO INFLATION DEVICE AT 8 ATMOSPHERES. THE PROCEDURE WAS COMPLETED WITH A NON-BSC DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19967277 ↗
  • MalfunctionOct 19, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 87% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 6.0X120X150-HYBRID STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 8 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17965305 ↗
  • MalfunctionOct 3, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED POPLITEAL ARTERY. A 6.0X120X150-HYBRID STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 8 ATMOSPHERES FOR A FEW SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS SUCCESSFULLY REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17865294 ↗
  • MalfunctionSep 10, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. A 6.0X120X150 STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED BEFORE REACHING NORMAL ATMOSPHERIC PRESSURE. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT STATUS WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12457509 ↗
  • InjuryFeb 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A DISSECTION OCCURRED. A 6MM X 120MM X 150CM STERLING BALLOON WAS SELECTED FOR USE IN AN ANGIOPLASTY AND STENTING PROCEDURE. THE PATIENT PRESENTED WITH STENOSIS IN THE SUPERFICIAL FEMORAL ARTERY (SFA). DURING THE PROCEDURE, A DISSECTION OCCURRED. THE DISSECTION MOST LIKELY OCCURRED POST ANGIOPLASTY. ONE ELUVIA STENT WAS IMPLANTED TO TREAT THE DISSECTION. TWO ADDITIONAL ELUVIA STENTS WERE IMPLANTED IN DIFFERENT AREAS OF THE SFA TO TREAT THE STENOSIS. THE PROCEDURE WAS COMPLETED, AND THE PATIENT WAS EXPECTED TO FULLY RECOVER.

    Patient problems: Vascular Dissection

    MAUDE report 11358818 ↗
  • MalfunctionFeb 27, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 6.0X120X150-HYBRID STERLING BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION AT 6 ATMOSPHERES IN THE SEVERELY CALCIFIED LESION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9763650 ↗
  • MalfunctionNov 12, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. TWO 6.0X120X150-HYBRID STERLING BALLOON CATHETERS WERE ADVANCED FOR DILATATION. HOWEVER, BOTH BALLOONS RUPTURED UPON INITIAL INFLATION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9310651 ↗
  • MalfunctionNov 12, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY. TWO 6.0X120X150-HYBRID STERLING BALLOON CATHETERS WERE ADVANCED FOR DILATATION. HOWEVER, BOTH BALLOONS RUPTURED UPON INITIAL INFLATION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS NOR INJURIES WERE REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9310648 ↗

Other Catheter, angioplasty, peripheral, transluminal devices