Advanix™ Pancreatic Stent M00537170Recalls & reported problems
Boston Scientific Corporation · Pancreatic Stent · Class 2 · Implant
All models of Advanix Pancreatic Stent (48)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729856641), received between Dec 27, 2017 and Sep 17, 2020 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Kinked1
- Material Deformation1
Patient problems
- No Consequences Or Impact To Patient2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2020Material Deformation; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING A PANCREATIC STENT PLACEMENT PROCEDURE IN THE PANCREATIC DUCT PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, AN ADVANIX PANCREATIC STENT WAS ADVANCED OVER A GUIDEWIRE INTO THE PATIENT, THE PHYSICIAN FELT RESISTANCE WHILE ATTEMPTING TO INSERT THE STENT. WHEN THE ENDOSCOPE SCREEN WAS CHECKED, THE STENT WAS FOUND TO BE KINKED. REPORTEDLY, THE STENT WAS REMOVED FROM THE PATIENT. A NEW ADVANIX PANCREATIC STENT WAS OPENED AND SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10547157 ↗InjurySep 8, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A 5 FR ADVANIX PANCREATIC STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE, PERFORMED ON (B)(6) 2020. REPORTEDLY, THE PATIENT HAD A PANCREATIC FISTULA AND THE ANATOMY WAS VERY CHALLENGING DUE TO THE PANCREATIC DUCTAL ANATOMY, MAKING THE PROCEDURE DIFFICULT AND COMPLEX. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE 5 FR ADVANIX PANCREATIC STENT FAILED TO DEPLOY. THE 5 FR STENT WAS REMOVED FROM THE DUCT AND A 4 FR ADVANIX PANCREATIC STENT WAS INSERTED. DURING THE PLACEMENT OF THE 4 FR PANCREATIC STENT, FLUOROSCOPY REVEALED A SMALL PIECE OF MATERIAL WHICH WAS ASSUMED TO BE A MAGNETIC RESONANCE MARKER FROM THE 5 FR STENT. THE PIECE SEEN ON FLUOROSCOPY WAS THE SIZE OF A MARKER, BUT WAS UNABLE TO BE RETRIEVED. AN ANGLE-TIP GUIDEWIRE WAS USED TO SUCCESSFULLY DEPLOY THE 4 FR ADVANIX PANCREATIC STENT TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PHYSICIAN EXAMINED THE 5 FR ADVANIX STENT AND NO PART APPEARED TO BE MISSING.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10501211 ↗MalfunctionDec 27, 2017Kinked; Kinked
NOTE: THIS REPORT PERTAINS TO ONE OF THREE DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT # 3005099803-2017-03865, 3005099803-2017-03869, 3005099803-2017-03866 FOR THE ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY PROCEDURE PERFORMED ON (B)(6) 2017. AS THE PHYSICIAN INTRODUCED A 7FRX 11CM ADVANIX PANCREATIC STENT INTO THE PATIENT¿S PANCREATIC DUCT, HE EXPERIENCED SOME RESISTANCE. THE STENT WAS ADVANCED INTO POSITION AND WHEN THE TECHNICIAN ATTEMPTED TO DEPLOY THE STENT THE DELIVERY SYSTEM AND ADVANIX PANCREATIC STENT BOTH BUCKLED/CRUNCHED UP WHICH WOULD NOT ALLOW THE STENT TO RELEASE FROM THE CATHETER. THE DEVICE WAS REMOVED FROM THE SCOPE. THEY THEN ATTEMPTED TO PLACE A 5FRX9CM ADVANIX PANCREATIC STENT OVER A GUIDEWIRE AND PUSH IT INTO POSITION WITH A HYDRATOME. DURING THE ADVANCING OF THE 5FRX9CM ADVANIX PANCREATIC STENT, IT TOO BUCKLED AND HAD TO BE REMOVED. FINALLY, A NAVIFLEX PANCREATIC DELIVERY SYSTEM WAS LOADED WITH THE 7FRX9CM ADVANIX PANCREATIC STENT AND ATTEMPTED TO BE PUSHED INTO POSITION. THE STENT WAS IN POSITION AN
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7148572 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page