Advanix™ Pancreatic Stent M00537260Recalls & reported problems

Boston Scientific Corporation · Pancreatic Stent · Class 2 · Implant

All models of Advanix Pancreatic Stent (48)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 08714729856856), received between Oct 17, 2019 and Sep 4, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Deformation6
  • Adverse Event Without Identified Device or Use Problem3
  • Use of Device Problem3
  • Positioning Failure2
  • Break1
  • Component Missing1
  • Difficult to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 4, 2026Use of Device Problem; Material Deformation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50600410 ↗
  • MalfunctionNov 24, 2025Positioning Failure; Use of Device Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem; Positioning Failure; Use of Device Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS TO BE IMPLANTED TO TREAT A MALIGNANT PANCREATIC DUCT STRICTURE IN THE PANCREATIC DUCT HEAD REGION DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE STENT FAILED TO DEPLOY. ANOTHER ADVANIX PANCREATIC STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. NOTE: IT WAS REPORTED THAT THE BARB FLAP COVER WAS NOT USED TO INSERT THE DEVICE THROUGH THE BIOPSY CAP. PER THE IFU, "SLIDE THE OUTER ORANGE SLEEVE OVER THE MOST PROXIMAL STENT BARB TO HOLD IT DOWN FOR STENT INSERTION." THE USER DID NOT FOLLOW THE INSTRUCTIONS FOR USE (IFU). NOTE: THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF STENT KINKED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23638985 ↗
  • MalfunctionAug 27, 2025Use of Device Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem; Use of Device Problem; Material Deformation; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A ADVANIX PANCREATIC STENT WAS TO BE IMPLANTED TO TREAT A STRICTURE IN THE PANCREATIC DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, AN ATTEMPT WAS MADE TO ADVANCE THE STENT OVER THE GUIDEWIRE. UPON REACHING THE AMPULLA, IT WAS OBSERVED THAT THE DISTAL PORTION OF THE STENT, SPECIFICALLY BENEATH THE LEADING BARB, APPEARED KINKED. ANOTHER ADVANIX PANCREATIC STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: IT WAS REPORTED THAT THE GUIDEWIRE WAS IN THE PANCREATIC DUCT, DURING THE ATTEMPTED DEPLOYMENT. PER THE INSTRUCTION FOR USE, "IF THE GUIDEWIRE IS NOT COMPLETELY RETRACTED INTO THE DELIVERY SYSTEM, THE STENT CANNOT BE FULLY DEPLOYED.". ADDITIONAL INFORMATION RECEIVED ON AUGUST 15, 2025: IT WAS REPORTED THAT THERE WAS A 10-MINUTE DELAY IN THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22907075 ↗
  • MalfunctionJul 9, 2025Break; Difficult to Advance; Break; Difficult to Advance

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT PIGTAIL LEADING BARB 7F 7CM WAS TO BE IMPLANTED IN THE PANCREATIC DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE PIGTAIL PART OF THE STENT WAS BROKEN, PREVENTING THE ORANGE PUSHER PART FROM ADVANCING DESPITE BEING PUSHED. A DIFFERENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED ON JULY 17, 2025. THE REPORTED COMPLAINT OCCURRED OUTSIDE THE BODY, PRIOR TO DEVICE INSERTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22465000 ↗
  • MalfunctionSep 5, 2024Component Missing; Component Missing

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS TO BE IMPLANTED IN THE PANCREATIC DUCT TO TREAT A PANCREATIC STONE DURING A PANCREATIC STONE REMOVAL PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE STENT WAS SUCCESSFULLY DEPLOYED. HOWEVER, THE ENDOSCOPIC MARKER ON THE DUODENAL SIDE OF THE STENT WAS DIFFICULT TO VIEW. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20152829 ↗
  • MalfunctionJun 23, 2023Material Deformation; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC THAT AN ADVANIX PANCREATIC STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE IN THE PANCREAS, PERFORMED ON (B)(6) 2023. DURING PREPARATION, IT WAS FOUND THAT THE DISTAL PART OF THE STENT WAS KINKED. ANOTHER ADVANIX PANCREATIC STENT WAS OPENED AND USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17190143 ↗
  • MalfunctionJul 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    NOTE: THIS REPORT IS ONE OF TWO COMPLAINTS THAT PERTAIN TO THE SAME EVENT. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A NAVIFLEX RX DELIVERY SYSTEM AND AN ADVANIX PANCREATIC STENT WERE USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY IN THE PANCREATIC DUCT PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, THE PHYSICIAN ATTEMPTED TO DEPLOY A PANCREATIC PIGTAIL TO ENSURE DRAINAGE. WHEN THE NAVIFLEX RX DELIVERY SYSTEM WAS ADVANCED TO THE STENT, THE GUIDE CATHETER BROKE INTO TWO. THE BROKEN PIECE OF GUIDE CATHETER WAS RETRIEVED WITH A SNARE. ALL BROKEN PIECE OF GUIDE CATHETER WAS REMOVED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ADVANIX-NAVIFLEX PANCREATIC STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. INVESTIGATION RESULTS REVEALED THE STENT WAS DAMAGED, THEREFORE THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT FOR THE ADVANIX PANCREATIC STENT. PLEASE SEE FOR FULL INVESTIGATION DETAILS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14911388 ↗
  • MalfunctionJun 29, 2021Positioning Failure; Material Deformation; Positioning Failure; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING A STENT PLACEMENT PROCEDURE IN THE MAIN PANCREATIC DUCT PERFORMED ON (B)(6) 2021. DURING PROCEDURE, IT WAS NOTED THAT THEY HAD DIFFICULTY RETRACTING THE NAVIFLEX DELIVERY SYSTEM. MORE FORCE WAS APPLIED BUT THE STENT BECAME KINKED AND COULD NOT BE RELEASED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ADVANIX PANCREATIC STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE AS "NO PATIENT INJURY".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12086142 ↗
  • MalfunctionOct 17, 2019Material Deformation; Material Deformation

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING A PROCEDURE, PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THEY ATTEMPTED TO LOAD THE STENT ONTO THE NAVIFLEX7FR, IT WAS DIFFICULT FOR THEM TO SET UP OR PREPARE. WHEN REPLACED WITH ANOTHER OF THE SAME DELIVERY SYSTEM, IT MOUNTED SMOOTHLY AND THE PROCEDURE WAS COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS; PANCREATIC STENT KINKED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9202337 ↗