Advanix™ Pancreatic Stent M00537280Recalls & reported problems
Boston Scientific Corporation · Pancreatic Stent · Class 2 · Implant
All models of Advanix Pancreatic Stent (48)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729856870), received between May 20, 2019 and Jan 21, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation4
- Break2
- Adverse Event Without Identified Device or Use Problem1
- Positioning Failure1
Patient problems
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 21, 2026Positioning Failure; Material Deformation; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS TO BE IMPLANTED IN THE MAIN PANCREATIC DUCT TO TREAT A PANCREATIC CANCER DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THAT DEVICE RELEASE WAS ATTEMPTED AT THE INTENDED SITE, BUT THE INNER TUBE COULD NOT BE WITHDRAWN. ANOTHER ADVANIX PANCREATIC STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS WHICH REVEALED THAT THE STENT WAS KINKED AND DAMAGED. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24130754 ↗MalfunctionDec 5, 2019Break; Material Deformation; Break; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY WITH PANCREATIC STENT IMPLANTATION PERFORMED IN THE PANCREATIC DUCT ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE STENT KINKED ON THE PUSH CATHETER, ALTHOUGH IT HAD BEEN MOISTENED. THE STENT WAS PIERCED AT THE SITE OF THE BEND, MAKING IT UNUSABLE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ADVANIX BILIARY STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9419040 ↗MalfunctionOct 14, 2019Material Deformation; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE IN THE PANCREATIC DUCT, PERFORMED ON (B)(6), 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY THE STENT, IT WAS NOTICED THAT THE STENT BUCKLED ABOUT HALF WAY UP. THE STENT WAS REMOVED OUT FROM THE PATIENT AND ANOTHER ADVANIX PANCREATIC STENT SUCCESSFULLY COMPLETED THE PROCEDURE. THERE WAS NO SERIOUS INJURY, NOR WERE THERE ANY ADVERSE PATIENT EFFECTS REPORTED AS A RESULT OF THIS EVENT. THE CONDITION OF THE PATIENT AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9187076 ↗MalfunctionOct 8, 2019Material Deformation; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) IN THE PANCREAS PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, DIFFICULTY REMOVING THE DELIVERY SYSTEM WAS ENCOUNTERED DUE TO THE ADHESION BETWEEN THE STENT AND THE CARRIER. THE NURSE HAD TO APPLY FORCE TO REMOVE THE CARRIER, RESULTING THE STENT AND THE CARRIER TO BE KINKED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ADVANIX PANCREATIC STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS GOOD.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9167119 ↗InjuryMay 20, 2019Break; Break
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE IN THE PANCREAS, PERFORMED ON (B)(6) 2019. ACCORDING TO THE COMPLAINANT, DURING THE STENT PLACEMENT PROCEDURE, THE DELIVERY SYSTEM WAS UNABLE TO DEPLOY THE STENT. AFTER TRYING TO REMOVE THE DELIVERY SYSTEM FROM THE PATIENT, THE PUSH CATHETER TWISTED SO THAT THE STENT WAS UNABLE TO BE DISCONNECTED. AFTER THAT, THE STENT WAS BROKEN INTO TWO PIECES. THE BROKEN PIECES OF STENT WERE RETRIEVED WITH THE USE OF A SNARE AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ADVANIX PANCREATIC STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8625478 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗