Advanix™ Pancreatic Stent M00537330Recalls & reported problems
Boston Scientific Corporation · Pancreatic Stent · Class 2 · Implant
All models of Advanix Pancreatic Stent (48)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729856924), received between Sep 12, 2017 and May 7, 2026 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation4
- Defective Device1
- Kinked1
- Migration1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient3
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 7, 2026Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A ADVANIX PANCREATIC STENT WAS TO BE IMPLANTED INTO THE BILE DUCT TO TREAT AN OBSTRUCTION CAUSED BY STONES DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PERFORMED ON (B)(6) 2026. PRIOR TO PROCEDURE AND UPON OPENING THE DEVICE PACKAGE, IT WAS OBSERVED THAT THERE WAS A TEAR IN THE STENT. THE PROCEDURE WAS COMPLETED USING ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25108650 ↗InjuryJun 20, 2024Material Deformation; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED IN THE PANCREATIC DUCT FOR DRAINAGE DURING AN ENDOSCOPIC RETROGRADE PANCREATIC DRAINAGE (ERPD) PROCEDURE PERFORMED ON (B)(6) 2024. DURING THE PROCEDURE, THE STENT WAS DEPLOYED; HOWEVER, IT WAS NOTED THAT THE DISTAL TIP WAS KINKED. THE STENT WAS REMOVED, AND ANOTHER ADVANIX PANCREATIC STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19578280 ↗MalfunctionMar 23, 2022Positioning Failure; Defective Device; Positioning Failure; Defective Device
NOTE: THIS REPORT IS ONE OF TWO COMPLAINTS THAT PERTAIN TO THE SAME EVENT (MFR. REPORT # 3005099803-2022-00628). IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING A PANCREATIC DUCT STENT PLACEMENT PROCEDURE IN THE PANCREAS, PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, WHEN THE PHYSICIAN ATTEMPTED TO DEPLOY THE STENT, THE STENT FAILED TO DEPLOY AND COULD NOT BE REMOVED FROM THE TORN GUIDE CATHETER. ANOTHER ADVANIX PANCREATIC STENT WAS OPENED AND SUCCESSFULLY COMPETED THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF STENT DAMAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13869290 ↗MalfunctionJun 11, 2020Migration; Migration
IT WAS REPORTED THAT AN ADVANIX PANCREATIC STENT WAS EXPLANTED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PERFORMED IN THE COMMON BILE DUCT/HEPATIC DUCT PERFORMED ON (B)(6) 2020. ON (B)(6) 2020, THE ADVANIX PANCREATIC STENT WAS SUCCESSFULLY IMPLANTED WITH NO ISSUES REPORTED. ON (B)(6) 2020 DURING A PLANNED STENT REMOVAL PROCEDURE, IT WAS NOTED THAT THE ADVANIX PANCREATIC STENT HAD MIGRATED PROXIMALLY INTO THE LIVER DUCTS. AFTER REMOVING THE MIGRATED STENT, A SMALL AMOUNT OF BLOODY BILE WAS NOTED UPON REMOVAL. NO TREATMENT WAS REQUIRED FOR THE BLEED AND THE BLEEDING WAS SELF LIMITING. A NEW ADVANIX PANCREATIC PIGTAIL STENT WAS PLACED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 10144009 ↗MalfunctionApr 16, 2020Material Deformation; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE BILIARY DRAINAGE PROCEDURE IN THE COMMON BILE DUCT PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE STENT COULD NOT PASS THROUGH THE STRICTURE SITE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE. INVESTIGATION RESULTS REVEALED THAT THE PANCREATIC STENT WAS KINKED THEREFORE THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9969492 ↗MalfunctionMar 30, 2020Material Deformation; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORTAION THAT AN ADVANIX PANCREATIC STENT WAS USED IN A PROCEDURE IN THE PANCREATIC DUCT, PERFORMED ON (B)(6), 2020. ACCORDING TO THE COMPLAINT, DURING THE PANCREATIC DUCT STENT PLACEMENT PROCEDURE, THE PIGTAIL OF THE STENT WAS KINKED. REPROTEDLY, THERE WAS NO HEALTH DAMAGE TO THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH ANOTHER ADVANIX PANCREATIC STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9897508 ↗MalfunctionSep 12, 2017Kinked; Kinked
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING PANCREATIC DUCT STENT PLACEMENT PROCEDURE PERFORMED IN THE PANCREATIC DUCT (EXACT DATE NOT REPORTED). ACCORDING TO THE COMPLAINANT, WHEN THEY ATTEMPTED TO DEPLOY THE STENT, GUIDEWIRE WAS PLACED IN THE PANCREATIC DUCT. HOWEVER, RESISTANCE WAS FELT FROM THE MIDPOINT AND IT WAS UNABLE TO ADVANCE ANY FURTHER SO THE DEVICE WAS RETRIEVED. WHEN THEY RETRIEVED THE ADVANIX PANCREATIC STENT, THE STENT TIP ON THE DISTAL SIDE WAS FOUND TO BE ¿DAMAGED¿. THE PROCEDURE WAS COMPLETED BY PLACING ANOTHER ADVANIX PANCREATIC STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION WAS REPORTED TO BE "GOOD".
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6858805 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page