Advanix™ Pancreatic Stent M00536560Recalls & reported problems
Boston Scientific Corporation · Pancreatic Stent · Class 2 · Implant
All models of Advanix Pancreatic Stent (48)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729857501), received between Oct 5, 2017 and Mar 4, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation5
- Positioning Failure3
- Kinked2
- Adverse Event Without Identified Device or Use Problem1
- Difficult or Delayed Positioning1
Patient problems
- No Consequences Or Impact To Patient4
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 4, 2025Positioning Failure; Material Deformation; Adverse Event Without Identified Device or Use Problem; Positioning Failure; Material Deformation; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS TO BE IMPLANTED IN THE PANCREATIC DUCT TO TREAT CHRONIC PANCREATITIS DURING A PANCREATIC DUCT STENT PLACEMENT. DURING THE PROCEDURE, THE STENT COULD NOT BE DEPLOYED. UPON INSPECTION, THE STENT WAS FOUND TO BE FLATTENED. ANOTHER ADVANIX PANCREATIC STENT WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21518554 ↗MalfunctionMay 18, 2023Positioning Failure; Material Deformation; Positioning Failure; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC THAT AN ADVANIX PANCREATIC STENT WAS ATTEMPTED TO BE PLACED DURING AN ENDOSCOPIC PANCREATIC STENT PLACEMENT PROCEDURE IN THE PANCREATIC DUCT PERFORMED ON (B)(6) 2023. DURING THE PROCEDURE, WHEN THE ADVANIX PANCREATIC STENT WAS LOADED INTO THE DELIVERY SYSTEM, THE LUMEN OF THE STENT FELT FLATTENED, HARD AND DIFFICULT TO INSERT. THE ADVANIX PANCREATIC STENT WAS ABLE TO ADVANCE INTO THE TARGET SITE FOR DEPLOYMENT, HOWEVER, DUE TO STRONG RESISTANCE, THE STENT COULD NOT BE RELEASED. THE PROCEDURE WAS COMPLETED WITH ANOTHER ADVANIX PANCREATIC STENT. THERE WERE NO PATIENT COMPLICATIONS OR INJURY REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16957334 ↗MalfunctionFeb 3, 2021Positioning Failure; Material Deformation; Positioning Failure; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING A PANCREATIC DUCT STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6), 2021. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE STENT WAS FOUND TO BE KINKED AND TWISTED LEADING THE STENT TO BE UNABLE TO RELEASE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ADVANIX PANCREATIC STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. IT HAPPENED DURING THE PROCEDURE AND IT WAS NOTICED THAT THE STENT WAS TWISTED WHEN THE DEVICE WAS UNPACKED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11274363 ↗MalfunctionSep 20, 2018Difficult or Delayed Positioning; Material Deformation; Difficult or Delayed Positioning; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE IN THE PANCREAS PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, WHEN THE ADVANIX PANCREATIC STENT WAS ATTEMPTED TO DEPLOY IN THE PANCREAS, SEVERE RESISTANCE WAS ENCOUNTERED WHEN THE STENT WAS RELEASED. REPORTEDLY, WHEN THE PAPILLA SIDE OF THE DEPLOYED PANCREATIC DUCT STENT WAS PULLED, THE STENT KINKED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE; A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7894010 ↗MalfunctionJul 2, 2018Kinked; Kinked
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX¿ PANCREATIC STENT WAS TO BE USED FOR PANCREATIC DUCT STENT PLACEMENT PROCEDURE PERFORMED IN THE PANCREATIC DUCT ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING PREPARATION, WHEN FLUSHING WAS PERFORMED BY SALINE, THE PHYSICIAN FOUND A SMALL BEND IN THE FLAP PART. THE PROCEDURE WAS COMPLETE WITH ANOTHER ADVANIX¿ PANCREATIC STENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE "GOOD".
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7656940 ↗MalfunctionDec 27, 2017Kinked; Kinked
NOTE: THIS REPORT PERTAINS TO ONE OF THREE DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT # 3005099803-2017-03865, 3005099803-2017-03869, 3005099803-2017-03866 FOR THE ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIO-PANCREATOGRAPHY PROCEDURE PERFORMED ON (B)(6) 2017. AS THE PHYSICIAN INTRODUCED A 7FRX 11CM ADVANIX PANCREATIC STENT INTO THE PATIENT¿S PANCREATIC DUCT, HE EXPERIENCED SOME RESISTANCE. THE STENT WAS ADVANCED INTO POSITION AND WHEN THE TECHNICIAN ATTEMPTED TO DEPLOY THE STENT THE DELIVERY SYSTEM AND ADVANIX PANCREATIC STENT BOTH BUCKLED/CRUNCHED UP WHICH WOULD NOT ALLOW THE STENT TO RELEASE FROM THE CATHETER. THE DEVICE WAS REMOVED FROM THE SCOPE. THEY THEN ATTEMPTED TO PLACE A 5FRX9CM ADVANIX PANCREATIC STENT OVER A GUIDEWIRE AND PUSH IT INTO POSITION WITH A HYDRATOME. DURING THE ADVANCING OF THE 5FRX9CM ADVANIX PANCREATIC STENT, IT TOO BUCKLED AND HAD TO BE REMOVED. FINALLY, A NAVIFLEX PANCREATIC DELIVERY SYSTEM WAS LOADED WITH THE 7FRX9CM ADVANIX PANCREATIC STENT AND ATTEMPTED TO BE PUSHED INTO POSITION. THE STENT WAS IN POSITION AN
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7148578 ↗MalfunctionOct 5, 2017Material Deformation; Material Deformation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ADVANIX PANCREATIC STENT USED DURING A PANCREATIC STENT REPLACEMENT PROCEDURE PERFORMED IN THE PANCREATIC DUCT ON (B)(6) 2017. ACCORDING TO THE COMPLAINANT, WHEN THE ADVANIX PANCREATIC STENT WAS ADVANCED TO THE TARGET AREA, THE DEVICE BECAME STUCK AND THE INNER CATHETER WAS UNABLE TO BE PULLED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE WITH NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS; PANCREATIC STENT DEFORMED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6916910 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page