Sterling™ H74939032200410Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Sterling (153)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729859932), received between Apr 15, 2019 and Feb 7, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture3
- Entrapment of Device1
- Inflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 7, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE BELOW THE KNEE VESSEL. A 2.0X40X150 MM STERLING BALLOON CATHETER WAS ADVANCED TO PERFORM DILATATION. DURING THE PROCEDURE, THE BALLOON WAS INFLATED MULTIPLE TIMES; HOWEVER, A PINHOLE RUPTURE OCCURRED IN THE BALLOON TOWARD THE END OF THE PROCEDURE. THE COMPROMISED DEVICE WAS SUCCESSFULLY WITHDRAWN, AND THE INTERVENTION WAS COMPLETED USING ANOTHER IDENTICAL CATHETER. NO COMPLICATIONS WERE REPORTED FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21337640 ↗MalfunctionDec 12, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT UNDERWENT LOWER LIMB ANGIOPLASTY. A 2.0X40X150 (4F) STERLING BALLOON CATHETER WAS SELECTED FOR USE. HOWEVER, DURING PREPARATION, A CONTRAST LEAKAGE AT THE BALLOON SHAFT WAS OBSERVED BEFORE INFLATING THE BALLOON. ADDITIONALLY, A PINHOLE WAS NOTED ON THE BALLOON MATERIAL. THE PROCEDURE WAS COMPLETED WITH A 2X60X150CM STERLING BALLOON CATHETER. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20914630 ↗MalfunctionApr 11, 2024Inflation Problem; Inflation Problem
IT WAS REPORTED THAT THE PROCEDURE WAS NOT COMPLETED. A 2.0X40X150 (4F) STERLING BALLOON WAS SELECTED FOR USE. DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON COULD NOT INFLATE IN THE TARGET LESION. CONSEQUENTLY, THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. NO COMPLICATIONS WERE REPORTED AND THE PATIENT WAS STABLE POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19093148 ↗MalfunctionAug 23, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 2.0X40X150MM STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED BEFORE REACHING THE RATED BURST PRESSURE. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8928535 ↗MalfunctionApr 15, 2019Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT CATHETER ENTRAPMENT ON THE GUIDEWIRE OCCURRED. VASCULAR ACCESS WAS GAINED WITH ANTEGRADE APPROACH. THE 80 PERCENT STENOSED LESION WAS LOCATED IN THE MODERATELY CALCIFIED AND NORMALLY TORTUOUS ANTERIOR TIBIALIS ARTERY IN THE LOWER LEG. A NON-BSC SHEATH WAS PLACED. A 20/300 V-18 CONTROL WIRE WAS ADVANCED. A 2.0X40X150 (4F) STERLING BALLOON CATHETER WAS INTRODUCED OVER THE GUIDEWIRE. THE STENOSIS WAS DILATED WITH THE STERLING BALLOON. AFTER SUCCESSFUL DILATION, AN ATTEMPT TO PULL BACK THE BALLOON WAS MADE; HOWEVER THIS WAS NOT POSSIBLE. THE BALLOON AND THE GUIDEWIRE WERE PULLED OUT OF THE PATIENT TOGETHER. THE PROCEDURE WAS CONSIDERED COMPLETED. NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S STATUS WAS OKAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8516303 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page