Skip to content

Sterling™ H74939032200490Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Sterling (153)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08714729859949), received between May 27, 2021 and May 27, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 27, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BLOOD CAME OUT OF THE INFLATION PORT. A 2MM X 40MM X 90CM STERLING BALLOON CATHETER WAS SELECTED FOR USE IN AN ANGIOPLASTY PROCEDURE BELOW THE KNEE. DURING THE PROCEDURE, THE BALLOON WAS ADVANCED AS USUAL INSIDE THE PATIENT AND THEN BLOOD CAME OUT OF THE INFLATION PORT. AS A RESULT, THE BALLOON WAS NEVER INFLATED. ANOTHER BALLOON WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11899000 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page