Sterling™ H74939032252210Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Sterling (153)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729860075), received between Dec 26, 2019 and Oct 21, 2024 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture3
  • Deflation Problem1
  • Device Markings/Labelling Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 21, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT A LABELING ISSUE OCCURRED. A 2.5X220X150 (4F) STERLING WAS SELECTED FOR ANGIOPLASTY OF THE TIBIAL ARTERY. AFTER THE BALLOON WAS INSERTED INTO THE PATIENT AND INFLATED TO TREAT THE LESION, IT WAS NOTED THAT THE BALLOON APPEARED TO BE TOO SMALL. UPON CHECKING THE SIZE LABEL ON THE STERLING HUB, IT WAS CONFIRMED THAT A 2.0X220 STERLING WAS ACTUALLY IN THE PACKAGE AND INFLATED IN THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER 2.5X220X150 STERLING DEVICE. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20495295 ↗
  • MalfunctionJan 20, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. CANNULATION OF THE SUPERFICIAL FEMORAL ARTERY BY FEMORAL SHEATH 6FR WAS PERFORMED. THE 50% STENOSED TARGET LESION WAS LOCATED IN THE NON-TORTUOUS AND MODERATELY CALCIFIED INTERIOR TIBIAL ARTERY. AFTER A USING 0.018 V18 GUIDEWIRE CROSSED THE LESION, A 2.5X220X150 STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, AFTER INITIAL INFLATION AT 4 ATMOSPHERES FOR 5 SECONDS, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME BALLOON WITH SUCCESSFUL ANGIOPLASTY DONE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND PATIENT STATUS WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16213846 ↗
  • InjuryMay 12, 2021Deflation Problem; Deflation Problem

    IT WAS REPORTED THAT BALLOON DEFLATION FAILURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE POSTERIOR TIBIAL ARTERY. A 2.5X220X150 STERLING BALLOON CATHETER WAS ADVANCED FOR POST DILATATION. HOWEVER, AFTER THE BALLOON WAS INFLATED AT 6 ATMOSPHERES, IT WOULD NOT DEFLATE. THE BALLOON WAS ATTEMPTED TO DEFLATE BY PULLING NEGATIVE PRESSURE BUT WAS UNSUCCESSFUL. THE DEVICE WAS FINALLY REMOVED WHILE IT WAS STILL INFLATED BY MANUAL EXTRACTION. THE PROCEDURE WAS COMPLETED WITH A NON-BOSTON SCIENTIFIC DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT WAS DOING WELL. IT WAS FURTHER REPORTED THAT THE BALLOON NEVER DEFLATED. A 50/50 SALINE-CONTRAST RATIO WAS USED. A VASCULAR SURGEON REMOVED THE BALLOON INTACT OVER THE COURSE OF FOUR HOURS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11813170 ↗
  • MalfunctionMar 18, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 2.5X220X150 (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION, HOWEVER; IT WAS NOTED THAT THE CONTRAST WAS DRIPPING AT THE END OF THE BALLOON WHEN IT WAS INFLATED AS THE BALLOON RUPTURED. THE DEVICE WAS COMPLETELY REMOVED AND NO PATIENT COMPLICATIONS WERE REPORTED. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE AND THE PATIENT'S STATUS WAS GOOD.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11520775 ↗
  • MalfunctionDec 26, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN MID TO DISTAL ANTERIOR TIBIAL ARTERY. A 2.5X220X150 (4F) STERLING BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED BEFORE REACHING THE RATED BURST PRESSURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS STABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9522203 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →