Skip to content

LithoVue M0067911000Recalls & reported problems

Boston Scientific Corporation · System Workstation · Class 2

All models of LithoVue (6)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729874522), received between Sep 10, 2020 and Feb 20, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image4
  • Appropriate Device Problem Term/Code Not Available2
  • Device Displays Incorrect Message2
  • Failure to Power Up2
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 20, 2026Failure to Power Up

    IT WAS REPORTED DURING THE URETEROSCOPIC PROCEDURE, IT WAS NOTED THAT THE UNIT WON'T TURN ON. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THERE WAS NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24399922 ↗
  • MalfunctionJul 25, 2024No Display/Image; Device Displays Incorrect Message; No Display/Image; Device Displays Incorrect Message

    IT WAS REPORTED THAT DURING THE PREPARATION, TWO LITHOVUE FLEXSCOPES WERE USED AND PLUGGED INTO THE MONITOR AND "HARDWARE MALFUNCTION" USER MESSAGE KEPT ON APPEARING ON THE MONITOR SCREEN. THE PROCEDURE WAS CANCELLED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19833788 ↗
  • MalfunctionNov 8, 2022Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A LITHOVUE CAPITAL WAS INTENDED TO BE USED IN A URETEROSCOPY PROCEDURE PERFORMED ON (B)(6), 2022. DURING PROCEDURE, THE LITHOVUE CAPITAL POWER ADAPTER WAS HAVING ISSUES AND WENT OUT. THE PROCEDURE WAS CANCELLED DUE TO THIS EVENT. THERE WAS NO SERIOUS INJURY/ADVERSE EFFECT TO PATIENT AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15753824 ↗
  • MalfunctionMay 27, 2022No Display/Image; Device Displays Incorrect Message; No Display/Image; Device Displays Incorrect Message

    NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES THAT WERE USED IN THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT NUMBER MFR. REPORT # 3005099803-2022-02815 FOR THE LITHOVUE CAPITAL AND MFR. REPORT # 3005099803-2022-02816 FOR THE LITHOVUE FLEXSCOPE. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A LITHOVUE CAPITAL AND A LITHOVUE FLEXSCOPE WAS USED IN A PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, A USER MESSAGE APPEARED ON THE SCREEN WHEN THE FLEXSCOPE WAS CONNECTED. THE PROCEDURE WAS CANCELLED DUE TO THIS EVENT. THERE WAS NO REPORTED SERIOUS INJURY/ADVERSE EFFECT TO PATIENT AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14522207 ↗
  • MalfunctionFeb 17, 2022No Display/Image; No Display/Image

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A LITHOVUE SCOPE WAS ATTEMPTED TO BE USED IN THE KIDNEY DURING A URETEROSCOPY PROCEDURE PERFORMED ON (B)(6) 2022. DURING PROCEDURE, THE SCOPE WAS PLUGGED INTO THE LITHOVUE WORKSTATION AND NO IMAGE APPEARED. ANOTHER SCOPE WAS OPENED TO TEST IF THIS WAS AN ISSUE WITH THE WORK STATION AND THEM SAME THING HAPPENED. THE PROCEDURE WAS CANCELLED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13548684 ↗
  • MalfunctionJan 5, 2021Failure to Power Up; Appropriate Device Problem Term/Code Not Available; Failure to Power Up; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A LITHOVUE TOUCH PC WAS USED IN A URETEROSCOPY PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE LITHOVUE TOUCH PC WAS USED AND THERE WAS NO IMAGE ON THE SCREEN. THE PROCEDURE WAS CANCELLED DUE TO THIS EVENT. THERE WAS NO SERIOUS INJURY/ADVERSE EFFECT TO PATIENT AS A RESULT OF THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 11118054 ↗
  • MalfunctionSep 10, 2020No Display/Image; Appropriate Device Problem Term/Code Not Available; No Display/Image; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A LITHOVUE TOUCH PC WAS USED DURING A KIDNEY STONE TREATMENT PROCEDURE PERFORMED ON (B)(6) 2020. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE LITHOVUE FLEXSCOPE WAS CONNECTED TO THE MONITOR AND A HARDWARE MALFUNCTION ERROR AS WELL AS SCOPE DISCONNECTED ERROR OCCURRED. THE MONITOR WAS UNPLUGGED AND PLUGGED BACK IN HOWEVER IT DID NOT RESOLVE THE ISSUE. A SECOND LITHOVUE FLEXSCOPE WAS OPENED AND THE SAME ISSUE OCCURRED. THE LITHOVUE SCOPES WERE TESTED WITH ANOTHER LITHOVUE MONITOR AND WORKED FINE THEREFORE IT WAS DETERMINED THE ISSUE WAS WITH THE TOUCH PC. THE PROCEDURE WAS CANCELLED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10514880 ↗

Other Ureteroscope and accessories, flexible/rigid devices

All Ureteroscope and accessories, flexible/rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page