FFR Link H7495551000Recalls & reported problems

Boston Scientific Corporation · FFR Signal Processing Module · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729890010), received between May 6, 2024 and Jun 25, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Communication or Transmission Problem2
  • Defective Device1
  • Failure to Power Up1
  • Material Deformation1
  • Nonstandard Device1
  • Unable to Obtain Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 25, 2026Communication or Transmission Problem

    IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. A FFR LINK AND GE PD OUT CABLE WERE SELECTED FOR USE. DURING CAPITAL EQUIPMENT MAINTENANCE, NO SIGNAL DISPLAYED. THE PD OUT CONNECTION PORT SEEMS TO BE DAMAGED ON THE FFR LINK. THE PROCEDURE WAS CANCELLED. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25631412 ↗
  • MalfunctionMar 4, 2026Material Deformation

    IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. A FFR LINK WAS SELECTED FOR USE. IT WAS REPORTED THAT THE POLE WAS DAMAGED. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24512512 ↗
  • MalfunctionJun 11, 2025Failure to Power Up; Failure to Power Up

    IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. THE FFR LINK WAS SELECTED FOR USE IN A PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY PROCEDURE. HOWEVER, DURING PREPARATION, THE DEVICE WOULD NOT START, AND THE POWER LED WAS RED AND BLINKING. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22183059 ↗
  • MalfunctionMar 14, 2025Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. AN FFR LINK WAS SELECTED FOR USE. DURING PROCEDURE, IT WAS NOTED THAT THE FFR LINK WOULD NOT GET A WAVEFORM. THE CASE WAS CANCELED AND THEY HAD THE PATIENT COME BACK THE NEXT DAY. NO COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21606592 ↗
  • MalfunctionNov 26, 2024Nonstandard Device; Defective Device; Nonstandard Device; Defective Device

    IT WAS REPORTED THAT THE PROCEDURE WAS CANCELLED. THE FFR LINK WAS SELECTED FOR TREATMENT. HOWEVER, THE POWER CODE WAS FOUND BROKEN AND THE DEVICE COULD NOT BE TURNED ON. THE PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. THERE WERE NO PATIENT COMPLICATIONS AND THE PATIENT WAS IN GOOD CONDITION. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS NOT CANCELLED. IT WAS COMPLETED WITH OTHER EQUIPMENT AND THE PATIENT WAS UNDER LOCAL ANESTHESIA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20780329 ↗
  • MalfunctionMay 6, 2024Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT THE PROCEDURE WAS NOT COMPLETED. AN FFR LINK WAS SELECTED FOR USE. DURING THE PROCEDURE, THE DEVICE WAS ABLE TO BE ZEROED SUCCESSFULLY BUT WAS UNABLE TO RECORD. AN "UNEXPECTED BEAT" ERROR WAS RECURRING. REBOOTING THE SYSTEM AND REPLACING THE COMET PRESSURE WIRE DID NOT RESOLVE THE ISSUE AND THE CASE WAS ABANDONED. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19249965 ↗

Other COMPUTER, DIAGNOSTIC, PROGRAMMABLE devices

All COMPUTER, DIAGNOSTIC, PROGRAMMABLE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.