Epic™ Biliary M00572270Recalls & reported problems
Boston Scientific Corporation · Stent System · Class 2 · Implant
All models of Epic Biliary (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729896890), received between Oct 12, 2018 and Dec 2, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Difficult or Delayed Positioning” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Difficult or Delayed Positioning2
- Break1
- Difficult to Remove1
- Material Deformation1
- Material Integrity Problem1
- Positioning Problem1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 2, 2019Difficult or Delayed Positioning; Material Deformation; Material Integrity Problem; Positioning Problem; Difficult or Delayed Positioning; Material Deformation; Material Integrity Problem; Positioning Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON NOVEMBER 06, 2019 THAT AN EPIC BILIARY STENT HAS BEEN IMPLANTED TO TREAT A MALIGNANT STENOSIS IN THE LOWER BILE DUCT DURING AN ENDOSCOPIC BILIARY STENT PLACEMENT PROCEDURE PERFORMED ON AN UNKNOWN DATE. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE SHEATH STOPPED MOVING AND THE STENT FAILED TO DEPLOY. THE SHEATH WAS PULLED WITH FORCE AND WAS ABLE TO REMOVE THE DEVICE FROM THE BILE DUCT. REPORTEDLY, THE STENT WAS REINSERTED AND WAS RELEASED ALL AT ONCE NEAR THE TARGET LESION. REPORTEDLY, THE STENT REMAINS IMPLANTED BUT A DIFFERENT STENT WAS PLACED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. NOTE: A PHOTO OF THE COMPLAINT DEVICE WAS RECEIVED, AND UPON REVIEW THE SHEATH WAS NOTED TO BE DAMAGED AND THE INNER SHAFT WAS KINKED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9396309 ↗MalfunctionOct 12, 2018Break; Difficult or Delayed Positioning; Difficult to Remove; Break; Difficult or Delayed Positioning; Difficult to Remove
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2018 THAT AN EPIC BILIARY ENDOSCOPIC STENT WAS IMPLANTED IN THE COMMON BILE DUCT DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING DEPLOYMENT, THE THUMBWHEEL OF THE HANDLE WAS BEING USED; THE STENT WAS ABLE TO BE RELEASED TO AN EXTENT BUT THEN COULD NOT BE FULLY RELEASED. A "SLAP" SOUND WAS HEARD AND THE DISTAL PART OF THE OUTER TUBE GOT SEPARATED. AT THIS TIME THE PHYSICIAN JUDGED THAT THE STENT AND DELIVERY SYSTEM COULD NOT BE RETRIEVED AS IT WAS, AND THE DELIVERY SYSTEM WAS CUT WITH PLIERS WHERE THE HANDLE MEETS THE OUTER AND INNER TUBES. THE OUTER AND INNER TUBE WERE REMOVED TOGETHER WITH THE ENDOSCOPE. IT WAS REPORTED THAT AFTER DELIVERY SYSTEM REMOVAL THE STENT WAS FOUND TO BE PLACED IN THE ORIGINALLY PLANNED LOCATION AND THE STENT REMAINED IMPLANTED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7960417 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page