Epic™ Biliary M00572290Recalls & reported problems
Boston Scientific Corporation · Stent System · Class 2 · Implant
All models of Epic Biliary (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 08714729896913), received between Apr 26, 2018 and Nov 20, 2025 — classified as 3 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Activation, Positioning or Separation Problem4
- Premature Activation4
- Detachment of Device or Device Component3
- Difficult to Advance3
- Material Deformation2
- Use of Device Problem2
- Difficult to Remove1
- Positioning Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 20, 2025Premature Activation; Premature Activation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN EPIC BILIARY STENT WAS TO BE IMPLANTED IN THE BILE DUCT TO TREAT A MALIGNANT HILAR TUMOR DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY PROCEDURE (ERCP) PERFORMED ON AN UNKNOWN DATE. THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE EPIC STENT WAS ADVANCED OVER THE GUIDEWIRE. HOWEVER, AS THE STENT EXITED THE SCOPE, IT WAS NOTED THAT IT HAD PARTIALLY DEPLOYED AT THE TIP OF THE DEVICE BEFORE THE DEPLOYMENT PROCESS HAD BEEN INITIATED. THE PREMATURE DEPLOYED STENT WAS REMOVED, AND THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE OF THE SAME MODEL. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23610422 ↗MalfunctionApr 24, 2025Detachment of Device or Device Component; Difficult to Advance; Detachment of Device or Device Component; Difficult to Advance
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN EPIC BILIARY ENDOSCOPIC STENT WAS TO BE IMPLANTED IN THE BILIARY TREE TO TREAT A BILE DUCT STRICTURE DURING THE ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE STENT WAS ATTEMPTED TO BE ADVANCED OVER THE WIRE, BUT IT WOULD NOT PASS, AND THE TIP OF CATHETER DETACHED FROM THE STENT. THE PROCEDURE WAS COMPLETED USING ANOTHER OF SAME DEVICE. THERE WAS NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21903935 ↗MalfunctionApr 30, 2024Premature Activation; Detachment of Device or Device Component; Premature Activation; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN EPIC BILIARY STENT WAS TO BE IMPLANTED IN THE BILE DUCT TO TREAT METASTATIC CANCER DURING A BILATERAL STENTING PROCEDURE PERFORMED ON (B)(6) 2024. THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, WHEN ADVANCING INTO THE WORKING CHANNEL OF THE DUODENOSCOPE, THE STENT PREMATURELY DEPLOYED. THE STENT WAS STUCK IN THE WORKING CHANNEL AND THE INNER SHEATH BROKE OFF. SUBSEQUENTLY, THE STENT WAS RETRIEVED WITH A SCOPE BRUSH. THE PROCEDURE WAS COMPLETED WITH JUST ONE EPIC BILIARY STENT OF A DIFFERENT SIZE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19217263 ↗MalfunctionJul 19, 2023Use of Device Problem; Activation, Positioning or Separation Problem; Difficult to Advance; Material Deformation; Use of Device Problem; Activation, Positioning or Separation Problem; Difficult to Advance; Material Deformation
NOTE: THIS REPORT PERTAINS TO ONE OF THE FOUR DEVICES USED ON THE SAME PATIENT. REFER TO MANUFACTURER REPORT # 3005099803-2023-03775, 3005099803-2023-03778, 3005099803-2023-03835, AND 3005099803-2023-03836 FOR THE ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2023, THAT FOUR (4) EPIC BILIARY STENTS WERE TO BE IMPLANTED IN THE COMMON BILE DUCT AT THE HEPATIC BIFURCATION TO TREAT A 2 CM MALIGNANT KLATSKIN'S TUMOR DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH SELF-EXPANDABLE METALLIC STENT (SEMS) IMPLANTATION PROCEDURE PERFORMED ON (B)(6) 2023. THE PATIENT'S ANATOMY WAS TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, TWO EPIC BILIARY STENTS WERE ATTEMPTED TO BE PLACED BILATERALLY AT THE SAME TIME. THE FIRST EPIC BILIARY STENT (THE SUBJECT OF MFR. REPORT # 3005099803-2023-03775) WAS LODGED IN THE WORKING CHANNEL OF THE SCOPE, THE DELIVERY SYSTEM COULD NOT BE ADVANCED, AND THE STENT WAS REMOVED FROM THE PATIENT PARTIALLY DEPLOYED ON THE DELIVERY SYSTEM. THE SECOND EPIC BILIARY STENT (THE SUBJECT OF MFR. REPORT # 3005099803-2023-03778) WAS ALSO DIFFICULT TO ADVANCE THROUGH THE SCOPE, AND THE O
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17350073 ↗MalfunctionJul 19, 2023Use of Device Problem; Activation, Positioning or Separation Problem; Difficult to Advance; Material Deformation; Use of Device Problem; Activation, Positioning or Separation Problem; Difficult to Advance; Material Deformation
NOTE: THIS REPORT PERTAINS TO ONE OF THE FOUR DEVICES USED ON THE SAME PATIENT. REFER TO MANUFACTURER REPORT # 3005099803-2023-03775, 3005099803-2023-03778, 3005099803-2023-03835, FOR THE ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6), 2023, THAT FOUR (4) EPIC BILIARY STENTS WERE TO BE IMPLANTED IN THE COMMON BILE DUCT AT THE HEPATIC BIFURCATION TO TREAT A 2 CM MALIGNANT KLATSKIN'S TUMOR DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH SELF-EXPANDABLE METALLIC STENT (SEMS) IMPLANTATION PROCEDURE PERFORMED ON (B)(6), 2023. THE PATIENT'S ANATOMY WAS TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, TWO EPIC BILIARY STENTS WERE ATTEMPTED TO BE PLACED BILATERALLY AT THE SAME TIME. THE FIRST EPIC BILIARY STENT (THE SUBJECT OF MFR. REPORT # 3005099803-2023-03775) WAS LODGED IN THE WORKING CHANNEL OF THE SCOPE, DELIVERY SYSTEM COULD NOT BE ADVANCED, AND THE STENT WAS REMOVED FROM THE PATIENT PARTIALLY DEPLOYED ON THE DELIVERY SYSTEM. THE SECOND EPIC BILIARY STENT (THE SUBJECT OF MFR. REPORT # 3005099803-2023-03778) WAS ALSO DIFFICULT TO ADVANCE THROUGH THE SCOPE, AND THE OUTER SHEATH BECAME BUCKLED A
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17349950 ↗InjuryJun 29, 2022Premature Activation; Premature Activation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2022 THAT AN EPIC BILIARY ENDOSCOPIC STENT WAS IMPLANTED TO TREAT A MALIGNANT STENOSIS IN THE BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2022. REPORTEDLY, THE PATIENT'S ANATOMY WAS NOT TORTUOUS AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. DURING THE PROCEDURE, THE EPIC BILIARY STENT ACCIDENTALLY DEPLOYED OUTSIDE THE AMPULLA. THERE WAS NO VISIBLE DAMAGE TO THE EPIC BILIARY STENT, WHICH REMAINED IMPLANTED. A 6CM WALLFLEX UNCOVERED STENT WAS IMPLANTED TO BRIDGE THE AMPULLA TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14858708 ↗MalfunctionApr 13, 2021Positioning Failure; Activation, Positioning or Separation Problem; Positioning Failure; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN EPIC BILIARY ENDOSCOPIC STENT WAS TO BE IMPLANTED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH STENT PLACEMENT PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, THE STENT WAS UNABLE TO DEPLOY. THE STENT WAS REMOVED FROM THE PATIENT AND ANOTHER EPIC BILIARY STENT WAS IMPLANTED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: THIS EVENT HAS BEEN DEEMED AN MDR REPORTABLE EVENT BASED ON THE INVESTIGATION RESULTS WHICH REVEALED THAT THE STENT WAS PARTIALLY DEPLOYED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11657007 ↗MalfunctionNov 5, 2018Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2018 THAT AN EPIC BILIARY STENT WAS TO BE USED TO TREAT A MALIGNANT STRICTURE DUE TO BILE DUCT CANCER DURING A BILIARY STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2018. REPORTEDLY, THE PATIENT'S ANATOMY WAS TORTUOUS. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE STENT WAS ATTEMPTED TO BE DEPLOYED IN THE LOWER PART OF THE BILE DUCT. THE HANDLE WAS ROTATED BUT THE STENT COULD NOT BE DEPLOYED. REPORTEDLY, THE STENT SUDDENLY BEGAN TO DEPLOY, AND A LONG PART OF THE STENT CAME OUT IN THE DUODENUM. THE STENT WAS REMOVED FROM THE PATIENT AND ANOTHER EPIC BILIARY STENT WAS PLACED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8039543 ↗InjuryJun 13, 2018Premature Activation; Premature Activation
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2018 THAT AN EPIC BILIARY STENT HAS BEEN IMPLANTED TO TREAT A STENOSIS IN THE BILE DUCT DUE TO GALL BLADDER CANCER DURING A STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2018. REPORTEDLY, THE PATIENT'S ANATOMY WAS TIGHT AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE AND SUBSEQUENT TO FLUSHING, AN ENDOSCOPE WAS INSERTED INSIDE THE PATIENT AND A GUIDEWIRE WAS PLACED. THE STENT WAS ADVANCED INTO THE TARGET SITE AND WAS ATTEMPTED TO BE DEPLOYED BY ROTATING THE THUMBWHEEL, HOWEVER, THE OUTER CATHETER DID NOT MOVE AND THE STENT COULD NOT BE DEPLOYED. AFTER FULLY ROTATING THE THUMBWHEEL, THE STENT STILL FAILED TO DEPLOY. WHILE REMOVING THE DELIVERY SYSTEM, THE STENT WAS SUDDENLY RELEASED WITH ABOUT 4 CM OF THE STENT HANGING OUT FROM THE PAPILLA. AN ARGON PLASMA COAGULATION WAS PERFORMED TO TRIM THE PORTION OF THE STENT PART OUTSIDE THE PAPILLA AND THE REMAINING STENT WAS LEFT IMPLANTED INSIDE THE PATIENT. THE PATIENT WAS REPORTED TO BE STABLE AND NO ADDITIONAL TREATMENT HAS BEEN PERFORMED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7598623 ↗InjuryApr 26, 2018Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2018 THAT AN EPIC BILIARY STENT WAS IMPLANTED TO TREAT A MALIGNANT STRICTURE IN THE MID COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) WITH STENT PLACEMENT PROCEDURE PERFORMED ON (B)(6) 2018. REPORTEDLY, THE PATIENT¿S ANATOMY WAS VERY TIGHT AND WAS NOT DILATED PRIOR TO STENT PLACEMENT. ACCORDING TO THE COMPLAINANT, DURING PROCEDURE, THE STENT WAS ABLE TO BE DEPLOYED. HOWEVER, THE DELIVERY SYSTEM WAS STUCK DURING WITHDRAWAL. THE GUIDEWIRE AND THE DELIVERY SYSTEM WERE PULLED OUT TOGETHER. WHILE PULLING OUT THE DELIVERY SYSTEM, THE INNER CATHETER SNAPPED. A PORTION OF THE INNER CATHETER DETACHED AND REMAINED WITHIN THE EPIC STENT. THE DETACHED PIECE OF THE INNER CATHETER WAS REMOVED FROM THE PATIENT WITH FORCEPS. THE EPIC STENT REMAINED IN THE INTENDED LOCATION WITH THE PATIENT AND AN 8X80 WALLFLEX BILIARY UNCOVERED STENT WAS PLACED INSIDE THE EPIC BILIARY STENT AS A BRIDGE TO ENSURE PROPER DRAINAGE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7466518 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page