DAKOTA M0063905010Recalls & reported problems

Boston Scientific Corporation · Nitinol Stone Retrieval Device · Class 2

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08714729937371), received between Jun 20, 2023 and Nov 28, 2025 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component2
  • Difficult to Open or Close2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 28, 2025Detachment of Device or Device Component; Difficult to Open or Close; Detachment of Device or Device Component; Difficult to Open or Close

    IT WAS REPORTED THAT A DAKOTA RETRIEVAL BASKET WAS USED DURING A URETEROSCOPIC STONE EXTRACTION PROCEDURE. REPORTEDLY, THE BASKET WAS OPENING AND CLOSING AT FIRST BUT THEN IT STOPPED DEPLOYING. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS. ANALYSIS OF THE RETURNED BASKET IDENTIFIED THAT THE GRASPER WAS DETACHED FROM THE SHEATH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23671552 ↗
  • InjuryJul 15, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A DAKOTA RETRIEVAL BASKET WAS USED DURING A LITHOTRIPSY PROCEDURE. DURING THE PROCEDURE, AFTER USING A LASER FIBER, AND A DAKOTA BASKET TO REMOVE FRAGMENTS, THE BASKET BROKE. REPORTEDLY, THE DISTAL BASKET WAS REMOVED AND A CYSTOSCOPE ADVANCES PER URETHRA TO REMOVE THE BROKEN BASKET FRAGMENTS. THE PROCEDURE WAS COMPLETED WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22508427 ↗
  • MalfunctionApr 30, 2025Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT A DAKOTA RETRIEVAL BASKET WAS USED DURING A BENIGN PROSTATIC HYPERPLASIA (BPH) PROCEDURE. DURING THE PROCEDURE, THE BASKET DID NOT OPEN. ADDITIONALLY, THERE WAS A STONE INSIDE THE BASKET WHEN IT FAILED TO OPEN AND WAS REMOVED USING ANOTHER BASKET. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE WITH NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21938444 ↗
  • InjuryAug 30, 2023Break; Break

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A DAKOTA BASKET WAS USED IN THE KIDNEY DURING A PERCUTANEOUS NEPHROLITHOTOMY PROCEDURE PERFORMED ON (B)(6) 2023. ON (B)(6) 2023, SURGERY WAS PERFORMED AND FOUND A PART OF THE END EFFECTOR WIRE OF THE DAKOTA WAS COMPLETELY DETACHED AND HAS REMAINED INSIDE THE PATIENT. THE WIRE WAS RETRIEVED SURGICALLY. IT WAS UNKNOWN THAT THE WIRE OF THE BASKET DETACHED UNTIL THE SURGERY. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17650197 ↗
  • InjuryJun 20, 2023Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT THE DAKOTA BASKET WAS USED DURING PROCEDURE PERFORMED ON (B)(6) 2023. DURING A PROCEDURE, THE DAKOTA BASKET BROKE OFF INSIDE THE PATIENT NEAR THE RIGHT URETER. THE PHYSICIAN SUCCESSFULLY LOCATED AND REMOVED ALL THE BROKEN PIECES OF THE BASKET. NO ADDITIONAL FRAGMENTS WERE FOUND DURING THE PROCEDURE. THE COLLECTED BASKET FRAGMENTS WERE SECURELY PACKAGED IN A BAG FOR REPORTING AND FURTHER INVESTIGATION. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. BOSTON SCIENTIFIC HAS BEEN UNABLE TO OBTAIN ADDITIONAL INFORMATION REGARDING THE EVENT TO DATE, DESPITE GOOD FAITH EFFORTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17165573 ↗

Other DISLODGER, STONE, BASKET, URETERAL, METAL devices

All DISLODGER, STONE, BASKET, URETERAL, METAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.