RHYTHMIA HDx™ M004RBINSTALL2EN0Recalls & reported problems

Boston Scientific Corporation · Mapping System · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729938682), received between Oct 28, 2020 and Nov 18, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Computer Operating System Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Computer Operating System Problem1
  • Data Problem1
  • Device Displays Incorrect Message1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 18, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE RHYTHMIA MAGNET WAS BROKEN, AND THE PROCEDURE WAS CANCELLED. DURING A REDO CRYOABLATION PROCEDURE TO TREAT ATRIAL FIBRILLATION (AF), A RHYTHMIA HDX WAS SELECTED FOR USE. IT WAS OBSERVED THAT THE RHYTHMIA MAGNET WAS NOT FUNCTIONING. AS A RESULT, THE PROCEDURE WAS CANCELLED. THE PATIENT WAS UNDER GENERAL ANESTHESIA WHEN THE ISSUE OCCURRED. THERE WERE NO PATIENT COMPLICATIONS. ONSITE SERVICE HAS BEEN REQUESTED; IT WAS IDENTIFIED THAT THERE WAS AN ISSUE WITH THE LOCALIZATION GENERATOR AND CABLE. IT IS UNKNOWN IF THE DEVICE OR ANY COMPONENTS WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20710931 ↗
  • MalfunctionJun 6, 2023Computer Operating System Problem; Computer Operating System Problem

    IT WAS REPORTED THAT DURING A PROCEDURE, A RHYTHMIA HDX MAPPING SYSTEM WAS SELECTED FOR USE. THE SYSTEM WAS LAGGING DURING THE SECOND CASE OF THE DAY. THE MOUSE CURSOR WAS SHOWING THE SPINNING CIRCLE CONSTANTLY AND INPUTTING ANY INFORMATION OR DATA WOULD TAKE SEVERAL SECONDS TO INPUT AND SHOW ON THE SCREEN. PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. PATIENT SEDATION STATUS IS UNKNOWN. CONSOLE IS NOT EXPECTED TO BE RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17075555 ↗
  • MalfunctionOct 28, 2020Device Displays Incorrect Message; Data Problem; Device Displays Incorrect Message; Data Problem

    IT WAS REPORTED THAT DUE TO A DEVICE ISSUE THE PROCEDURE WAS CANCELLED. DURING A PROCEDURE AN BACKPATCH WAS SELECTED FOR USE. WHEN THE PATIENT WAS REMOVED FROM THE PROCEDURE TABLE THERE WAS NO TRACKING OF THE CATHETER ON THE RHYTHMIA SYSTEM. THE RHYTHMIA MAPPING SPECIALIST (RMS) REBOOTED THE WORKSTATION AND SIGNAL STATION WHICH DID NOT RESOLVE THE ISSUE. REPLACING THE CABLE AND THE CATHETER ALSO DID NOT RESOLVE THE ISSUE. THEREFORE THE ABLATION PORTION OF THE PROCEDURE WAS CANCELLED AND THE PHYSICIAN PROCEEDED TO IMPLANT THE CARDIAC RESYNCHRONIZATION THERAPY DEVICE IN THE PATIENT SUCCESSFULLY WITHOUT ANY PATIENT COMPLICATIONS. ADDITIONAL INFORMATION REVEALED THAT AFTER THE CASE THE RMS REPLACED THE BACKPATCH AND THE PROBLEM WAS RESOLVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10747612 ↗

Other COMPUTER, DIAGNOSTIC, PROGRAMMABLE devices

All COMPUTER, DIAGNOSTIC, PROGRAMMABLE devices →