RHYTHMIA HDx™ M004RBINSTALL2EN0Recalls & reported problems
Boston Scientific Corporation · Mapping System · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08714729938682), received between Oct 28, 2020 and Nov 18, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Computer Operating System Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Computer Operating System Problem1
- Data Problem1
- Device Displays Incorrect Message1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 18, 2024Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE RHYTHMIA MAGNET WAS BROKEN, AND THE PROCEDURE WAS CANCELLED. DURING A REDO CRYOABLATION PROCEDURE TO TREAT ATRIAL FIBRILLATION (AF), A RHYTHMIA HDX WAS SELECTED FOR USE. IT WAS OBSERVED THAT THE RHYTHMIA MAGNET WAS NOT FUNCTIONING. AS A RESULT, THE PROCEDURE WAS CANCELLED. THE PATIENT WAS UNDER GENERAL ANESTHESIA WHEN THE ISSUE OCCURRED. THERE WERE NO PATIENT COMPLICATIONS. ONSITE SERVICE HAS BEEN REQUESTED; IT WAS IDENTIFIED THAT THERE WAS AN ISSUE WITH THE LOCALIZATION GENERATOR AND CABLE. IT IS UNKNOWN IF THE DEVICE OR ANY COMPONENTS WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20710931 ↗MalfunctionJun 6, 2023Computer Operating System Problem; Computer Operating System Problem
IT WAS REPORTED THAT DURING A PROCEDURE, A RHYTHMIA HDX MAPPING SYSTEM WAS SELECTED FOR USE. THE SYSTEM WAS LAGGING DURING THE SECOND CASE OF THE DAY. THE MOUSE CURSOR WAS SHOWING THE SPINNING CIRCLE CONSTANTLY AND INPUTTING ANY INFORMATION OR DATA WOULD TAKE SEVERAL SECONDS TO INPUT AND SHOW ON THE SCREEN. PROCEDURE WAS NOT COMPLETED DUE TO THIS EVENT. PATIENT SEDATION STATUS IS UNKNOWN. CONSOLE IS NOT EXPECTED TO BE RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17075555 ↗MalfunctionOct 28, 2020Device Displays Incorrect Message; Data Problem; Device Displays Incorrect Message; Data Problem
IT WAS REPORTED THAT DUE TO A DEVICE ISSUE THE PROCEDURE WAS CANCELLED. DURING A PROCEDURE AN BACKPATCH WAS SELECTED FOR USE. WHEN THE PATIENT WAS REMOVED FROM THE PROCEDURE TABLE THERE WAS NO TRACKING OF THE CATHETER ON THE RHYTHMIA SYSTEM. THE RHYTHMIA MAPPING SPECIALIST (RMS) REBOOTED THE WORKSTATION AND SIGNAL STATION WHICH DID NOT RESOLVE THE ISSUE. REPLACING THE CABLE AND THE CATHETER ALSO DID NOT RESOLVE THE ISSUE. THEREFORE THE ABLATION PORTION OF THE PROCEDURE WAS CANCELLED AND THE PHYSICIAN PROCEEDED TO IMPLANT THE CARDIAC RESYNCHRONIZATION THERAPY DEVICE IN THE PATIENT SUCCESSFULLY WITHOUT ANY PATIENT COMPLICATIONS. ADDITIONAL INFORMATION REVEALED THAT AFTER THE CASE THE RMS REPLACED THE BACKPATCH AND THE PROBLEM WAS RESOLVED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10747612 ↗
Other COMPUTER, DIAGNOSTIC, PROGRAMMABLE devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page