Skip to content

CRE™ RX M00558890Recalls & reported problems

Boston Scientific Corporation · Biliary Balloon Dilatation Catheter · Class 2

All models of CRE RX (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08714729965442), received between Jan 22, 2019 and Aug 12, 2026 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Entrapment of Device2
  • Tear, Rip or Hole in Device Packaging2
  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Advance1
  • Problem with Sterilization1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 12, 2026Detachment of Device or Device Component; Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50217663 ↗
  • MalfunctionJul 16, 2026Tear, Rip or Hole in Device Packaging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25850511 ↗
  • MalfunctionJul 16, 2026Tear, Rip or Hole in Device Packaging

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25850272 ↗
  • InjuryApr 28, 2026Entrapment of Device; Problem with Sterilization; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE RX DILATATION BALLOON WAS USED IN THE ESOPHAGUS, STOMACH AND DUODENUM DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE FOR DIAGNOSTICS PERFORMED ON (B)(6) 2026 DURING A PREVIOUS PROCEDURE PERFORMED ON AN UNKNOWN DATE, WHEN A CRE RX DILATATION BALLOON WAS INSERTED INTO THE WORKING CHANNEL OF THE SCOPE, THE BALLOON GOT STUCK HALFWAY. THE BALLOON WAS REMOVED. IT WAS REPORTED THAT RX TUNNEL BECAME LOOSE FROM THE PURPLE SHAFT AND WAS DETACHED AND LEFT INSIDE THE SCOPE UNNOTICED. DURING ANOTHER PROCEDURE ON (B)(6) 2026, WHILE USING A BIOPSY FORCEPS, IT WAS REPORTED THAT THE DETACHED RX TUNNEL WAS PUSHED OUT FROM THE WORKING CHANNEL OF THE SCOPE AND INTO THE PATIENT. UPON CHECKING, IT WAS POSITIVELY IDENTIFIED TO HAVE COME FROM THE CRE RX BALLOON FROM THE PREVIOUS PROCEDURE. THE DETACHED PART WAS SUCCESSFULLY RETRIEVED WITH A GRASPER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023650 ↗
  • MalfunctionApr 28, 2026Entrapment of Device; Detachment of Device or Device Component

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE RX DILATATION BALLOON WAS USED DURING A PROCEDURE PERFORMED ON AN UNKNOWN DATE. DURING THE PROCEDURE, WHEN A CRE RX DILATATION BALLOON WAS INSERTED INTO THE WORKING CHANNEL OF THE SCOPE, THE BALLOON GOT STUCK HALFWAY. THE BALLOON WAS REMOVED. IT WAS REPORTED THAT RX TUNNEL BECAME LOOSE FROM THE PURPLE SHAFT AND WAS DETACHED AND LEFT INSIDE THE SCOPE UNNOTICED. DURING A SEPARATE PROCEDURE ON MARCH 27, 2026, THE BIOPSY FORCEPS PUSHED OUT A FOREIGN MATERIAL FROM THE WORKING CHANNEL OF THE SCOPE AND INTO THE PATIENT. UPON CHECKING, IT WAS POSITIVELY IDENTIFIED TO HAVE COME FROM THE CRE RX BALLOON FROM THE PREVIOUS PROCEDURE. THE DETACHED PART WAS SUCCESSFULLY RETRIEVED WITH A GRASPER. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT AND THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023038 ↗
  • MalfunctionJan 22, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE RX BILIARY DILATATION BALLOON WAS USED DURING A DILATION OF THE PAPIL OF VATER PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, WHEN THE PACKAGE WAS OPENED, IT WAS NOTED THAT THE BLACK RX TUNNEL OF THE DEVICE WAS DETACHED FROM THE CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE RX BILIARY DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8267918 ↗

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates