CRE™ RX M00558940Recalls & reported problems
Boston Scientific Corporation · Biliary Balloon Dilatation Catheter · Class 2
All models of CRE RX (9)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08714729965497), received between Nov 19, 2018 and Feb 6, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Adverse Event Without Identified Device or Use Problem2
- Entrapment of Device2
- Break1
- Burst Container or Vessel1
- Difficult to Advance1
- Difficult to Remove1
- Inflation Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 6, 2026Entrapment of Device; Use of Device Problem; Detachment of Device or Device Component; Difficult to Advance
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE RX DILATATION BALLOON WAS ATTEMPTED TO BE USED IN THE STOMACH AND ESOPHAGUS DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED IN (B)(6) 2025. THE SPECIFIC DATE IS UNKNOWN. DURING THE PROCEDURE, A BILIARY BALLOON DILATATION CATHETER WAS INSERTED INTO AN ENDOSCOPE WITH A 2.8 MM WORKING CHANNEL, WHICH WAS TOO SMALL FOR THE DEVICE. THE CATHETER MET SIGNIFICANT RESISTANCE DURING INSERTION AND DID NOT EXIT THE DISTAL TIP OF THE SCOPE. WHEN THE TEAM PULLED THE DEVICE BACK OUT, THE BLACK RX SHEATH, WHICH IS NOT DESIGNED TO DETACH, WAS TORN OFF AND REMAINED IN THE BIOPSY CHANNEL. IT WAS REPORTED THAT NO PART OF THE DETACHED PIECE FELL INTO THE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE RX DILATATION BALLOON DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: THE INSTRUCTION FOR USE (IFU) STATES THAT THE CRE RX BILIARY BALLOON DILATATION CATHETER IS DESIGNED TO PASS THROUGH AN ENDOSCOPE WITH A MINIMUM 3.7MM WORKING CHANNEL. HOWEVER, THE COMPLAINANT REPORTED THAT THEY USED THE DEVICE IN A SCOPE WITH TOO SMALL OF A WORKING CHANNEL 2.8MM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24279791 ↗MalfunctionJan 6, 2026Burst Container or Vessel; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE RX DILATATION BALLOON WAS USED IN THE COMMON BILE DUCT DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE TO TREAT STONES ON (B)(6) 2025. DURING THE PROCEDURE, THE BALLOON POPPED UPON INFLATION DURING DILATION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD FAITH EFFORTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23985137 ↗MalfunctionFeb 6, 2025Inflation Problem; Material Puncture/Hole; Inflation Problem; Material Puncture/Hole
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE RX DILATATION BALLOON WAS ATTEMPTED TO BE USED DURING A PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE BALLOON DID NOT INFLATE. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE RX DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THIS EVENT HAS BEEN DEEMED A REPORTABLE EVENT BASED ON THE INVESTIGATION FINDING OF A BALLOON PINHOLE AND THE ANATOMY LOCATION AND PROCEDURE TYPE ARE UNKNOWN AND MAY HAVE OCCURRED IN THE ESOPHAGUS. PLEASE SEE BLOCK H11 FOR FULL INVESTIGATION DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21324004 ↗MalfunctionAug 9, 2023Material Twisted/Bent; Adverse Event Without Identified Device or Use Problem; Material Twisted/Bent; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE RX BILIARY DILATATION BALLOON WAS USED DURING A PROCEDURE PERFORMED ON (B)(6) 2023. IT WAS REPORTED THAT THERE WAS A KINK AT THE DISTAL PART OF THE BALLOON SHEATH. THE SHEATH MUST BE CLIPPED IN THE OPPOSITE DIRECTION TO INFLATE THE BALLOON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. NO FURTHER INFORMATION HAS BEEN OBTAINED DESPITE GOOD-FAITH EFFORTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17491139 ↗MalfunctionMay 13, 2022Entrapment of Device; Difficult to Remove; Detachment of Device or Device Component; Entrapment of Device; Difficult to Remove; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE RX BILIARY DILATATION BALLOON WAS USED IN THE STOMACH AND PANCREAS DURING A DILATATION OF HOTAXIOS PROCEDURE PERFORMED ON (B)(6) 2022. DURING THE PROCEDURE, IT WAS NOTICED THAT THE DARK PART OF THE BALLOON CATHETER, THE RX TUNNEL, STAYED INSIDE THE WORKING CHANNEL WHEN THE PHYSICIAN ATTEMPTED TO REMOVE THE BALLOON. THE DILATATION WAS COMPLETED AT THIS TIME. THE ENDOSCOPE WAS REMOVED AND THE NURSE TOOK OUT THE BLACK PART USING A CLEANING BRUSH AND THE PROCEDURE WAS COMPLETED WITH A THERAPEUTIC GASTROSCOPE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14400411 ↗MalfunctionDec 17, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A CRE RX BILIARY DILATATION BALLOON WAS USED IN THE AMPULLA OF VATER DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6), 2020. ACCORDING TO THE COMPLAINANT, WHEN THE DEVICE WAS REMOVED FROM THE PATIENT AFTER SUCCESSFUL DILATION, THE BLACK RX TUNNEL UNKNOWINGLY DETACHED FROM THE CATHETER INTO THE SCOPE. IT WAS ONLY DISCOVERED WHEN AN EXTRACTOR BALLOON WAS ADVANCED DOWN THE SCOPE AND PUSHED THE RX TUNNEL OUT OF THE TIP OF THE SCOPE, WHICH THEN APPEARED ON THE ENDOSCOPIC IMAGE. THE EXTRACTOR BALLOON WAS THEN WITHDRAWN FROM THE SCOPE, AND THE BLACK RX TUNNEL CAME OUT OF THE SCOPE WITH IT. THE DILATION PROCEDURE WAS CONTINUED AND COMPLETED WITH THE EXTRACTOR AT THIS TIME. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11030123 ↗MalfunctionNov 19, 2018Break; Break
NOTE: THIS REPORT PERTAINS TO ONE OF TWO DEVICES USED DURING THE SAME PROCEDURE. MANUFACTURER REPORT # 3005099803-2018-61371 PERTAINS TO THE FIRST CRE RX DILATATION BALLOON, AND MANUFACTURER REPORT # 3005099803-2018-61367 PERTAINS TO THE SECOND CRE RX DILATATION BALLOON. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT TWO CRE RX DILATATION BALLOON WERE USED IN THE STOMACH DURING AN ESOPHAGOGASTRODUODENOSCOPY (EGD) PROCEDURE PERFORMED ON (B)(6) 2018. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, IT WAS NOTED THAT THERE WAS DIFFICULTY IN ADVANCING THE DEVICE DOWN THE SCOPE. REPORTEDLY, THE DEVICE GOT STUCK IN THE WORKING CHANNEL AND THEY HAD TROUBLE GETTING IT OUT. IT WAS THEN THAT THE CATHETER SNAPPED. NONE OF THE CATHETER WAS INSIDE THE PATIENT. THE SCOPE WAS TAKEN OUT OF THE PATIENT, AND THE DEVICE WAS REMOVED FROM THE SCOPE. THE SAME ISSUE OCCURRED WITH THE SECOND CRE RX DILATATION BALLOON. THE PROCEDURE WAS COMPLETED WITH ANOTHER CRE RX DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8086549 ↗
Other CATHETER, BILIARY, DIAGNOSTIC devices
- Hurricane™ RX M00545900Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls1 death705 reports
- MUSTANG™ H74939171060470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC384 reports
- CRE™ PRO M00558710Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC1 recall284 reports
- Hurricane™ RX M00545920Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls211 reports
- MUSTANG™ H74939171070470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC265 reports
- MUSTANG™ H74939171050470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC212 reports
- Hurricane™ RX M00545940Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC2 recalls109 reports
- MUSTANG™ H74939171100470Boston Scientific Corporation · CATHETER, BILIARY, DIAGNOSTIC203 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page