Athletis™ H74939347060470Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Athletis (26)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 08714729973799), received between May 24, 2021 and Aug 12, 2026 — classified as 14 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture12
- Entrapment of Device2
- Leak/Splash2
Patient problems
- No Clinical Signs, Symptoms or Conditions14
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 12, 2026Material Rupture
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50225927 ↗MalfunctionAug 25, 2025Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 100% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED UPPER ARM SHUNT. A 6.0MM X 40MM X 75CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THIRD INFLATION AT 20 ATMOSPHERES FOR 10 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT INJURIES REPORTED, AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22882490 ↗MalfunctionApr 17, 2025Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT BALLOON BECAME STUCK WITH THE WIRE. THE 90% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN A MILDLY TORTUOUS AND MILDLY CALCIFIED VESSEL. A 6.0MM X 40MM X 75CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR SHUNT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY. HOWEVER, THE BALLOON BECAME STUCK WITH 150CM NON-BOSTON SCIENTIFIC GUIDEWIRE. THE DEVICE WAS REMOVED, BUT IT WAS CONSIDERABLY DIFFICULT. BOTH THE BALLOON AND WIRE WERE REMOVED TOGETHER. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21861809 ↗MalfunctionNov 13, 2024Leak/Splash; Material Rupture; Leak/Splash; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE ARTERIOVENOUS FISTULA OF BRACHIAL ARTERY. A 6.0MM X 40MM X 75CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING INFLATION AT 10 ATMOSPHERES, THE BALLOON BURST AND LEAKED CONTRAST. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS FINE POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20678939 ↗MalfunctionSep 16, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MODERATELY CALCIFIED COMMON FEMORAL ARTERY (CFA). A 6.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 6 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20224644 ↗MalfunctionApr 26, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED FOREARM VESSEL. A 6.0MM X 40MM X 75CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON THIRD INFLATION AT 40 ATMOSPHERES. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19197597 ↗MalfunctionApr 4, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED BRACHIAL VEIN. A 6.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON TWELVE INFLATIONS AT 30 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19047217 ↗MalfunctionMar 26, 2024Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT THE BALLOON BECAME STUCK WITH WIRE. THE 90% STENOSED TARGET LESIONS WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL IN THE FOREARM. A 6.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON BECAME STUCK WITH A 0.035 NON-BOSTON SCIENTIFIC GUIDEWIRE. THE GUIDEWIRE WAS REMOVED FROM THE CATHETER OUTSIDE THE BODY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18981986 ↗MalfunctionFeb 26, 2024Leak/Splash; Material Rupture; Leak/Splash; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE STENOSED TARGET LESION WAS LOCATED IN THE ARTERIOVENOUS FISTULA. A 6.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. HOWEVER, WHILE DILATING THE BALLOON USING THE INFLATOR, THE CENTER PART OF THE BALLOON BURST, AND THE CONTRAST AGENT LEAKED OUT WITHOUT DILATATION. AFTER SEVERAL ATTEMPTS TO DILATE, THE CONTRAST AGENT CONTINUED TO LEAK OUT AND THE BALLOON WAS NOT DILATED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18779996 ↗MalfunctionAug 17, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND NON-CALCIFIED VESSEL FOREARM. A 6.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION AT 40 ATMOSPHERES, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM. THE PROCEDURE WAS COMPLETED WITH THE ORIGINAL DEVICE. THERE WERE NO PATIENT COMPLICATIONS NOR INJURIES REPORTED, AND THE PATIENT CONDITION WAS GOOD POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17568531 ↗MalfunctionApr 13, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE RADIAL ARTERIOVENOUS FISTULA. A 6.0MM X 40MM X 75CM ATHLETIS BALLOON DILATATION CATHETER WAS ADVANCED. HOWEVER, DURING THE PROCEDURE, THE BALLOON BURST INSIDE PATIENT. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT CONDITION WAS STABLE. IT WAS FURTHER REPORTED THAT THE TARGET LESION WAS 100% STENOSED AND WAS MILDLY TORTUOUS AND SEVERELY CALCIFIED. THE BALLOON WAS INFLATED TWICE BEFORE IT RUPTURED AND NO FRAGMENTS WERE LEFT INSIDE THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16739881 ↗MalfunctionNov 4, 2022Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN A MILDLY TORTUOUS AND MILDLY CALCIFIED CEPHALIC VEIN. A 6.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS SELECTED FOR THE VASCULAR ACCESS INTERVENTIONAL THERAPY CASE. HOWEVER, DURING THE FIRST INFLATION AT 20 ATMOSPHERES FOR 10 SECONDS DURATION TIME, THE BALLOON RUPTURED. THE DEVICE WAS ABLE TO BE REMOVED USING THE NORMAL METHOD WITHOUT ANY ISSUES. ANOTHER ATHLETIS BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15729833 ↗MalfunctionDec 20, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. A 6.0MM X 40MM X 75CM ATHLETIS BALLOON WAS SELECTED FOR A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY IN A SHUNT. PATIENT ANATOMY CONDITIONS REVEALED 99% STENOSIS, MODERATE TORTUOSITY AND SEVERE CALCIFICATION. THE BALLOON WAS INFLATED THREE TIMES BELOW 32 ATMOSPHERES (ATMS) FOR 30 SECONDS. ON THE THIRD INFLATION, THE BALLOON RUPTURED WHEN INFLATED TO 32ATMS. THE BALLOON WAS COMPLETELY REMOVED USING THE NORMAL METHOD. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13042442 ↗MalfunctionMay 24, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT THE BALLOON BURST. THE TARGET LESION WAS LOCATED IN THE SUPERFICIAL FEMORAL ARTERY (SFA). STENOSIS WAS ESTIMATED AT 50% TARGET LESION WAS 80% CALCIFIED WITH NO TORTUOSITY. A 6.0MM X 40MM X 75CM ATHLETIS BALLOON WAS SELECTED FOR USE IN A SFA ANGIOPLASTY PROCEDURE TO POST-DILATE A STENT. AFTER THE STENT WAS IMPLANTED, THE ATHLETIS BALLOON WAS INFLATED ONCE TO A PRESSURE BETWEEN 10-12 ATMOSPHERES. HOWEVER, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH RECANALIZATION. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11873371 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page