Skip to content

Athletis™ H74939347061050Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Athletis (26)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 08714729973843), received between Sep 7, 2023 and Aug 22, 2024 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture10
  • Entrapment of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 22, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SHUNT VEIN. A 6.0MM X 100MM X 50CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, ON THE FIRST INFLATION AT 30 ATMOSPHERES FOR 40 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM AND WAS REPLACED FOR ANOTHER OF THE SAME MODEL TO COMPLETE THE PROCEDURE. NO COMPLICATIONS REPORTED, AND PATIENT STATUS WAS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20040110 ↗
  • MalfunctionJul 9, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19704061 ↗
  • MalfunctionMay 17, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 35 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19346480 ↗
  • MalfunctionMar 22, 2024Entrapment of Device; Entrapment of Device

    IT WAS REPORTED THAT DEVICE ENTRAPMENT OCCURRED. THE 95% STENOSED TARGET LESION WAS MODERATELY TORTUOUS AND SEVERELY CALCIFIED. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR A SHUNT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE. DURING TREATMENT, THE DEVICE WAS JOINT TOGETHER WITH A .035 GUIDEWIRE. BOTH DEVICES WERE REMOVED AS A SINGLE UNIT. NO DAMAGE WAS NOTED ON THE BALLOON AND WAS REPLACED WITH ANOTHER OF THE SAME MODEL TO COMPLETE THE PROCEDURE. NO COMPLICATIONS WERE REPORTED, AND PATIENT STATUS WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18959705 ↗
  • MalfunctionMar 21, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED VESSEL. A 6.0MM X 100MM X 50CM ATHLETIS WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 20 ATMOSPHERES FOR 20 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18952387 ↗
  • MalfunctionFeb 19, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS MODERATELY TORTUOUS AND SEVERELY. THE PATIENT UNDERWENT A SHUNT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY WHERE A 6.0MM X 100MM X 50CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 30 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18731425 ↗
  • MalfunctionJan 17, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 35 ATMOSPHERES FOR 60 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18527502 ↗
  • MalfunctionJan 8, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN A MODERATELY TORTUOUS AND SEVERELY CALCIFIED VEIN. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED. HOWEVER, DURING THE FIRST INFLATION AT 34 ATMOSPHERES FOR 40 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18466090 ↗
  • MalfunctionDec 11, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN A MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 30 ATMOSPHERES FOR 40 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18307185 ↗
  • MalfunctionOct 10, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN A MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL. A 6.0MM X 100MM X 50CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE INITIAL INFLATION AT 10 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17903142 ↗
  • MalfunctionSep 7, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SHUNT VEIN. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATION CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE INITIAL INFLATION AT 30 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT INJURIES REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17702999 ↗

Other Catheter, angioplasty, peripheral, transluminal devices