Athletis™ H74939347061050Recalls & reported problems
Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2
All models of Athletis (26)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 08714729973843), received between Sep 7, 2023 and Aug 22, 2024 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture10
- Entrapment of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 22, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SHUNT VEIN. A 6.0MM X 100MM X 50CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, ON THE FIRST INFLATION AT 30 ATMOSPHERES FOR 40 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM AND WAS REPLACED FOR ANOTHER OF THE SAME MODEL TO COMPLETE THE PROCEDURE. NO COMPLICATIONS REPORTED, AND PATIENT STATUS WAS GOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20040110 ↗MalfunctionJul 9, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD WITHOUT ISSUE AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19704061 ↗MalfunctionMay 17, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 35 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19346480 ↗MalfunctionMar 22, 2024Entrapment of Device; Entrapment of Device
IT WAS REPORTED THAT DEVICE ENTRAPMENT OCCURRED. THE 95% STENOSED TARGET LESION WAS MODERATELY TORTUOUS AND SEVERELY CALCIFIED. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR A SHUNT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE. DURING TREATMENT, THE DEVICE WAS JOINT TOGETHER WITH A .035 GUIDEWIRE. BOTH DEVICES WERE REMOVED AS A SINGLE UNIT. NO DAMAGE WAS NOTED ON THE BALLOON AND WAS REPLACED WITH ANOTHER OF THE SAME MODEL TO COMPLETE THE PROCEDURE. NO COMPLICATIONS WERE REPORTED, AND PATIENT STATUS WAS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18959705 ↗MalfunctionMar 21, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED VESSEL. A 6.0MM X 100MM X 50CM ATHLETIS WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 20 ATMOSPHERES FOR 20 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18952387 ↗MalfunctionFeb 19, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS MODERATELY TORTUOUS AND SEVERELY. THE PATIENT UNDERWENT A SHUNT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY WHERE A 6.0MM X 100MM X 50CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE FIRST INFLATION AT 30 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18731425 ↗MalfunctionJan 17, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE FIRST INFLATION AT 35 ATMOSPHERES FOR 60 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18527502 ↗MalfunctionJan 8, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN A MODERATELY TORTUOUS AND SEVERELY CALCIFIED VEIN. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED. HOWEVER, DURING THE FIRST INFLATION AT 34 ATMOSPHERES FOR 40 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18466090 ↗MalfunctionDec 11, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN A MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON FIRST INFLATION AT 30 ATMOSPHERES FOR 40 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS IN GOOD CONDITION AFTER THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18307185 ↗MalfunctionOct 10, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 95% STENOSED TARGET LESION WAS LOCATED IN A SHUNT IN A MODERATELY TORTUOUS AND SEVERELY CALCIFIED VESSEL. A 6.0MM X 100MM X 50CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THE INITIAL INFLATION AT 10 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. THERE WERE NO COMPLICATIONS REPORTED AND THE PATIENT WAS IN GOOD CONDITION POST-PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17903142 ↗MalfunctionSep 7, 2023Material Rupture; Material Rupture
IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED SHUNT VEIN. A 6.0MM X 100MM X 50CM ATHLETIS PTA BALLOON DILATION CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING THE INITIAL INFLATION AT 30 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT INJURIES REPORTED, AND THE PATIENT WAS IN GOOD CONDITION POST PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17702999 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page