Skip to content

Athletis™ H74939347100470Recalls & reported problems

Boston Scientific Corporation · PTA Balloon Dilatation Catheter · Class 2

All models of Athletis (26)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 08714729974215), received between May 19, 2021 and Mar 9, 2026 — classified as 16 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture12
  • Inflation Problem3
  • Leak/Splash2
  • Device-Device Incompatibility1
  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 9, 2026Fluid/Blood Leak; Inflation Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 11-FEB-2026. IT WAS REPORTED THAT BALLOON FAILED TO INFLATE AND A LEAK OCCURRED. A 0.0MM X 40MM X 75CM ATHLETIS BALLOON WAS SELECTED FOR USE IN A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) PROCEDURE. THE TARGET LESION WAS LOCATED IN THE SUBCLAVIAN VEIN, AND WAS REPORTED TO BE 80% STENOSED WITH SEVERE TORTUOSITY AND MODERATE CALCIFICATION. DURING THE PROCEDURE, IT WAS NOTED THAT THE BALLOON DID NOT INFLATE. UPON WITHDRAWAL, A CONTRAST AGENT LEAK FROM THE BALLOON WAS OBSERVED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED A BALLOON PINHOLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24551848 ↗
  • MalfunctionSep 12, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE LEFT CENTRAL VEIN. A 10.0MM X 40MM X 75CM ATHLETICS BALLOON CATHETER WAS SELECTED FOR USE IN A LEFT CENTRAL VENOUS CATHETERIZATION. DURING THE INFLATION, THE BALLOON RUPTURED JUST BEFORE REACHING THE NEEDED PRESSURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE VESSEL WAS TORTUOUS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23038062 ↗
  • MalfunctionJul 18, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT A BALLOON RUPTURE OCCURRED. THE 80-90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND SEVERELY CALCIFIED ARTERIOVENOUS FISTULA. A 10.0MM X 40MM X 75CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON RUPTURED UPON THIRD INFLATION AT 14 ATMOSPHERES FOR 30 SECONDS. THE DEVICE WAS REMOVED WITHOUT ANY PROBLEM, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22543777 ↗
  • MalfunctionMay 20, 2025Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2025. IT WAS REPORTED THAT THE BALLOON WAS LEAKING CONTRAST AND DID NOT INFLATE. THE STENOSED TARGET LESION WAS LOCATED IN THE ARTERIOVENOUS FISTULA. A 10.0MM X 40MM X 75CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, WHILE DILATING THE BALLOON WITH THE INFLATOR, IT WAS CONFIRMED THAT THE CONTRAST AGENT WAS LEAKING. AS A RESULT, THE BALLOON DID NOT EXPAND AND WAS DEEMED DEFECTIVE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT. HOWEVER, DEVICE ANALYSIS REVEALED A BALLOON PINHOLE LOCATED APPROXIMATELY 3MM DISTAL OF THE PROXIMAL MARKERBAND.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22056624 ↗
  • MalfunctionApr 11, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 10.0MM X 40MM X 75CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATION. DURING THE PROCEDURE, THE BALLOON BURST AT 30 ATMOSPHERES. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21831140 ↗
  • MalfunctionDec 18, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. A 10.0MM X 40MM X 75CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATION. HOWEVER, THE BALLOON BURST JUST BEFORE REACHING PRESSURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20970698 ↗
  • MalfunctionNov 4, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN THE RADIAL ARTERIOVENOUS FISTULA. A 10.0MM X 40MM X 75CM ATHLETIS PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATATION. DURING INFLATION, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS STABLE POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20600017 ↗
  • MalfunctionSep 25, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE BALLOON RUPTURE OCCURRED. THE 70% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND NON CALCIFIED SUBCLAVIAN VEIN. A 10.0MM X 40MM X 75CM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATATION. DURING THE PROCEDURE, THE DEVICE WAS USED WITHIN THE STANDARD AIR PRESSURE RANGE, YET THE BALLOON STILL BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20300044 ↗
  • MalfunctionMay 17, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURED. THE 60% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND MILDLY CALCIFIED LEFT CENTRAL VEIN. A 10.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING THIRD INFLATION AT 14 ATMOSPHERES FOR 1 MINUTE, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED USING NORMAL METHOD AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS GOOD POST PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19343627 ↗
  • MalfunctionApr 11, 2024Inflation Problem; Inflation Problem

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(6) 2024. IT WAS REPORTED THAT BALLOON INFLATION ISSUE OCCURRED. A 10.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. BUT AS SOON AS THE INFLATION STARTED, THE PRESSURE WENT AWAY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED BALLOON PINHOLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19093210 ↗
  • MalfunctionOct 17, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE PATIENT UNDERWENT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA). THE 70% STENOSED TARGET LESION WAS LOCATED IN THE MILDLY TORTUOUS AND SEVERELY CALCIFIED RIGHT CENTRAL ARTERY. A 10.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED; HOWEVER, DURING THE SECOND INFLATION AT 20 ATMOSPHERES FOR 30 SECONDS, THE BALLOON RUPTURED. THE DEVICE WAS REMOVED, AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT CONDITION WAS GOOD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17952544 ↗
  • MalfunctionAug 31, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. DURING THE PROCEDURE, A 10 MM X 40 MM ATHLETIS BALLOON CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, BEFORE REACHING THE RATED BURST PRESSURE, THE BALLOON BURST. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THERE WERE NO FURTHER INFORMATION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17665232 ↗
  • MalfunctionJul 11, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MODERATELY CALCIFIED RADIAL ARTERIOVENOUS FISTULA. A 10.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATATION. HOWEVER, DURING INFLATION, THE BALLOON BURST. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED, AND THE PATIENT WAS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17302192 ↗
  • MalfunctionJun 2, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT BALLOON RUPTURE OCCURRED. THE TARGET LESION WAS LOCATED IN RADIAL ARTERY. A 10.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS ADVANCED FOR DILATION. HOWEVER, DURING INFLATION AT 25 ATMOSPHERES, THE BALLOON RUPTURED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17054359 ↗
  • MalfunctionApr 4, 2023Leak/Splash; Leak/Splash

    REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 14-MAR-2023. IT WAS REPORTED THAT BALLOON LEAK OCCURRED. THE PATIENT UNDERWENT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY . THE 70% STENOSED TARGET LESION WAS LOCATED IN A MODERATELY CALCIFIED AND MODERATELY TORTUOUS ARTERIOVENOUS FISTULA. A 10.0MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS FOR DILATATION. HOWEVER, WHEN THE PHYSICIAN TRIED TO INFLATE THE BALLOON, A BALLOON LEAK WAS FOUND UNDER ANGIOGRAPHY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS WAS FINE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED BALLOON PINHOLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16677049 ↗
  • MalfunctionMay 19, 2021Material Rupture; Device-Device Incompatibility; Material Rupture; Device-Device Incompatibility

    IT WAS REPORTED THAT THE BALLOON BURST INSIDE A STENT. A 10MM X 40MM X 75CM ATHLETIS PTA BALLOON DILATATION CATHETER WAS SELECTED FOR USE IN AN ANGIOPLASTY PROCEDURE TO TREAT AN OCCLUSION IN THE SUPERFICIAL FEMORAL ARTERY (SFA). DURING THE PROCEDURE, THE BALLOON BURST WHILE DILATING A STENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED. IT WAS FURTHER REPORTED THAT THE ATHLETIS BALLOON WAS INFLATED 3 TIMES FOR 50 SECOND DURATIONS AT 40ATM INSIDE A NON-BOSTON SCIENTIFIC TRANSJUGULAR INTRAHEPATIC PORTOSYSTEMIC SHUNT (TIPS) STENT. THE DEVICE WAS SUCCESSFULLY REMOVED FROM THE PATIENT, AND A MUSTANG BALLOON WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11850048 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page