LABSYSTEM™ PRO M00420020810Recalls & reported problems
Boston Scientific Corporation · LS10K WIN10 COMPUTER · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729996934), received between May 31, 2022 and May 31, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Communication or Transmission Problem2
- Use of Device Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 31, 2022Use of Device Problem; Communication or Transmission Problem; Use of Device Problem; Communication or Transmission Problem
IT WAS REPORTED THAT DURING A PROCEDURE USING THE LS PRO RECORDING SYSTEM THE WAS NOT CORRECTLY MEASURING THE BLOOD PRESSURE OF THE PATIENT. AT THE START OF THE PROCEDURES THE BLOOD PRESSURE WAS NORMAL AND THEN DURING THE PROCEDURE THE BLOOD PRESSURE WOULD BECOME UNDERESTIMATED. WHEN THEY CHECKED THE CALIBRATED ZERO WAS NOT CORRECT AND WAS LOWER THAN EXPECTED AND NEEDED TO BE RESET. ONCE THE RESET HAS BEEN PERFORMED THE PROCEDURES WERE COMPLETED SUCCESSFULLY. DURING THESE PROCEDURES THERE WERE NO OTHER SYSTEMS IN USE TO MONITOR THE VITAL SIGNS OF THE PATIENT. THE RECORDING SYSTEM IS NOT EXPECTED TO BE RETURNED AS IT IS GOING TO BE REPAIRED ON SITE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14553560 ↗MalfunctionMay 31, 2022Use of Device Problem; Communication or Transmission Problem; Use of Device Problem; Communication or Transmission Problem
IT WAS REPORTED THAT DURING A PROCEDURE USING THE LS PRO RECORDING SYSTEM THE WAS NOT CORRECTLY MEASURING THE BLOOD PRESSURE OF THE PATIENT. AT THE START OF THE PROCEDURES THE BLOOD PRESSURE WAS NORMAL AND THEN DURING THE PROCEDURE THE BLOOD PRESSURE WOULD BECOME UNDERESTIMATED. WHEN THEY CHECKED THE CALIBRATED ZERO WAS NOT CORRECT AND WAS LOWER THAN EXPECTED AND NEEDED TO BE RESET. ONCE THE RESET HAS BEEN PERFORMED THE PROCEDURES WERE COMPLETED SUCCESSFULLY. DURING THESE PROCEDURES THERE WERE NO OTHER SYSTEMS IN USE TO MONITOR THE VITAL SIGNS OF THE PATIENT. THE RECORDING SYSTEM IS NOT EXPECTED TO BE RETURNED AS IT IS GOING TO BE REPAIRED ON SITE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14552799 ↗
Other Computer, diagnostic, programmable devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.