Skip to content

Armada B1140-060Recalls & reported problems

Abbott Vascular Inc. · Armada 35 PTA Catheter 14 mm x 60 mm x 80 cm / Over-The-Wire · Class 2

All models of Armada (213)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08717648155031), received between Jul 6, 2018 and Nov 11, 2023 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture6
  • Difficult to Remove1
  • Material Separation1
  • Off-Label Use1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient2
  • Foreign Body In Patient1
  • Tachycardia1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 11, 2023Off-Label Use; Difficult to Remove; Material Rupture; Material Separation; Off-Label Use; Difficult to Remove; Material Rupture; Material Separation

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE LEFT ILIAC VEIN. THE 14X60 MM ARMADA PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER WAS INFLATED THREE TIMES TO 11 ATMOSPHERES BUT THE BALLOON RUPTURED. THE DEVICE WAS FULLY DEFLATED UPON REMOVAL; HOWEVER, THE DEVICE BECAME STUCK ON THE END OF THE INTRODUCER SHEATH IN THE RIGHT INTERNAL JUGULAR VEIN DURING REMOVAL. A KNIFE WAS USED TO CUT THE FLUSH PORT OF THE DEVICE AND THEN THE ACCESS SHEATH WAS REMOVED. A LARGER SHEATH WAS THEN ATTEMPTED AND THE DISTAL MARKER OF THE BALLOON WAS ABLE TO BE CAPTURED. WHEN THE DEVICE WAS RETRIEVED, IT WAS INSPECTED AND IT WAS FOUND THAT SOME OF THE BALLOON MATERIAL HAD SEPARATED, ABOUT 1-5 MM OF MATERIAL. AN ANGIOGRAM WAS PERFORMED AND A CAVA-GRAM. INTRAVASCULAR ULTRASOUND (IVUS) WAS ALSO PERFORMED AND A FOCUSED SCAN OF THE NECK, AND NO DEBRIS OR FOREIGN BODY WAS SEEN; HOWEVER, A SMALL FRAGMENT LIKELY REMAINS IN THE PATIENT. A BEDSIDE ECHOCARDIOGRAM WAS PERFORMED AS THERE WAS SINUS TACHYCARDIA WHICH OCCURRED RELATED TO GUIDE WIRE AND CATHETER MANIPULATION. THE PATIENT WAS HELD FOR OBSERVATION BUT THEY ARE WELL AND ASYMPTOMATIC. SUBSEQUENT TO THE INITIALLY FILED REPORT IT WAS REPORTED THAT TH

    Patient problems: Foreign Body In Patient; Tachycardia

    MAUDE report 18117103 ↗
  • MalfunctionSep 28, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT TARGET LESIONS IN THE RIGHT AND LEFT ILIAC VEINS. THE FIRST ARMADA DILATATION CATHETER WAS ADVANCED INTO THE RIGHT ILIAC VEIN WITHOUT DIFFICULTY AND INFLATED THREE TIMES. ON THE THIRD INFLATION, THE BALLOON WAS INFLATED TO 4 ATMOSPHERES (ATM) AND RUPTURED. THE DEVICE WAS REMOVED WITHOUT DIFFICULTY AND THERE WAS NO ADVERSE PATIENT EFFECT. DILATATION WAS COMPLETED USING A 16 X 60 MM NON-ABBOTT DILATATION CATHETER. THE SECOND ARMADA DILATATION CATHETER WAS ADVANCED INTO THE LEFT ILIAC VEIN WITHOUT DIFFICULTY AND INFLATED ONE TIME, TO 2 ATM; HOWEVER, A BALLOON RUPTURE OCCURRED. THERE WAS NO DIFFICULTY REMOVING THE DEVICE AND NO ADVERSE PATIENT EFFECT. DILATATION WAS COMPLETED USING THE 16 X 60 MM NON-ABBOTT DILATATION CATHETER. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12544641 ↗
  • MalfunctionSep 28, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT TARGET LESIONS IN THE RIGHT AND LEFT ILIAC VEINS. THE FIRST ARMADA DILATATION CATHETER WAS ADVANCED INTO THE RIGHT ILIAC VEIN WITHOUT DIFFICULTY AND INFLATED THREE TIMES. ON THE THIRD INFLATION, THE BALLOON WAS INFLATED TO 4 ATMOSPHERES (ATM) AND RUPTURED. THE DEVICE WAS REMOVED WITHOUT DIFFICULTY AND THERE WAS NO ADVERSE PATIENT EFFECT. DILATATION WAS COMPLETED USING A 16 X 60 MM NON-ABBOTT DILATATION CATHETER. THE SECOND ARMADA DILATATION CATHETER WAS ADVANCED INTO THE LEFT ILIAC VEIN WITHOUT DIFFICULTY AND INFLATED ONE TIME, TO 2 ATM; HOWEVER, A BALLOON RUPTURE OCCURRED. THERE WAS NO DIFFICULTY REMOVING THE DEVICE AND NO ADVERSE PATIENT EFFECT. DILATATION WAS COMPLETED USING THE 16 X 60 MM NON-ABBOTT DILATATION CATHETER. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12544640 ↗
  • MalfunctionSep 17, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THIS WAS A PROCEDURE TO TREAT A LESION IN THE LEFT COMMON ILIAC VEIN. AN ARMADA 35 14MM/60MM WAS INSERTED, BUT DURING THE FIRST INFLATION, THE BALLOON RUPTURED AT 4 ATMOSPHERES. THE DEVICE WAS REMOVED AND REPLACED WITH THE SAME SIDE ARMADA 35. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12492347 ↗
  • MalfunctionFeb 4, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE RIGHT COMMON ILIAC VEIN. A NON-ABBOTT STENT WAS IMPLANTED. THEN ONE 14 X 60 MM ARMADA 35 AND ONE 14 X 40 MM ARMADA 35 BALLOON DILATATION CATHETERS (BDC) WERE ADVANCED SIMULTANEOUSLY THROUGH TWO DIFFERENT ACCESS SITES FOR POST-DILATATION. HOWEVER, BOTH BALLOONS RUPTURED AT 6 ATMOSPHERES DURING THE FIRST INFLATION. BOTH ARMADA BDC DEVICES WERE SIMPLY REMOVED. THE PROCEDURE ENDED AT THIS POINT. THE PATIENT IS STABLE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9664277 ↗
  • MalfunctionJul 6, 2018Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS TO POST-DILATE A PREVIOUSLY IMPLANTED UNSPECIFIED STENT IN AN UNSPECIFIED VEIN WITH A 14 X 60 MM ARMADA 35 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER. THE PTA CATHETER WAS ADVANCED TO THE STENT, BUT WHEN INFLATED, THE BALLOON RUPTURED. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7667796 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page