Armada B2040-200Recalls & reported problems
Abbott Vascular Inc. · Armada 35 LL PTA Catheter 4 mm x 200 mm x 135 cm / Over-The-Wire · Class 2
All models of Armada (213)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08717648155147), received between Aug 19, 2020 and Mar 6, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture4
- Improper or Incorrect Procedure or Method2
- Leak/Splash2
- Difficult to Remove1
- Inflation Problem1
- Material Separation1
- Material Split, Cut or Torn1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 6, 2026Improper or Incorrect Procedure or Method
IT WAS REPORTED THE PROCEDURE WAS TO TREAT A MILDLY CALCIFIED AND TORTUOUS LESION IN THE SUPERFICIAL FEMORAL ARTERY. THE 4.0X200MM ARMADA 35 BALLOON CATHETER WAS USED HOWEVER IT WAS NOTED THE DEVICE WAS USED PAST THE EXPIRATION DATE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24531860 ↗MalfunctionFeb 24, 2026Inflation Problem; Leak/Splash; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MODERATELY CALCIFIED, MILDLY TORTUOUS LEFT SUPERFICIAL FEMORAL ARTERY LESION. DURING PREPARATION WHILE PERFORMING AIR ASPIRATION OF THE 4X200 MM ARMADA 35 BALLOON DILATATION CATHETER (BDC) AIR WAS NOTED. THE PHYSICIAN TURNED THE BALLOON NEGATIVE MULTIPLE TIMES TO SUCK IN THE AIR; HOWEVER CONTINUOUS AIR WAS NOTED. HOWEVER, THE PHYSICIAN DECIDED TO USE THE BALLOON ANYWAY. THE BDC WAS ADVANCED TO THE TARGET LESION FOR PRE-DILATION WITHOUT ISSUE. INFLATION WAS NOTED TO BE SLOW, BUT THE BALLOON WAS ABLE TO FULLY INFLATE. THE BDC WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24435916 ↗MalfunctionJun 28, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT 65% STENOSED LESION IN THE SUPERFICIAL FEMORAL ARTERY (SFA) WITH MODERATE CALCIFICATION AND MILD TORTUOSITY. THE 4X200MM ARMADA 35 LL BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION AND THE BALLOON WAS INFLATED; HOWEVER, A RUPTURE OCCURRED WHEN ATTEMPTING IN INFLATE THE BALLOON TO NOMINAL PRESSURE, 8 ATMOSPHERES (ATM). THEREFORE, THE BDC WAS REMOVED FROM THE PATIENT. ANOTHER ARMADA 35LL BALLOON WAS USED TO CONTINUE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19635826 ↗MalfunctionFeb 20, 2024Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN AN UNSPECIFIED ARTERY. THE 4 X 200 MM ARMADA 35 BALLOON DILATATION CATHETER (BDC) WAS ADVANCED WITHOUT ISSUE. DURING THE ATTEMPT TO INFLATE THE BALLOON, THE BALLOON DID NOT HOLD PRESSURE AND WAS LEAKING. AFTER REMOVAL, A HOLE TEAR WAS NOTED IN THE BALLOON. ANOTHER ARMADA WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18743712 ↗MalfunctionAug 16, 2022Difficult to Remove; Material Rupture; Material Separation; Difficult to Remove; Material Rupture; Material Separation
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE MINIMALLY CALCIFIED LEFT SUPERFICIAL FEMORAL ARTERY (SFA). A RIGHT FEMORAL ACCESS WAS OBTAINED AND ULTRASOUND WAS PERFORMED. THE ARMADA PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER WAS INFLATED TO NOMINAL PRESSURE AND RUPTURED AFTER 2 INFLATIONS TO 8 ATMOSPHERES (ATM). THERE WAS RESISTANCE WITH THE 6FR SHEATH DURING REMOVAL AND SOME OF THE BALLOON MATERIAL SEPARATED. THE MATERIAL WAS SEPARATED BUT STILL ATTACHED TO THE SHAFT SO THE WHOLE BALLOON WAS PULLED OUT THROUGH THE SHEATH. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. ANOTHER ARMADA WAS USED TO COMPLETE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15233084 ↗MalfunctionNov 24, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT AN 80% IN-STENT RESTENOSED LESION IN THE RIGHT SUPERFICIAL FEMORAL ARTERY. A 4.0X200MM ARMADA 35 BALLOON CATHETER WAS PREPARED PER THE INSTRUCTIONS FOR USE WITHOUT ANY ISSUES. THE BALLOON CATHETER WAS INFLATED ONCE AT 8 ATMOSPHERES; HOWEVER, THE BALLOON RUPTURED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NEW 4.0X200MM ARMADA 35 BALLOON CATHETER FOR BALLOON ANGIOPLASTY. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10891196 ↗MalfunctionAug 19, 2020Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MILDLY TORTUOUS, MODERATELY CALCIFIED DE NOVO SUPERIOR FEMORAL ARTERY (SFA) THAT IS 40% TO 60% STENOSED. A 4.0X200MM ARMADA 35 WAS NOTED TO RUPTURE AT FIRST INFLATION AT NOMINAL PRESSURE. ANOTHER ARMADA WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10426438 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page