Abbott Vascular Inc. · Armada 35 LL PTA Catheter 5 mm x 200 mm x 135 cm / Over-The-Wire · Class 2

All models of Armada (213)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 08717648155239), received between Nov 17, 2016 and May 19, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture4
  • Leak/Splash2
  • Contamination /Decontamination Problem1
  • Deformation Due to Compressive Stress1
  • Difficult to Remove1
  • Improper or Incorrect Procedure or Method1
  • Inflation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient2
  • No Patient Involvement1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2025Leak/Splash; Deformation Due to Compressive Stress; Leak/Splash; Deformation Due to Compressive Stress

    T WAS REPORTED THAT DURING PREPARATION OF THE 5X200 MM ARMADA 35 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER, THE BALLOON WAS NOTED TO HAVE A PINHOLE AND A SMALL KINK IN THE BALLOON. THE DEVICE WAS NOT USED AND THERE WAS NO PATIENT INVOLVEMENT. A NEW, SAME SIZE ARMADA WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22046897 ↗
  • MalfunctionJan 16, 2025Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT THE SUPERFICIAL FEMORAL ARTERY WITH MILD TORTUOSITY, MODERATE CALCIFICATION AND 80% STENOSIS. THE 5.0X200MM ARMADA 35 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY CATHETER WAS PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND ADVANCED TO THE LESION. THE BALLOON WAS INFLATED ONCE TO 5 ATMOSPHERES YET WAS UNABLE TO HOLD PRESSURE THUS UNABLE TO INFLATE. A LEAK WAS NOTED AT THE CONNECTION BETWEEN THE DEVICE AND THE PUMP. THEREFORE, ANOTHER SAME SIZE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21170715 ↗
  • MalfunctionNov 1, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE SUPERFICIAL FEMORAL ARTERY (SFA) WITH 95 STENOSIS, HEAVY CALCIFICATION AND MODERATE TORTUOSITY. THE 5.0X200MM ARMADA 35 BALLOON DILATATION CATHETER (BDC) WAS PREPARED PER INSTRUCTIONS FOR USE (IFU) AND ADVANCED WITHOUT RESISTANCE, HOWEVER, RUPTURED AT NOMINAL PRESSURE DURING THE FIRST INFLATION. A 5.0X100MM ARMADA 35 BDC WAS USED AND A SUPERA SELF-EXPANDING STENT (SES) WAS DEPLOYED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15709030 ↗
  • MalfunctionSep 27, 2021Difficult to Remove; Material Rupture; Improper or Incorrect Procedure or Method; Difficult to Remove; Material Rupture; Improper or Incorrect Procedure or Method

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A DE NOVO LESION IN THE HEAVILY CALCIFIED, MODERATELY TORTUOUS, 80% STENOSED LEFT SUPERFICIAL FEMORAL ARTERY. A NON-ABBOTT 6F SHEATH WAS ADVANCED WITH A 5X200MM ARMADA 35 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER, AND THE BALLOON RUPTURED DURING THE FIRST INFLATION AT NOMINAL PRESSURE. THE PTA CATHETER AND SHEATH FACED RESISTANCE WITH EACH OTHER DURING REMOVAL, SO FORCE WAS APPLIED. AN UNSPECIFIED ARMADA 35 BALLOON WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12533887 ↗
  • MalfunctionFeb 19, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT THREE ARMADA 35 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETERS (5X200MM, 5X20MM, 6X200MM) AND TWO ABSOLUTE PRO SELF-EXPANDING STENT SYSTEMS (SESS) (6X150MM, 6X60MM) WERE RECEIVED BY THE CUSTOMER WITH BLOOD MARKS ON THE OUTSIDE PACKAGING. THE DEVICES WERE NOT USED AND THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Patient Involvement

    MAUDE report 9727218 ↗
  • MalfunctionMay 18, 2017Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN THE LEFT SUPERFICIAL FEMORAL ARTERY THAT WAS MILDLY TORTUOUS, HIGHLY CALCIFIED AND 60% STENOSED. THE ARMADA 35 BALLOON BURST AFTER THE FIRST INFLATION BELOW RATED BURST PRESSURE. ANOTHER ARMADA 35 WAS USED TO COMPLETE THE PROCEDURE. THE DEVICE WAS PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE WITH NO ISSUES OR LEAKS NOTED. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6575589 ↗
  • MalfunctionNov 17, 2016Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED SUPERFICIAL FEMORAL ARTERY. A 5 X 200 MM ARMADA 35 BALLOON CATHETER WAS ADVANCED TO THE LESION FOR PRE-DILATATION: HOWEVER, THE BALLOON RUPTURED DURING THE FIRST INFLATION BELOW THE RATED BURST PRESSURE. IT IS UNKNOWN EXACTLY WHAT PRESSURE. ANOTHER ARMADA BALLOON CATHETER WAS USED FOR PRE-DILATATION AND AN UNKNOWN STENT WAS SUCCESSFULLY DEPLOYED TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6110472 ↗