Armada B2120-080Recalls & reported problems
Abbott Vascular Inc. · Armada 35 PTA Catheter 12 mm x 80 mm x 135 cm / Over-The-Wire · Class 2
All models of Armada (213)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08717648155574), received between Feb 28, 2018 and Jan 13, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture2
- Detachment Of Device Component1
- Inflation Problem1
- Leak/Splash1
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 13, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED DURING THE PROCEDURAL PREPARATION OF A 12X80MM ARMADA 35 BALLOON DILATATION CATHETER (BDC) A LEAK WAS NOTED WHEN THE DEVICE WAS PURGED WITH SALINE. A NEW SAME SIZED ARMADA 35 BDC WAS USED AND THE PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT INVOLVEMENT NOR CLINICAL DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21144183 ↗MalfunctionDec 16, 2020Inflation Problem; Material Rupture; Inflation Problem; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY TORTUOUS, MODERATELY CALCIFIED ILIAC ARTERY THAT WAS 70% STENOSED. A 12X80MM ARMADA 35 BALLOON WAS ADVANCED TO THE LESION AND WHEN PRESSURIZED AT 8 ATMOSPHERES THE ARMADA WAS NOTED AS HAVING FAILED TO INFLATE AND RUPTURED. ANOTHER DEVICE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11019884 ↗MalfunctionFeb 28, 2018Detachment Of Device Component; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE ILIAC ARTERY. A 12 X 80MM ARMADA 35 BALLOON DILATATION CATHETER (BDC) INFLATED ONE TIME AT 10 ATMOSPHERES AND RUPTURED. NO REPORTED RESISTANCE DURING ADVANCEMENT OR REMOVAL. WHEN DEVICE WAS REMOVED THE SHAFT OF THE BALLOON APPEARED SEPARATED SINCE THE BALLOON WAS LOOPED ON THE GUIDE WIRE. THE DISTAL SHAFT OF THE BDC SEPARATED INTO TWO PIECES. THE BALLOON WAS ALSO NOTED TO BE SEPARATED FROM THE CATHETER AND THE BALLOON MARKERS WERE ALSO DETACHED FROM THE INNER MEMBER. THE DEVICE WAS SIMPLY WITHDRAWN, AS THE ENTIRE DEVICE CAME OUT WITH THE WIRE. THE PROCEDURE CONTINUED. ANGIOJET THROMBECTOMY WAS PERFORMED, WHICH WAS UNRELATED TO THE USE OF THE DEVICE. ANOTHER ARMADA 35 WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7304708 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page