Abbott Vascular Inc. · Armada 35 PTA Catheter 14 mm x 60 mm x 135 cm / Over-The-Wire · Class 2

All models of Armada (213)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717648155604), received between Oct 15, 2019 and Oct 20, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Rupture2
  • Difficult to Advance1
  • Inflation Problem1
  • Material Separation1

Patient problems

  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 20, 2019Inflation Problem; Material Rupture; Difficult to Advance; Inflation Problem; Material Rupture; Difficult to Advance

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A DE NOVO LESION WITH 50% STENOSIS LOCATED IN A MODERATELY TORTUOUS, MODERATELY CALCIFIED RIGHT COMMON ILLIAC ARTERY. A 14X60MM ARMADA 35 BALLOON DILATATION CATHETER WAS ADVANCED WITH RESISTANCE MET WITH THE ANATOMY BUT ULTIMATELY REACHED THE LESION AND WAS INFLATED. BEFORE THE BALLOON REACHED 4 ATMOSPHERES (ATM), AT ABOUT 3 ATM, THE BALLOON RUPTURED. INFLATION HAD BEEN NOTED AS SLOW. ANOTHER SAME SIZE ARMADA 35 WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9212550 ↗
  • MalfunctionOct 15, 2019Material Rupture; Material Separation; Material Rupture; Material Separation

    IT WAS REPORTED THAT THE ARMADA 35 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER RUPTURED AT 3 ATMOSPHERES (ATM). THE BALLOON ALSO SEPARATED FROM THE CATHETER. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9194776 ↗

Other Catheter, angioplasty, peripheral, transluminal devices