Skip to content

Armada A1025-040Recalls & reported problems

Abbott Vascular Inc. · Armada 14 PTA Catheter 2.5 mm x 40 mm x 90 cm / Over-The-Wire · Class 2

All models of Armada (213)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717648155734), received between May 1, 2024 and Jun 20, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem1
  • Difficult to Advance1
  • Difficult to Remove1
  • Material Too Soft/Flexible1
  • Material Twisted/Bent1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 20, 2025Stretched; Difficult to Advance; Material Too Soft/Flexible; Stretched; Difficult to Advance; Material Too Soft/Flexible

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE LEFT PERONEAL ARTERY BELOW THE KNEE WITH NO TORTUOSITY, MILD CALCIFICATION AND 65% STENOSIS. WHEN THE 2.5X40 MM ARMADA 14 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER WAS ADVANCED OVER THE UNSPECIFIED .014 GUIDE WIRE, THE BALLOON SHAFT DID NOT HAVE AS MUCH SUPPORT AS USUAL. UPON REMOVAL FROM THE ANATOMY IT WAS FOUND THAT THE SHAFT WAS COMPLETELY COMPRESSED AND KINKED. ANOTHER ARMADA 14 DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. RETURNED DEVICE ANALYSIS IDENTIFIED THAT THE INNER AND OUTER MEMBER WERE STRETCHED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22298595 ↗
  • MalfunctionMay 1, 2024Deflation Problem; Difficult to Remove; Material Twisted/Bent; Deflation Problem; Difficult to Remove; Material Twisted/Bent

    IT WAS REPORTED THAT THE PROCEDURE WAS TO PERFORM RECANALIZATION AND PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) OF THE LOWER LEFT LEG USING THE ANTEGRADE APPROACH. THREE ARMADA BALLOON DILATATION CATHETERS (BDC) WERE ATTEMPTED TO BE USED IN THE PROCEDURE; HOWEVER, TWO OF THE ARMADA BALLOONS FAILED TO DEFLATE PROPERLY AND BOTH WERE DIFFICULT TO REMOVE. AFTER REMOVAL, THE BALLOONS WERE NOTED TO BE TWISTED. IN THE PHYSICIAN OPINION THE MATERIAL OF THE PTA BALLOONS IS WEAKER CAUSING BALLOON TWISTING [TORSION]. A THIRD ARMADA BDC WAS ATTEMPTED TO BE USED IN THE PROCEDURE; HOWEVER, THE BALLOON WAS NOTED TO HAVE A LEAK AT THE HUB. ANOTHER ARMADA 14 BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19222987 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page