Armada A2015-120Recalls & reported problems
Abbott Vascular Inc. · Armada 14 PTA Catheter 1.5 mm x 120 mm x 150 cm / Over-The-Wire · Class 2
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 08717648155932), received between Oct 2, 2018 and May 20, 2024 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture6
- Difficult to Advance4
- Difficult to Remove2
- Improper or Incorrect Procedure or Method1
- Inflation Problem1
- Leak/Splash1
- Material Twisted/Bent1
Patient problems
- No Consequences Or Impact To Patient5
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 20, 2024Material Rupture; Difficult to Advance; Material Rupture; Difficult to Advance
IT WAS REPORTED THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED LEFT ANTERIOR TIBIAL ARTERY. DURING PRE-DILATATION, THE 1.5X20X150CM ARMADA 14 BALLOON WAS ADVANCED BUT THERE WAS NOTED RESISTANCE DURING ADVANCEMENT DUE TO CALCIUM, HOWEVER IT REACHED THE LESION. DURING THE FIRST INFLATION, THE BALLOON RUPTURED AT 6 ATMOSPHERES (ATM). A 1.5X120X150CM ARMADA 14 BALLOON WAS THEN USED BUT THERE WAS NOTED RESISTANCE DURING ADVANCEMENT DUE TO CALCIUM, HOWEVER IT REACHED THE LESION. DURING THE FIRST INFLATION, THE BALLOON RUPTURED AT 5ATM. A NON-ABBOTT DEVICE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19353762 ↗MalfunctionAug 31, 2022Material Rupture; Improper or Incorrect Procedure or Method; Material Rupture; Improper or Incorrect Procedure or Method
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MODERATELY CALCIFIED, MILDLY TORTUOUS, 30% STENOSED, DE NOVO LESION IN THE LEFT ANTERIOR TIBIAL ARTERY. A 1.5X120MM ARMADA 14 BALLOON DILATATION CATHETER (BDC) RUPTURED DURING THE FIRST INFLATION TO 24 ATMOSPHERES (ATM). IT IS SUSPECTED THAT THE RUPTURE WAS DUE TO THE CALCIFIED LESION. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. ANOTHER ARMADA BDC WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15331495 ↗MalfunctionDec 14, 2021Material Rupture; Material Rupture
IT WAS REPORTED THIS WAS A PROCEDURE TO TREAT A HEAVILY CALCIFIED AND MILDLY TORTUOUS LEFT TIBIAL ARTERY. AN ARMADA 14 DILATATION CATHETER WAS INSERTED AND ADVANCED TO THE TARGET LESION. HOWEVER, DURING INFLATION, IT RUPTURED AT 10 ATMOSPHERES. THE PHYSICIAN STATED THE BALLOON RUPTURED WAS CAUSED BY A SHARP PIECE OF CALCIFICATION. THE DEVICE WAS REMOVED AND REPLACED WITH A NON-ABBOTT DEVICE AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13003264 ↗MalfunctionFeb 25, 2021Difficult to Remove; Difficult to Advance; Difficult to Remove; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS TREAT AN ANTERIOR TIBIAL ARTERY (ATA). A 1.5X120MM ARMADA PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY CATHETER WAS ADVANCED OVER THE 14 COMMAND GUIDE WIRE TO CROSS A LESION; HOWEVER, THE GUIDE WIRE RETRACTED DUE TO THE CALCIFICATION AND BACK INTO THE BALLOON CATHETER. THE GUIDE WIRE WOULD NOT ADVANCE THROUGH THE DISTAL PORTION OF THE BALLOON CATHETER AND BECAME STUCK WITH THE BALLOON CATHETER. THE GUIDE WIRE AND CATHETER WERE REMOVED TOGETHER. ANOTHER UNSPECIFIED BALLOON AND COMMAND GUIDE WIRE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11379139 ↗MalfunctionFeb 4, 2020Difficult to Remove; Difficult to Advance; Material Twisted/Bent; Difficult to Remove; Difficult to Advance; Material Twisted/Bent
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MODERATELY TORTUOUS, MODERATELY CALCIFIED POPLITEAL ARTERY (BTK-BELOW THE KNEE). THE LESION WAS RECANALIZED SUCCESSFULLY WITH THE TIP OF THE COMMAND GUIDE WIRE (GW) IN THE ATERIA DORSALIS PEDIS. AN ARMADA 14 BALLOON WAS THEN USED TO TREAT THE LESION IN THE BTK AREA. DURING ADVANCING THE BALLOON, RESISTANCE WAS MET WITH THE COMMAND GW. IT WAS THEN ATTEMPTED TO STRETCH THE GW TO ASSIST IN ADVANCING THE BALLOON EASIER. THE BALLOON WAS THEN ABLE TO ADVANCE OVER THE GUIDE WIRE AND BE INFLATED AT THE LESION. AS THE GW WAS ATTEMPTED TO BE REMOVED, THEY NOTED THE TIP OF THE WIRE WAS NOT MOVING BECAUSE IT HAD BECOME SEPARATED. THEN DURING ATTEMPTED REMOVAL OF THE ARMADA, THE BALLOON CATHETER GOT TWISTED IN ITSELF AND IT WAS NOT ABLE TO BE REMOVED THROUGH THE 6F SHEATH. IT WAS NOTED THAT RESISTANCE WAS MET WITH THE GW AS WELL. IT WAS THEN DECIDED TO REMOVE THE GW, BALLOON AND SHEATH ALTOGETHER AS A SINGLE UNIT. THE PROCEDURE WAS DEEMED INCOMPLETE. THE NEXT DAY A RE-INTERVENTION WITH A SNARE DEVICE WAS DONE TO REMOVE THE SEPARATED TIP. IT WAS NOTED THAT THE TIP WAS KINKED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9664398 ↗MalfunctionDec 17, 2019Material Rupture; Difficult to Advance; Material Rupture; Difficult to Advance
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A 70% STENOSED AND TORTUOUS LESION IN THE DISTAL POSTERIOR TIBIAL ARTERY. RESISTANCE WAS FELT WHEN ADVANCING A 1.5X120MM ARMADA 14 BALLOON CATHETER AND THE BALLOON RUPTURED DURING THE FIRST INFLATION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NEW 1.5X120MM ARMADA 14 BALLOON CATHETER. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9477714 ↗MalfunctionJun 10, 2019Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A HEAVILY CALCIFIED LESION IN THE RIGHT ANTERIOR TIBIAL ARTERY. A 1.5X120MM ARMADA 14 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER WAS ADVANCED WITHOUT ISSUE; HOWEVER, THE BALLOON BURST DURING THE FIRST INFLATION TO 12 ATMOSPHERES. THE DEVICE WAS REMOVED WITHOUT REPORTED ISSUE. A SECOND 1.5X120MM ARMADA 14 PTA CATHETER WAS THEN ADVANCED WITHOUT ISSUE; HOWEVER, DURING THE INFLATION, A LEAK WAS NOTED WHERE THE HUB AND PROXIMAL SHAFT MEET. THE DEVICE WAS REMOVED WITHOUT REPORTED ISSUE, AND A NON-ABBOTT BALLOON DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS, AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8683597 ↗MalfunctionJun 10, 2019Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A HEAVILY CALCIFIED LESION IN THE RIGHT ANTERIOR TIBIAL ARTERY. A 1.5X120MM ARMADA 14 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER WAS ADVANCED WITHOUT ISSUE; HOWEVER, THE BALLOON BURST DURING THE FIRST INFLATION TO 12 ATMOSPHERES. THE DEVICE WAS REMOVED WITHOUT REPORTED ISSUE. A SECOND 1.5X120MM ARMADA 14 PTA CATHETER WAS THEN ADVANCED WITHOUT ISSUE; HOWEVER, DURING THE INFLATION, A LEAK WAS NOTED WHERE THE HUB AND PROXIMAL SHAFT MEET. THE DEVICE WAS REMOVED WITHOUT REPORTED ISSUE, AND A NON-ABBOTT BALLOON DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS, AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8683572 ↗MalfunctionOct 2, 2018Inflation Problem; Material Rupture; Inflation Problem; Material Rupture
IT WAS REPORTED THAT THE 1.5X120 MM ARMADA 14 DILATATION CATHETER WAS ADVANCED TO THE NON-TORTUOUS, UNCALCIFIED DISTAL ANTERIOR TIBIAL ARTERY. TWO ATTEMPTS WERE MADE TO INFLATE THE ARMADA 14 TO 8 ATMOSPHERES, BUT THE BALLOON DID NOT FULLY INFLATE. THE ARMADA 14 WAS REMOVED WITHOUT INCIDENT AND A LEAK WAS NOTED COMING FROM A HOLE ON THE ARMADA BALLOON. ANOTHER ARMADA WAS USED SUCCESSFULLY TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7928265 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.