Armada F2012-020Recalls & reported problems

Abbott Vascular Inc. · Armada 14 XT PTA Catheter 1.2 mm x 20 mm x 145 cm / Over-The-Wire · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08717648182839), received between Jun 14, 2019 and Jun 19, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Difficult to Advance” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance3
  • Material Rupture3
  • Difficult to Remove2
  • Deformation Due to Compressive Stress1
  • Material Twisted/Bent1
  • Stretched1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 19, 2024Difficult to Remove; Difficult to Remove

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE DISTAL ANTERIOR TIBIAL ARTERY WITH HEAVY CALCIFICATION, HEAVY TORTUOSITY AND 100% TOTAL OCCLUSION. THE 1.2X20 MM ARMADA XT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER BECAME STUCK WITH A NON-ABBOTT GUIDE WIRE AND BOTH DEVICES HAD TO BE REMOVED FROM THE ANATOMY TOGETHER AS A UNIT. THE VESSEL ACCESS WAS LOST AND A NEW WIRE AND BALLOON WERE USED TO COMPLETE THE PROCEDURE; HOWEVER, THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19567749 ↗
  • MalfunctionMar 28, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE RIGHT ANTERIOR TIBIAL ARTERY WITH HEAVY CALCIFICATION. THE 1.2X20MM ARMADA BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION; HOWEVER, THE BALLOON RUPTURED DURING THE FIRST INFLATION AT 10 ATMOSPHERES (ATM). THEREFORE, THE BDC WAS REMOVED FROM THE PATIENT. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER BALLOON. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18991949 ↗
  • MalfunctionJul 5, 2023Difficult to Remove; Deformation Due to Compressive Stress; Difficult to Advance; Material Twisted/Bent; Difficult to Remove; Deformation Due to Compressive Stress; Difficult to Advance; Material Twisted/Bent

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MODERATELY CALCIFIED, MODERATELY TORTUOUS, 80% STENOSED LESION IN THE DORSALIS PEDIS. A 1.2X20MMX145CM ARMADA 14 OVER THE WIRE (OTW) BALLOON DILATATION CATHETER (BDC) WAS ADVANCED OVER AN HT COMMAND 14 GUIDE WIRE (GW) AND ATTEMPTING TO NAVIGATE AROUND THE PEDAL ARCH, WHEN THE SHAFT OF THE BDC BECAME DEFORMED AND ADVANCING BECAME DIFFICULT. IT WAS DECIDED TO REMOVE THE GUIDE WIRE AND INJECT CONTRAST, HOWEVER THE WIRE COULD NOT BE REMOVED FROM THE BDC. THE GUIDE WIRE AND BDC WERE REMOVED AS A SINGLE UNIT, AND IT WAS NOTED THAT THE BALLOON VISIBLY BECAME BUNCHED UP ON THE WIRE CAUSING IT TO BECOME STUCK. A SECONDARY BALLOON AND GUIDE WIRE WERE USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAYS IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17259665 ↗
  • MalfunctionApr 13, 2023Stretched; Difficult to Advance; Stretched; Difficult to Advance

    SUBSEQUENT TO THE INITIALLY FILED REPORT, THE DEVICE WAS RECEIVED, AND THE BALLOON WAS TIGHTLY FOLDED INDICATING THAT THE BALLOON HAD NOT BEEN INFLATED. THE ACCOUNT STATED THAT THE BALLOON WAS NEVER INFLATED AS IT WAS USED TO CROSS A TIGHT LESION. AN UNSPECIFIED WIRE WAS SWAPPED AND THE BALLOON CAME OUT; AND, THE BALLOON WAS RE-INSERTED AND ATTEMPTED TO BE ADVANCED THROUGH THE TIGHT LESION. THE BALLOON FELT WEAK AND STRETCHED IN A SEGMENT AND WAS DECIDED NOT TO BE USED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16733872 ↗
  • MalfunctionOct 7, 2019Material Rupture; Difficult to Advance; Material Rupture; Difficult to Advance

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED LESION IN THE MILDLY TORTUOUS ANTERIOR TIBIAL ARTERY (ATA). THE 1.2X20 MM ARMADA 14 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER WAS ADVANCED WITH DIFFICULTY TO THE LESION. REPORTEDLY, THERE WERE NO ISSUES DURING PREPARATION OF THE DEVICE. THE CATHETER WAS INFLATED ONCE TO 8 ATMOSPHERES (ATM) AND RUPTURED. ANOTHER ARMADA 14 PTA CATHETER WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9164126 ↗
  • MalfunctionJun 14, 2019Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE ARMADA 14 XT BALLOON DILATATION CATHETER (BDC) WAS ADVANCED WITHOUT RESISTANCE, INFLATED TWICE TO 10 ATMOSPHERES AND RUPTURED. A NON-ABBOTT BDC WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8701821 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →