Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08717648207464), received between Dec 6, 2021 and May 11, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Rupture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 11, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED, 60% STENOSED LESION IN THE SUPERFICIAL FEMORAL ARTERY (SFA). THE 6.0X200MM ARMADA 18 BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE LESION WITHOUT RESISTANCE, HOWEVER, THE BALLOON RUPTURED AT 8 ATMOSPHERES (ATM) DURING THE FIRST INFLATION. THE BDC WAS REMOVED WITHOUT ISSUE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16915632 ↗
  • MalfunctionDec 6, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED, 95% STENOSED LESION IN THE LEFT SUPERIOR FEMORAL ARTERY (SFA). A 6X200MM ARMADA 18 BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION AND UPON THE FIRST INFLATION TO 10 ATMOSPHERES (ATM), THE BALLOON ENCOUNTERED CALCIUM AND RUPTURED. THE BDC WAS REMOVED WITHOUT ISSUE AND ANOTHER ARMADA 18 WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12943324 ↗

Other Catheter, angioplasty, peripheral, transluminal devices