Skip to content

Armada 1013464-120Recalls & reported problems

Abbott Vascular Inc. · Armada 18 PTA Catheter 2.5 mm x 120 mm x 150 cm / Over-The-Wire · Class 2

All models of Armada (213)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 08717648207815), received between Dec 16, 2025 and Dec 16, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Rupture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 16, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THE PROCEDURE WAS TO TREAT A MODERATELY STENOSED LESION IN THE LEFT PROXIMAL ANTERIOR TIBIAL ARTERY. THE 2.50X120MM ARMADA 18 BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION HOWEVER THE BALLOON RUPTURED DURING THE SECOND INFLATION AT 10 ATMOSPHERES. THE BDC WAS REMOVED AND ANOTHER ARMADA 18 BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23817263 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page