Armada 1013465-040Recalls & reported problems

Abbott Vascular Inc. · Armada 18 PTA Catheter 3.0 mm x 40 mm x 150 cm / Over-The-Wire · Class 2

All models of Armada (213)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 08717648207860), received between May 8, 2017 and May 31, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Leak/Splash2
  • Difficult to Advance1
  • Difficult to Remove1
  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 31, 2024Difficult to Remove; Difficult to Advance; Difficult to Remove; Difficult to Advance

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE TORTUOUS ILIAC ARTERY WHERE THERE WERE STENTS PRESENT AS WELL AS CALCIUM AND ANEURYSMS, VIA CONTRALATERAL APPROACH. THE 3X40 MM ARMADA 18 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER WAS ADVANCED OVER A COMMAND GUIDE WIRE THROUGH A 6FR SHEATH. THE ARMADA BALLOON WAS UNABLE TO BE ADVANCED TO THE LESION AND WAS ALSO UNABLE TO BE REMOVED FROM THE GUIDE WIRE. THE DEVICES WERE REMOVED TOGETHER AND THE COMMAND GUIDE WIRE WAS ATTEMPTED TO BE RE-ADVANCED; HOWEVER, IT WAS NOT POSSIBLE TO RECROSS THE LESION AND THE PROCEDURE WAS ABORTED. THERE WERE NO REPORTED ADVERSE PATIENT EFFECTS AND THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. THE PATIENT WILL COME BACK FOR ANOTHER PROCEDURE, ATTEMPTING ANTEGRADE APPROACH. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19436083 ↗
  • MalfunctionNov 7, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED SUPERFICIAL FEMORAL ARTERY (SFA). THE 3.0X40MM ARMADA 18 BALLOON DILATATION CATHETER (BDC) RUPTURED DURING THE FIRST INFLATION AT 4 ATMOSPHERES (ATM). THE BDC WAS REMOVED WITHOUT ISSUE AND A 7.0X40MM ARMADA 35 BDC WAS INFLATED TO 4 ATMS; HOWEVER, RUPTURED DURING THE FIRST INFLATION. THE ARMADA 35 BDC WAS REMOVED; HOWEVER, RESISTANCE WAS FELT WITH THE 6FR SHEATH. THE BALLOON WAS NOTED TO STILL REMAIN IN THE PATIENT. SNARING OF THE SEPARATED PORTION WAS ATTEMPTED; HOWEVER, NOT SUCCESSFUL. THE PATIENT WAS TAKEN INTO SURGERY TO REMOVE THE BALLOON; HOWEVER, IT IS UNKNOWN IF THE SEPARATED PORTION WAS SUCCESSFULLY REMOVED. THERE IS NO CLINICALLY SIGNIFICANT DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15749558 ↗
  • MalfunctionAug 29, 2018Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE POPLITEAL ARTERY WITH HEAVY CALCIFICATION. A 3.0 X 40 MM ARMADA 18 BALLOON CATHETER WAS ADVANCED OVER AN UNSPECIFIED GUIDE WIRE AND CROSSED THE TARGET LESION. THEN WHEN THE BALLOON CATHETER WAS PRESSURIZED FOR THE FIRST INFLATION, BLOOD WAS OBSERVED COMING BACK IN THE INFLATION DEVICE [LEAK]. THE BALLOON DID NOT SEEM TO HOLD PRESSURE. THE BALLOON CATHETER WAS IMMEDIATELY REMOVED. ANOTHER UNSPECIFIED BALLOON CATHETER WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7828747 ↗
  • MalfunctionMay 8, 2017Leak/Splash

    IT WAS REPORTED THAT DURING PREPARATION OF THE 3.0 X 40 MM ARMADA 18 DILATATION CATHETER, AIR WAS PULLED BACK INTO THE SYRINGE AND IT WAS THOUGHT THAT THERE WAS A LEAK AT THE HUB. THE DEVICE WAS NOT USED IN THE PATIENT ANATOMY AND THERE WAS NO ADVERSE PATIENT EFFECT. A SECOND ARMADA 18 WAS USED WITHOUT ISSUE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Patient Involvement

    MAUDE report 6553055 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →