Armada 1013465-060Recalls & reported problems

Abbott Vascular Inc. · Armada 18 PTA Catheter 3.0 mm x 60 mm x 150 cm / Over-The-Wire · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 08717648207877), received between Jun 9, 2021 and Sep 18, 2025 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture6
  • Leak/Splash3
  • Improper or Incorrect Procedure or Method1
  • Inflation Problem1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Foreign Body In Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A MILDLY CALCIFIED LESION IN THE ANTERIOR TIBIAL ARTERY. A 3.0X60MM ARMADA BALLOON DILATATION CATHETER (BDC) WAS PREPARED AND ADVANCED TO THE LESION WITH NO NOTED ISSUES, AND ATTEMPTED TO BE USED FOR VESSEL DILATATION. HOWEVER DURING THE FIRST INFLATION TO 8ATMS THE BDC WAS NOTED TO BE LEAKING FROM THE SHAFT OF THE DEVICE NEAR THE Y-PORT. THEREFORE THE BDC WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH THE USE OF THE ANOTHER ABBOTT DEVICE. THERE WERE NO ADVERSE PATIENT EFFECTS NOR CLINICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23090690 ↗
  • MalfunctionJun 2, 2025Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN A VESSEL IN THE LOWER EXTREMITY. A 3.0X60MM ARMADA 18 BALLOON DILATATION CATHETER (BDC) WAS ATTEMPTED TO BE USED FOR VESSEL DILATATION; HOWEVER, DURING THE FIRST INFLATION THE BALLOON WAS NOTED TO NOT INFLATE AND A BALLOON RUPTURE WAS NOTED. THERE WERE NO ADVERSE PATIENT EFFECTS NOR CLINICALLY SIGNIFICANT DELAY IN PROCEDURE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22124150 ↗
  • MalfunctionMar 26, 2025Inflation Problem; Leak/Splash; Inflation Problem; Leak/Splash

    IT WAS REPORTED THAT THE PROCEDURE PERFORMED WAS TO TREAT A RESTENOSIS, MODERATELY CALCIFIED RIGHT COMMON ILIAC ARTERY THAT IS 50% STENOSED. THE 3.0X60MM ARMADA BALLOON DILATATION CATHETER (BDC) THROUGH A 7F SHEATH, ACCESSED THE RIGHT COMMON FEMORAL ARTERY AND ADVANCED TO THE LESION. WHEN ATTEMPTING TO INFLATE THE ARMADA BALLOON, THE BALLOON WOULD LOSE PRESSURE AND COMPLETELY FAILED TO INFLATE. A LEAK AT THE HUB WAS ALSO OBSERVED. A NEW 3.0X60MM ARMADA BDC WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21701331 ↗
  • MalfunctionMay 24, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE PERONEAL ARTERY. THE 3X60MM ARMADA 18 BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION AND THE BALLOON WAS ATTEMPTED TO BE INFLATED TWICE, HOWEVER, THE BALLOON WAS NOTED TO BE LEAKING CONTRAST. THEREFORE, THE BDC WAS REMOVED FROM THE PATIENT. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. AFTER REMOVAL, OUTSIDE THE PATIENT THE BALLOON WAS INJECTED WITH CONTRAST AGAIN AND A HOLE WAS NOTED IN THE BALLOON. ANOTHER ARMADA WAS USED TO COMPLETE THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19393055 ↗
  • MalfunctionNov 3, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE LOWER EXTREMITY. FOUR BALLOON DILATATION CATHETERS (BDCS), 7.0X80MM ARMADA 35, 6.0X80MM ARMADA 35, AND TWO 3.0X60MM ARMADA 18S WERE INFLATED AT AN UNKNOWN PRESSURE BUT ALL BURST BEFORE REACHING NOMINAL PRESSURE. ANOTHER ABBOTT BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18067511 ↗
  • MalfunctionNov 3, 2023Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE LOWER EXTREMITY. FOUR BALLOON DILATATION CATHETERS (BDCS), 7.0X80MM ARMADA 35, 6.0X80MM ARMADA 35, AND TWO 3.0X60MM ARMADA 18S WERE INFLATED AT AN UNKNOWN PRESSURE BUT ALL BURST BEFORE REACHING NOMINAL PRESSURE. ANOTHER ABBOTT BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18067285 ↗
  • MalfunctionNov 2, 2021Material Rupture; Material Rupture

    SUBSEQUENT TO THE INITIAL REPORT, THE FOLLOWING INFORMATION WAS RECEIVED: IT WAS REPORTED THAT THE 3X60MM ARMADA 18 BALLOON DILATATION CATHETER (BDC) WAS PREPARED FOR USE WITHOUT ISSUE. THE LESION WAS MODERATELY CALCIFIED. THE BDC WAS ADVANCED AND POSITIONED AT THE TARGET LESION AND INFLATION WAS INITIATED; HOWEVER, THE BALLOON DID NOT HOLD PRESSURE AND CONTRAST COULD BE SEEN EXITING THE BALLOON. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12738592 ↗
  • MalfunctionNov 1, 2021Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT DURING PREP OF A 3X60MM ARMADA 18 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) CATHETER, BUBBLES WERE NOTED, AND THE DEVICE FAILED TO HOLD PRESSURE. THE DEVICE WAS NOT USED AND THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. RETURNED DEVICE ANALYSIS IDENTIFIED A TEAR IN THE BALLOON. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12733078 ↗
  • InjuryJun 9, 2021Material Rupture; Material Separation; Improper or Incorrect Procedure or Method; Material Rupture; Material Separation; Improper or Incorrect Procedure or Method

    SUBSEQUENT TO THE PREVIOUSLY FILED REPORT. IT WAS CONFIRMED THAT PART OF THE BALLOON DETACHED; PART OF THE MATERIAL WAS ABLE TO BE REMOVED, BUT THE PHYSICIAN WAS NOT ABLE TO REMOVE THE BALLOON MARKER AND DEPLOYED A SUPERA STENT TO COMPRESS THE MARKER AGAINST THE ARTERY WALL.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11971595 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →