Armada 1013465-150Recalls & reported problems
Abbott Vascular Inc. · Armada 18 PTA Catheter 3.0 mm x 150 mm x 150 cm / Over-The-Wire · Class 2
All models of Armada (213)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 08717648207914), received between Sep 22, 2017 and Feb 27, 2024 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Material Rupture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Rupture4
- Leak/Splash2
- Product Quality Problem1
Patient problems
- No Consequences Or Impact To Patient4
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 27, 2024Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE LOWER EXTREMITY ARTERY. THE 3.0X150MM ARMADA 18 BALLOON DILATATION CATHETER (BDC) RUPTURED AT 6 ATMOSPHERES (ATMS). THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18792828 ↗MalfunctionSep 15, 2021Material Rupture; Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT THE ANTERIOR TIBIAL ARTERY. TWO BALLOONS (3X120MM ARMADA 14 AND 3X150MM ARMADA 18) WERE USED AND BOTH RUPTURED WHEN THEY WERE INFLATED ONCE AT 14 ATMOSPHERES. ANOTHER SMALLER SIZED BALLOON WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12476838 ↗MalfunctionFeb 19, 2020Leak/Splash; Product Quality Problem; Leak/Splash; Product Quality Problem
IT WAS REPORTED THAT THE ARMADA 18 WAS SUCCESSFULLY INFLATED AT 8 ATMOSPHERES IN THE POSTERIOR TIBIAL ARTERY AND ANOTHER DISTAL ARTERY WITHOUT INCIDENT. THE ARMADA 18 WAS REMOVED FROM THE PATIENT AND THE CASE CONTINUED. THE PHYSICIAN WAS INTENDING TO USE THE ARMADA 18 A THIRD TIME IN THE PATIENT WHEN SHE NOTICED THAT THE BALLOON DID NOT APPEAR TO HAVE RE-WRAPPED CORRECTLY. IT WAS DECIDED TO USE AN ARMADA 35 TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED. ANALYSIS OF THE RETURNED DEVICE IDENTIFIED A LEAK.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9728251 ↗MalfunctionFeb 17, 2020Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN AN UNSPECIFIED CORONARY ARTERY. THE 3.0 X 150 MM ARMADA 18 BALLOON CATHETER WAS ADVANCED IN THE VESSEL; HOWEVER, DURING THE 3RD INFLATION IT WAS NOT POSSIBLE TO KEEP THE PRESSURE STABLE DUE TO A LEAK AT THE HUB. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE AND NO ADVERSE PATIENT EFFECT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9717656 ↗MalfunctionSep 22, 2017Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED PERONEAL ARTERY. A 3.0 X 150 MM ARMADA 18 BALLOON CATHETER RUPTURED DURING THE FIRST INFLATION BELOW NOMINAL PRESSURE. A SECOND 3.0 X 150 MM ARMADA 18 BALLOON CATHETER ALSO RUPTURED DURING THE FIRST INFLATION BELOW NOMINAL PRESSURE. BOTH DEVICES WERE PREPPED PER THE INSTRUCTIONS FOR USE AND THERE WAS NO RESISTANCE ADVANCING THE DEVICES. A NON-ABBOTT .035 BALLOON CATHETER WAS SUCCESSFULLY USED TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6888273 ↗MalfunctionSep 22, 2017Material Rupture
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A HEAVILY CALCIFIED PERONEAL ARTERY. A 3.0 X 150 MM ARMADA 18 BALLOON CATHETER RUPTURED DURING THE FIRST INFLATION BELOW NOMINAL PRESSURE. A SECOND 3.0 X 150 MM ARMADA 18 BALLOON CATHETER ALSO RUPTURED DURING THE FIRST INFLATION BELOW NOMINAL PRESSURE. BOTH DEVICES WERE PREPPED PER THE INSTRUCTIONS FOR USE AND THERE WAS NO RESISTANCE ADVANCING THE DEVICES. A NON-ABBOTT .035 BALLOON CATHETER WAS SUCCESSFULLY USED TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6888265 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page