Armada 1013468-040Recalls & reported problems

Abbott Vascular Inc. · Armada 18 PTA Catheter 5.0 mm x 40 mm x 150 cm / Over-The-Wire · Class 2

All models of Armada (213)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 08717648208171), received between Jun 1, 2017 and Feb 27, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture3
  • Difficult To Position1
  • Difficult to Remove1
  • Leak/Splash1

Patient problems

  • No Consequences Or Impact To Patient3
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 27, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION IN THE PROXIMAL SUPERFICIAL FEMORAL ARTERY (SFA). THE 5X40MM ARMADA 18 BALLOON DILATATION CATHETER (BDC) WAS ADVANCED TO THE TARGET LESION AND THE PRESSURE WAS APPLIED TO THE BALLOON. THE BALLOON DID NOT INFLATE AND A HOLE IN THE BALLOON WAS SUSPECTED. THE BDC WAS REMOVED FROM THE PATIENT. ANOTHER 5X40MM ARMADA 18 BDC WAS USED TO CONTINUE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND NO CLINICALLY SIGNIFICANT DELAY REPORTED IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18793564 ↗
  • MalfunctionJul 21, 2021Material Rupture; Material Rupture

    IT WAS REPORTED THE PROCEDURE WAS TO TREAT IN-STENT RESTENOSIS WITH NO CALCIFICATION IN THE LEFT SUPERFICIAL FEMORAL ARTERY (SFA). THE 5.0 X 40 MM ARMADA 18 BALLOON CATHETER WAS ADVANCED WITHOUT ISSUE. THE BALLOON RUPTURED AT 8 ATMOSPHERES (ATMS) DURING THE FIRST INFLATION. THE ARMADA 18 BALLOON CATHETER WAS REMOVED FROM THE PATIENT ANATOMY INTACT AND REPLACED WITH A 5.0 X 60 MM ARMADA 18 BALLOON CATHETER. THERE WAS NO ADVERSE PATIENT EFFECT. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12203416 ↗
  • MalfunctionSep 5, 2020Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED TO TREAT A LESION IN THE HEAVILY CALCIFIED SUPERFICIAL FEMORAL ARTERY. STANDARD PRE-DILATATION AND POST-DILATATION WERE PERFORMED, ALONG WITH LASER ATHERECTOMY. A 5X40MM ARMADA 18 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON RUPTURED AT 4 ATMOSPHERES. AN UNSPECIFIED BALLOON WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10498043 ↗
  • MalfunctionApr 13, 2018Difficult To Position; Difficult to Remove

    IT WAS REPORTED THAT THE ARMADA 018 BALLOON DILATATION CATHETER BECAME STUCK ON THE 018 STEEL CORE GUIDE WIRE DURING ADVANCEMENT TO THE UNCALCIFIED, NON-TORTOUS POPLITEAL ARTERY. THE ARMADA 018 WAS UNABLE TO BE REMOVED FROM THE GUIDE WIRE, SO BOTH WERE REMOVED FROM THE ANATOMY AS A UNIT. OUTSIDE THE ANATOMY, THE DEVICES WERE REMOVED FROM EACH OTHER. THE SAME GUIDE WIRE WAS USED WITH A NEW ARMADA 018 TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. THERE WAS NO ADDITIONAL INFORMATION PROVIDED

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7430666 ↗
  • MalfunctionJun 1, 2017Leak/Splash

    IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A SUPERFICIAL FEMORAL ARTERY. A 5.0 X 40 MM ARMADA 18 BALLOON CATHETER WAS ADVANCED TO THE LESION; HOWEVER, WHEN INFLATING, THE CATHETER WOULD NOT HOLD PRESSURE. OUTSIDE THE ANATOMY THE BALLOON WAS PUT IN WATER AND INFLATED AND THERE WAS NO BALLOON RUPTURE NOTED BUT THE CATHETER STILL WOULD NOT HOLD PRESSURE . ANOTHER ARMADA 18 BALLOON CATHETER WAS SUCCESSFULLY USED TO COMPLETE THE PROCEDURE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6605319 ↗

Other Catheter, angioplasty, peripheral, transluminal devices

All Catheter, angioplasty, peripheral, transluminal devices →