Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10381780126256), received between Feb 25, 2022 and Sep 6, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Remains Activated” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Remains Activated1
- Output below Specifications1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 6, 2024Output below Specifications; Output below Specifications
THIS IS 2 OF 2 REPORTS LINKED TO MFG REPORT NUMBER 3006697299-2024-00135: A FACILITY REPORTED THAT THE CUSA CLARITY FOOTSWITCH (C7002) VOLTAGE WAS CHECKED PRIOR TO SURGERY AND WAS BELOW THE SPECIFIED VALUE. THE DEVICE WAS REBOOTED, BUT IT COULD NOT BE USED, AND ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. SURGICAL DELAY WAS MORE THAN 30 MINUTES. THE FOLLOWING DAY, A COMPANY REPRESENTATIVE VISITED THE FACILITY TO CHECK THE DEVICE, AND RE-INSTALLED THE SOFTWARE WHICH RESOLVED THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20159191 ↗MalfunctionFeb 25, 2022Device Remains Activated; Device Remains Activated
A FACILITY REPORTED THAT DURING THE OPERATION OF THE CUSA CLARITY FOOTSWITCH (C7002) WHEN PRESSING THE ORANGE PEDAL THE PEDAL STAYS IN ACTIVATION MODUS. THEY CHANGED THE FOOT PEDAL TO ANOTHER CUSA CLARITY. THE EVENT LED TO INCREASE SURGERY TIME FOR 5 MINUTES; HOWEVER, THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13600087 ↗
Other Instrument, ultrasonic surgical devices
- Harmonic ACE HARH36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical4 deaths7,609 reports
- Harmonic HARHD36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical1 recall2 deaths5,037 reports
- Harmonic ACE HAR36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2,726 reports
- Harmonic HAR1136Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2 deaths1,538 reports
- THUNDERBEAT 5 mm,35 cm, Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls1,373 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls2 deaths1,128 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical2 recalls1 death970 reports
- Harmonic ACE HARH23Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical935 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page