CUSA® Clarity C7300Recalls & reported problems
Integra Lifesciences (Ireland) Limited · CUSA® Clarity Quick Connect Cartridge and Tubing Set · Class U
All models of CUSA Clarity (21)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10381780126287), received between Jan 19, 2023 and Jan 31, 2025 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Inability to Irrigate” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Inability to Irrigate3
- Break1
- Device Handling Problem1
- Failure to Align1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Bacterial Infection1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryJan 31, 2025Device Handling Problem; Device Handling Problem
MalfunctionOct 3, 2024Inability to Irrigate; Inability to Irrigate
THIS REPORT IS LINKED TO MFG REPORT NUMBER 3006697299-2024-00144 AND 3006697299-2024-00152, AND IS FOR THE 2ND SURGERY: AFTER RECEIVING ADDITIONAL INFORMATION ON MFG REPORT# 3006697299-2024-00144 REGARDING IRRIGATION ISSUES DURING SURGERY INVOLVING C7300 CUSA CLARITY QUICK CONNECT TUBING (C7300) INDICATING THAT THE DEVICE ONLY WORKED SEQUENTIALLY WHEN IRRIGATION WAS SET TO A MAXIMUM SPEED OF 20 ML/MIN), THE CUSTOMER EXPLAINED THAT THIS TYPE OF ISSUE WHICH INCREASED THE DURATION OF SURGERY BY MORE THAN 30 MINUTES CONTINUED TO HAPPEN AGAIN AND OCCURRED DURING 3 SURGICAL PROCEDURES IN TOTAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20362925 ↗MalfunctionOct 3, 2024Inability to Irrigate; Inability to Irrigate
THIS REPORT IS LINKED TO MFG REPORT NUMBER 3006697299-2024-00144 AND 3006697299-2024-00151, AND IS FOR THE 3RD SURGERY: AFTER RECEIVING ADDITIONAL INFORMATION ON MFG REPORT# 3006697299-2024-00144 REGARDING IRRIGATION ISSUES DURING SURGERY INVOLVING C7300 CUSA CLARITY QUICK CONNECT TUBING (C7300) INDICATING THAT THE DEVICE ONLY WORKED SEQUENTIALLY WHEN IRRIGATION WAS SET TO A MAXIMUM SPEED OF 20 ML/MIN), THE CUSTOMER EXPLAINED THAT THIS TYPE OF ISSUE WHICH INCREASED THE DURATION OF SURGERY BY MORE THAN 30 MINUTES CONTINUED TO HAPPEN AGAIN AND OCCURRED DURING 3 SURGICAL PROCEDURES IN TOTAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20362100 ↗MalfunctionSep 24, 2024Inability to Irrigate; Inability to Irrigate
A FACILITY REPORTED THAT THEY EXPERIENCED IRRIGATION ISSUES WITH THE CUSA CLARITY QUICK CONNECT TUBING SET (C7300) DURING A SURGICAL PROCEDURE. THE IRRIGATION ONLY WORKED SEQUENTIALLY WHEN IRRIGATION WAS SET AT MAX SPEED 20 ML/MIN. CUSTOMER TRIED SWITCHING EVERY PART (MACHINE, PEDAL, HANDPIECE, SALINE SOLUTION, TIP) AND AFTER SWITCHING THE TUBING, THE MACHINE STARTED WORKING NORMALLY. THIS FAULT OCCURRED TWICE AND THE CHANGING OF THE TUBING HAS WORKED BOTH TIMES WHICH LED THEM TO BELIEVE THAT THE LOT HAS A MANUFACTURING ISSUE. THE SURGERY WAS DELAYED BY APPROXIMATELY 1-2 HOURS. THE FACILITY HAS DISCARDED THE USED, CONTAMINATED TUBING SO THEY ARE NOT AVAILABLE TO BE SENT FOR EVALUATION. HOWEVER, THE REMAINING STERILE TUBING WILL BE SENT FOR INSPECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20292870 ↗MalfunctionSep 7, 2023Break; Break
MalfunctionJan 19, 2023Failure to Align; Failure to Align
A FACILITY REPORTED THAT PRIOR TO A LIVER RESECTION PROCEDURE, THE CUSA CLARITY QUICK CONNECT TUBING SET (C7300) IRRIGATION TUBE WAS MISALIGNED. SURGICAL TIME WAS EXTENDED MORE THAN 30 MINUTES. THE DEVICE WAS CHANGED TO A NEW ONE AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16211395 ↗
Other Instrument, ultrasonic surgical devices
- Harmonic ACE HARH36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical4 deaths7,609 reports
- Harmonic HARHD36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical1 recall2 deaths5,037 reports
- Harmonic ACE HAR36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2,726 reports
- Harmonic HAR1136Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2 deaths1,538 reports
- THUNDERBEAT 5 mm,35 cm, Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls1,373 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls2 deaths1,128 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical2 recalls1 death970 reports
- Harmonic ACE HARH23Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical935 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page