CUSA® Clarity C7300Recalls & reported problems

Integra Lifesciences (Ireland) Limited · CUSA® Clarity Quick Connect Cartridge and Tubing Set · Class U

All models of CUSA Clarity (21)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10381780126287), received between Jan 19, 2023 and Jan 31, 2025 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Inability to Irrigate” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Inability to Irrigate3
  • Break1
  • Device Handling Problem1
  • Failure to Align1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Bacterial Infection1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 31, 2025Device Handling Problem; Device Handling Problem

    N/A.

    Patient problems: Bacterial Infection

    MAUDE report 21283179 ↗
  • MalfunctionOct 3, 2024Inability to Irrigate; Inability to Irrigate

    THIS REPORT IS LINKED TO MFG REPORT NUMBER 3006697299-2024-00144 AND 3006697299-2024-00152, AND IS FOR THE 2ND SURGERY: AFTER RECEIVING ADDITIONAL INFORMATION ON MFG REPORT# 3006697299-2024-00144 REGARDING IRRIGATION ISSUES DURING SURGERY INVOLVING C7300 CUSA CLARITY QUICK CONNECT TUBING (C7300) INDICATING THAT THE DEVICE ONLY WORKED SEQUENTIALLY WHEN IRRIGATION WAS SET TO A MAXIMUM SPEED OF 20 ML/MIN), THE CUSTOMER EXPLAINED THAT THIS TYPE OF ISSUE WHICH INCREASED THE DURATION OF SURGERY BY MORE THAN 30 MINUTES CONTINUED TO HAPPEN AGAIN AND OCCURRED DURING 3 SURGICAL PROCEDURES IN TOTAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20362925 ↗
  • MalfunctionOct 3, 2024Inability to Irrigate; Inability to Irrigate

    THIS REPORT IS LINKED TO MFG REPORT NUMBER 3006697299-2024-00144 AND 3006697299-2024-00151, AND IS FOR THE 3RD SURGERY: AFTER RECEIVING ADDITIONAL INFORMATION ON MFG REPORT# 3006697299-2024-00144 REGARDING IRRIGATION ISSUES DURING SURGERY INVOLVING C7300 CUSA CLARITY QUICK CONNECT TUBING (C7300) INDICATING THAT THE DEVICE ONLY WORKED SEQUENTIALLY WHEN IRRIGATION WAS SET TO A MAXIMUM SPEED OF 20 ML/MIN), THE CUSTOMER EXPLAINED THAT THIS TYPE OF ISSUE WHICH INCREASED THE DURATION OF SURGERY BY MORE THAN 30 MINUTES CONTINUED TO HAPPEN AGAIN AND OCCURRED DURING 3 SURGICAL PROCEDURES IN TOTAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20362100 ↗
  • MalfunctionSep 24, 2024Inability to Irrigate; Inability to Irrigate

    A FACILITY REPORTED THAT THEY EXPERIENCED IRRIGATION ISSUES WITH THE CUSA CLARITY QUICK CONNECT TUBING SET (C7300) DURING A SURGICAL PROCEDURE. THE IRRIGATION ONLY WORKED SEQUENTIALLY WHEN IRRIGATION WAS SET AT MAX SPEED 20 ML/MIN. CUSTOMER TRIED SWITCHING EVERY PART (MACHINE, PEDAL, HANDPIECE, SALINE SOLUTION, TIP) AND AFTER SWITCHING THE TUBING, THE MACHINE STARTED WORKING NORMALLY. THIS FAULT OCCURRED TWICE AND THE CHANGING OF THE TUBING HAS WORKED BOTH TIMES WHICH LED THEM TO BELIEVE THAT THE LOT HAS A MANUFACTURING ISSUE. THE SURGERY WAS DELAYED BY APPROXIMATELY 1-2 HOURS. THE FACILITY HAS DISCARDED THE USED, CONTAMINATED TUBING SO THEY ARE NOT AVAILABLE TO BE SENT FOR EVALUATION. HOWEVER, THE REMAINING STERILE TUBING WILL BE SENT FOR INSPECTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20292870 ↗
  • MalfunctionSep 7, 2023Break; Break

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17702493 ↗
  • MalfunctionJan 19, 2023Failure to Align; Failure to Align

    A FACILITY REPORTED THAT PRIOR TO A LIVER RESECTION PROCEDURE, THE CUSA CLARITY QUICK CONNECT TUBING SET (C7300) IRRIGATION TUBE WAS MISALIGNED. SURGICAL TIME WAS EXTENDED MORE THAN 30 MINUTES. THE DEVICE WAS CHANGED TO A NEW ONE AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16211395 ↗

Other Instrument, ultrasonic surgical devices