CUSA® Clarity C7415SEARecalls & reported problems

Integra Lifesciences (Ireland) Limited · CUSA® Clarity 36kHz Curved Extended MicroTip™ Plus · Class U

All models of CUSA Clarity (21)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10381780126447), received between Apr 3, 2025 and May 8, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Crack” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 8, 2026Crack

    THIS REPORT IS 2 OF 2 FOR THE SECOND PRODUCT REPORTED AND IS LINKED TO MFG REPORT NUMBER 3006697299-2026-00044: A FACILITY REPORTED THAT THE CUSA CLARITY 36KHZ CURVED EXTENDED MICROTIP¿ (C7415S) WAS USED IN AN ENDOSCOPIC INTRACRANIAL TUMOR RESECTION WITH CUSA CONSOLE C7000 AND CUSA HANDPIECE C7036. HOWEVER, AFTER 5 MINUTES OF USE THE DEVICE BROKE (SETTING OF THE DEVICE WHEN IT HAPPENED: AMPLITUDE 80, TISSUE SELECT 0, ASPIRATION 70, IRRIGATION 3). A SECOND PRODUCT WAS USED BUT ALSO BROKE AFTER 10 MINUTES OF USE. THE MAXILLARY SINUS AND ROOT OF THE TIP WERE TOUCHING AND THAT MAY HAVE PUT A STRAIN ON THE PRODUCT. A THIRD PRODUCT WAS USED TO COMPLETE THE SURGERY, AND A CT SCAN WAS TAKEN TO ENSURE THAT NO PARTS WERE LEFT IN THE PATIENT. THERE WAS NO SURGICAL DELAY OR INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25125651 ↗
  • MalfunctionApr 3, 2025Crack; Crack

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21766204 ↗

Other Instrument, ultrasonic surgical devices