CUSA® Clarity C7421SRecalls & reported problems
Integra Lifesciences (Ireland) Limited · CUSA® Clarity 23kHz Single-Sided Bone Tip · Class U
All models of CUSA Clarity (21)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10381780536369), received between Nov 15, 2024 and Nov 15, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “Radiation Leak” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Radiation Leak1
Patient problems
- Unspecified Tissue Injury1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryNov 15, 2024Radiation Leak; Radiation Leak
A FACILITY REPORTED THAT THE USA CLARITY 23KHZ SINGLE SIDED BONE TIP (C7421SEA) WAS USED FOR AN "INTRACRANIAL TUMOR REMOVAL SURGERY¿, AND WHEN CUTTING THE INTERNAL AUDITORY MEATUS, DAMAGE WAS OBSERVED IN THE DURA MATER. THE SETTING OF THE DEVICE WHEN THE ISSUE OCCURRED: AMPLITUDE 80, TISSUE SELECT 0, ASPIRATION 90, IRRIGATION 10. THE SURGEON WAS NOT SURE IF THE DEVICE TOUCHED THE DURA MATER AND SUSPECTED THAT THE DEVICE IRRADIATED ULTRASONIC WAVES. HOWEVER, THE SURGEON WAS INITIALLY PLANNING TO MAKE AN INCISION IN THE PART OF DURA MATER THAT WAS DAMAGED; THEREFORE, THE EVENT DID NOT INTERFERE WITH THE OVERALL PROCEDURE. THERE WAS NO DELAY IN SURGERY, AND THE SAME DEVICE WAS USED TO COMPLETE THE SURGERY. ADDITIONAL INFORMATION RECEIVED INDICATING THE FOLLOWING: THE DEVICE DID NOT FALL OUT OF THE "DUTY CYCLE". ITS OUTPUT SETTING DURING PROCEDURE WAS AS FOLLOWS; AMPLITUDE 80, TISSUE SELECT 0, ASPIRATION 90, IRRIGATION 10. WHEN THE SALES REP ATTENDED THE SURGERY, IT SEEMED LIKE THE SURGENT DID NOT PUT TOO MUCH LOAD ON THE TIP. OTHER OUTPUT DEVICES WERE NOT USED AT THE SAME TIME.
Patient problems: Unspecified Tissue Injury
MAUDE report 20699155 ↗
Other Instrument, ultrasonic surgical devices
- Harmonic ACE HARH36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical4 deaths7,609 reports
- Harmonic HARHD36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical1 recall2 deaths5,037 reports
- Harmonic ACE HAR36Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2,726 reports
- Harmonic HAR1136Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical2 deaths1,538 reports
- THUNDERBEAT 5 mm,35 cm, Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls1,373 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical3 recalls2 deaths1,128 reports
- THUNDERBEAT 5 mm,35 cm,Front-actuated Grip Type S TB-0535FCSOlympus Medical Systems Corp. · Instrument, ultrasonic surgical2 recalls1 death970 reports
- Harmonic ACE HARH23Ethicon Endo-Surgery, LLC · Instrument, ultrasonic surgical935 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page