This Dexcom G4 Platinum transmitter is the wireless part of a continuous glucose monitoring system. Many reports describe the transmitter losing its connection to a pump, and some mention incorrect messages on the display. The records include 1,935 reports, 12 injuries and no deaths, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,935
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1,935 adverse event reports that list this exact device (DI 10386270000047), received between Aug 11, 2015 and May 22, 2024 — classified as 12 injuries, 1,923 malfunctions.
The most frequently reported specific device problem is “Wireless Communication Problem” (1,581 reports).
2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Wireless Communication Problem1,581
- Device Displays Incorrect Message232
- Invalid Sensing56
- Device Operates Differently Than Expected38
- Use of Device Problem27
- Imprecision14
- No Device Output4
- Failure to Sense3
Patient problems
- No Consequences Or Impact To Patient1,632
- No Known Impact Or Consequence To Patient289
- Burn(s)6
- Shock6
- Hypoglycemia5
- Loss of consciousness3
- No Clinical Signs, Symptoms or Conditions2
- Injury1
Recent adverse event reports
The 25 most recent of 1,935 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 22, 2024No Device Output; No Device Output
SUBSEQUENT TO THE INITIAL MDR, IT WAS DETERMINED THAT A REPORT WAS SUBMITTED IN ERROR. UPON FURTHER REVIEW, IT WAS DETERMINED THAT THE PATIENT ALLEGATION DOES NOT MEET THE CRITERIA OF A REPORTABLE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19373678 ↗MalfunctionFeb 17, 2021Communication or Transmission Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR AN UNSPECIFIED DURATION OF TIME. THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11338499 ↗MalfunctionJan 11, 2021Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A FAILED TRANSMITTER ERROR. THE TRANSMITTER HAD BEEN IN USE LESS THAN 3 MONTHS. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 11146697 ↗MalfunctionDec 9, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR AN UNSPECIFIED DURATION OF TIME. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10977220 ↗MalfunctionNov 10, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR GREATER THAN 1 HOUR. TRANSMITTER WAS REPLACED TO RESOLVE THE ISSUE. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10816262 ↗MalfunctionOct 30, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR AN UNSPECIFIED DURATION OF TIME. MULTIPLE CONTACT ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO OBTAIN ADDITIONAL INFORMATION REGARDING THE ISSUE; HOWEVER, THE CUSTOMER DID NOT RESPOND. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10763336 ↗MalfunctionOct 29, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR AN UNSPECIFIED DURATION OF TIME. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10756332 ↗MalfunctionOct 22, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR AN UNSPECIFIED DURATION OF TIME. REPORTEDLY, THE CUSTOMER WAS AROUND A RADIO WAVE TOWER, AND CONNECTION IS REGAINED ONCE LEAVING. NO ADDITIONAL PATIENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10718050 ↗MalfunctionOct 21, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR AN UNSPECIFIED DURATION OF TIME. REPORTEDLY, THE CUSTOMER WAS ATTEMPTING TO USE AN INCOMPATIBLE TRANSMITTER. CUSTOMER EDUCATED ON THE INCOMPATIBILITY OF THE TRANSMITTER WITH THE CUSTOMER'S CURRENT SOFTWARE. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10711644 ↗MalfunctionOct 6, 2020Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A FAILED TRANSMITTER ERROR. REPORTEDLY, THE TRANSMITTER HAD BEEN IN USE MORE THAN 6 MONTHS. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. TANDEM CUSTOMER TECHNICAL SUPPORT INSTRUCTED CUSTOMER TO USE A NEW TRANSMITTER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10638180 ↗MalfunctionSep 16, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR AN UNSPECIFIED DURATION OF TIME. THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10541264 ↗MalfunctionSep 15, 2020Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A FAILED TRANSMITTER ERROR. REPORTEDLY, THE TRANSMITTER HAD BEEN IN USE MORE THAN 3 MONTHS. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. TANDEM CUSTOMER TECHNICAL SUPPORT INSTRUCTED CUSTOMER TO USE A NEW TRANSMITTER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10537501 ↗MalfunctionAug 7, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR GREATER THAN 1 HOUR. REPORTEDLY, THERE WAS OBSTRUCTION BETWEEN THE PUMP AND THE TRANSMITTER. TRANSMITTER WAS REPLACED TO RESOLVE THE ISSUE. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10380107 ↗MalfunctionAug 4, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT INTERMITTENT LOSS OF CONNECTION ISSUES OCCURRED BETWEEN THE TRANSMITTER AND THE PUMP. REPORTEDLY, THE TRANSMITTER HAD BEEN IN USE LONGER THAN 90 DAYS; CUSTOMER SUPPORT INSTRUCTED CUSTOMER TO CHANGE TRANSMITTER. NO ADDITIONAL PATIENT INFORMATION WAS AVAILABLE
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10362448 ↗MalfunctionJul 30, 2020Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A FAILED TRANSMITTER ERROR. THE TRANSMITTER HAD BEEN IN USE LESS THAN 3 MONTHS. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10347774 ↗MalfunctionJul 16, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT INTERMITTENT LOSS OF CONNECTION ISSUES OCCURRED BETWEEN THE TRANSMITTER AND THE PUMP. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10283421 ↗MalfunctionJul 9, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR GREATER THAN 1 HOUR. REPORTEDLY, THE TRANSMITTER HAD BEEN IN USE LONGER THAN 6 MONTHS. MULTIPLE CONTACT ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO OBTAIN ADDITIONAL INFORMATION REGARDING THE ISSUE; HOWEVER, THE CUSTOMER DID NOT RESPOND. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10253418 ↗MalfunctionJul 8, 2020Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A FAILED TRANSMITTER ERROR. THE TRANSMITTER HAD BEEN IN USE LESS THAN 3 MONTHS. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10246640 ↗MalfunctionJul 2, 2020Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT INTERMITTENT INVALID TRANSMITTER ID ALERTS OCCURRED. DESPITE MULTIPLE ATTEMPTS, A SENSOR SESSION WAS UNABLE TO BE INITIATED. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10224297 ↗MalfunctionJul 1, 2020Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A FAILED TRANSMITTER ERROR. THE TRANSMITTER HAD BEEN IN USE LESS THAN 3 MONTHS. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE. A ROOT CAUSE COULD NOT BE DETERMINED. A REPLACEMENT TRANSMITTER WAS SENT TO THE CUSTOMER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10219435 ↗MalfunctionJun 19, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR AN UNSPECIFIED DURATION OF TIME. THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10176081 ↗MalfunctionJun 18, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR GREATER THAN 1 HOUR. TRANSMITTER WAS REPLACED TO RESOLVE THE ISSUE. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10168357 ↗MalfunctionJun 5, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR GREATER THAN 1 HOUR. TRANSMITTER WAS REPLACED TO RESOLVE THE ISSUE. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10125069 ↗MalfunctionMay 21, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR GREATER THAN 1 HOUR. THE TRANSMITTER ULTIMATELY REGAINED CONNECTION WITH THE PUMP. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10079218 ↗MalfunctionMay 20, 2020Wireless Communication Problem; Wireless Communication Problem
IT WAS REPORTED THAT THE TRANSMITTER LOST CONNECTION WITH THE PUMP FOR GREATER THAN 1 HOUR. TRANSMITTER WAS REPLACED TO RESOLVE THE ISSUE. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10075251 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.
- Z-1560-2020 · Feb 3, 2020 · TerminatedDexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a
- Z-1561-2020 · Feb 3, 2020 · TerminatedDexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
Other Sensor, Glucose, Invasive devices
- Dexcom G4 Platinum Sensor 9500-27Dexcom, Inc. · Sensor, Glucose, Invasive2 recalls2 deaths99,426 reports
- Dexcom G5 Transmitter Kit STT-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 death58,932 reports
- Dexcom G5 Mobile IOS App SW10611Dexcom, Inc. · Sensor, Glucose, Invasive54,544 reports
- Dexcom G5 Transmitter 9438-06Dexcom, Inc. · Sensor, Glucose, Invasive15,360 reports
- Dexcom G5 Mobile Receiver MT22719Dexcom, Inc. · Sensor, Glucose, Invasive1 death12,594 reports
- Dexcom G4 SBT Transmitter 9438-05Dexcom, Inc. · Sensor, Glucose, Invasive1 death9,041 reports
- Dexcom G5 Mobile Receiver MT22719-BLUDexcom, Inc. · Sensor, Glucose, Invasive7,685 reports
- Dexcom G5 Mobile Receiver Kit STK-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 recall6,345 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page