Skip to content

Dexcom G4 PLATINUM Receiver with Share Kit US STK-DR-001Recalls & reported problems

Dexcom, Inc. · G4 PLATINUM Receiver with Share Kit US Adult Black · Class 3

All models of Dexcom G4 Platinum (41) · Dexcom G4 PLATINUM Receiver with Share Kit US (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Dexcom G4 PLATINUM receiver is the display unit that goes with a Share Kit for glucose monitoring. The 1,533 reports most often describe the receiver screen freezing, error icons appearing, the receiver shutting down unexpectedly, and wireless connection problems. A few reports also list low blood sugar among patient problems. The reports list no deaths and 12 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,533
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,533 adverse event reports that list this exact device (DI 10386270000153), received between Aug 17, 2016 and Dec 12, 2020 — classified as 12 injuries, 1,521 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (584 reports).

2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device Displays Incorrect Message584
  • Defective Alarm236
  • Wireless Communication Problem172
  • Unable to Obtain Readings161
  • No Audible Alarm121
  • Inappropriate or Unexpected Reset81
  • Device Inoperable75
  • Unexpected Shutdown47

Patient problems

  • No Consequences Or Impact To Patient1,519
  • Hypoglycemia10
  • Loss of consciousness7
  • Confusion/ Disorientation1
  • Patient Problem/Medical Problem1
  • Seizures1
  • Sweating1

Recent adverse event reports

The 25 most recent of 1,533 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 12, 2020Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE RECEIVER DISPLAY WAS FROZEN. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10995899 ↗
  • MalfunctionDec 11, 2020Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10991484 ↗
  • MalfunctionNov 2, 2020Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10770696 ↗
  • MalfunctionSep 24, 2020Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10580541 ↗
  • MalfunctionSep 22, 2020No Audible Alarm; No Audible Alarm

    THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. RECEIVER CHARGE AND BOOT TESTS WERE PERFORMED AND PASSED. FUNCTIONAL TESTS WERE PERFORMED AND PASSED. MANUAL FUNCTIONAL "TRY IT" TESTS WERE PERFORMED AND PASSED. AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. THE AUDIO SPEAKER RESISTANCE WAS MEASURED AND PASSED. THE RECEIVER LOG WAS DOWNLOADED AND REVIEWED FINDING NO ERROR RELATED TO THE COMPLAINT. THE ALLEGATION WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10566845 ↗
  • MalfunctionSep 10, 2020Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10515069 ↗
  • MalfunctionSep 10, 2020Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10514491 ↗
  • MalfunctionSep 9, 2020Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10507282 ↗
  • MalfunctionAug 17, 2020Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10413926 ↗
  • MalfunctionAug 17, 2020Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10413801 ↗
  • MalfunctionAug 4, 2020Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT INTERMITTENT AUDIO OUTPUT OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10364495 ↗
  • MalfunctionJul 6, 2020Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10233396 ↗
  • MalfunctionJun 3, 2020Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10114583 ↗
  • MalfunctionMay 29, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10102794 ↗
  • MalfunctionMay 29, 2020Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10102793 ↗
  • MalfunctionMay 29, 2020No Audible Alarm; No Audible Alarm

    IT WAS REPORTED THAT THE RECEIVER HAD NO AUDIO OUTPUT. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10100985 ↗
  • MalfunctionMay 18, 2020Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10067026 ↗
  • MalfunctionMay 8, 2020Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10038840 ↗
  • MalfunctionMay 8, 2020Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10038217 ↗
  • MalfunctionMay 1, 2020Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE RECEIVER DISPLAY WAS FROZEN. DATA WAS RECEIVED BUT NOT INVESTIGATED AS DATA WILL NOT CONFIRM THE ISSUE. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10017192 ↗
  • MalfunctionMar 24, 2020Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT INTERMITTENT AUDIO OUTPUT OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9871149 ↗
  • MalfunctionMar 18, 2020Unexpected Shutdown; Unexpected Shutdown

    IT WAS REPORTED THAT UNEXPECTED RECEIVER SHUTDOWN OCCURRED. THE PRODUCT WAS EVALUATED. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASS. ATTEMPT TO CHARGE AND BOOT THE UNIT WAS PERFORMED AND FAILED DUE TO DAMAGED BATTERY. . FUNCTIONAL TEST WAS UNABLE TO PERFORM. INTERNAL TEST WAS PERFORMED PASSED THE LOG WAS DOWNLOADED AND REVIEWED FINDING ERROR RELATED TO COMPLAINT. THE ALLEGATION WAS CONFIRMED. THE PROBABLE CAUSE WAS DETERMINED TO BE FAIL-RTT LOSS. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9849075 ↗
  • MalfunctionMar 16, 2020Wireless Communication Problem; Wireless Communication Problem

    IT WAS REPORTED THAT SIGNAL LOSS OVER ONE HOUR OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9837153 ↗
  • MalfunctionMar 2, 2020Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9775189 ↗
  • MalfunctionFeb 3, 2020Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE RECEIVER WAS OVERHEATING. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9659861 ↗

FDA clearance and classification

Its record lists premarket approval application P120005, decided on Oct 5, 2012 with FDA decision code APWD (not a standard approval code). Check the FDA record for its current status. FDA record ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

FDA definition of this device type: The device is a glucose-monitoring device indicated for detecting trends and tracking patterns in adults (age 18 and older) with diabetes. The device is indicated for use as an adjunctive device to complement, not replace, information obtained from standard home glucose monitoring devices. The system aids in the detection of episodes of hyperglycemia and hypoglycemia, facilitating both acute and long-term therapy adjustments.

Recalls under the same FDA submission

These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.

  • Z-1561-2020 · Feb 3, 2020 · Terminated
    Dexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
  • Z-1560-2020 · Feb 3, 2020 · Terminated
    Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a

Common questions about the Dexcom G4 PLATINUM Receiver with Share Kit US STK-DR-001

Has the Dexcom G4 PLATINUM Receiver with Share Kit US STK-DR-001 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10386270000153), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Dexcom G4 PLATINUM Receiver with Share Kit US STK-DR-001?

The FDA's MAUDE database lists 1,533 adverse event reports for this device, received between Aug 17, 2016 and Dec 12, 2020. In those reports, the most common specific device problems are “Device Displays Incorrect Message” (584), “Defective Alarm” (236) and “Wireless Communication Problem” (172); the most common patient problems are “Hypoglycemia” (10), “Loss of consciousness” (7) and “Confusion/ Disorientation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Dexcom G4 PLATINUM Receiver with Share Kit US STK-DR-001?

None of the reports are classified as deaths, 12 as injuries and 1,521 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Dexcom G4 PLATINUM Receiver with Share Kit US STK-DR-001, and is it still sold?

The FDA's device database lists Dexcom, Inc. as the company responsible for it (brand Dexcom G4 PLATINUM Receiver with Share Kit US, model STK-DR-001). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Sensor, Glucose, Invasive devices

All Sensor, Glucose, Invasive devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page