Skip to content

Dexcom G4 PLATINUM Receiver with Share Kit US STK-DR-PNKRecalls & reported problems

Dexcom, Inc. · G4 PLATINUM Receiver with Share Kit US Adult Pink · Class 3

Quick summary

This Dexcom G4 Platinum receiver shows glucose readings from a sensor worn on the body. The FDA database holds 542 reports for it, many describing incorrect error messages, alarm problems and wireless connection issues. Most reports describe no impact on the patient, with 4 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
542
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 542 adverse event reports that list this exact device (DI 10386270000160), received between Sep 10, 2015 and Mar 4, 2020 — classified as 4 injuries, 538 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (221 reports).

2 other recalls cover devices cleared under the same FDA submission (P120005); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Device Displays Incorrect Message221
  • Defective Alarm74
  • No Audible Alarm59
  • Wireless Communication Problem50
  • Device Inoperable42
  • Unable to Obtain Readings35
  • Inappropriate or Unexpected Reset32
  • Use of Device Problem22

Patient problems

  • No Consequences Or Impact To Patient538
  • Hypoglycemia4
  • Confusion/ Disorientation2
  • Loss of consciousness1

Recent adverse event reports

The 25 most recent of 542 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 4, 2020Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9789691 ↗
  • MalfunctionNov 7, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9293556 ↗
  • MalfunctionAug 2, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8854634 ↗
  • MalfunctionAug 1, 2019Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT INTERMITTENT AUDIO OUTPUT OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8851148 ↗
  • MalfunctionJul 17, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT ERROR ICON DISPLAYED OCCURRED. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8799273 ↗
  • MalfunctionJun 24, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT A "HW" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8728294 ↗
  • MalfunctionMay 20, 2019Unable to Obtain Readings; Unable to Obtain Readings

    IT WAS REPORTED THAT A "HWBBT" ERROR AND A "CALL TECH SUPPORT " ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS PROVIDED FOR EVALUATION. CONFIRMATION OF THE ALLEGATION AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8623891 ↗
  • MalfunctionFeb 27, 2019Defective Alarm; Defective Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, THE RECEIVER HAD INTERMITTENT VIBRATION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING INTERMITTENT VIBRATION COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8374931 ↗
  • MalfunctionFeb 26, 2019Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, AN ERR67 WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS NOT AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8370266 ↗
  • MalfunctionFeb 20, 2019Unexpected Shutdown; Unexpected Shutdown

    THE RECEIVER WAS RETURNED FOR EVALUATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. A RECEIVER CHARGE AND BOOT WAS PERFORMED AND PASSED. A FUNCTIONAL TEST WAS PERFORMED AND PASSED. A RECEIVER DOWNLOAD WAS PERFORMED AND PASSED. THE RECEIVER CASE WAS OPENED, AND AN INTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. CONFIRMATION OF THE COMPLAINT WAS NOT CONFIRMED. THE PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8356304 ↗
  • MalfunctionFeb 19, 2019Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, AN ERR HWBBT WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS NOT AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8349635 ↗
  • MalfunctionFeb 5, 2019Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, AN "ERR 67" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8304577 ↗
  • MalfunctionJan 31, 2019Defective Alarm; Defective Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT ON (B)(6) 2019, THE RECEIVER HAD INTERMITTENT VIBRATION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING INTERMITTENT VIBRATION COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8292929 ↗
  • MalfunctionJan 31, 2019Defective Alarm; Defective Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT ON (B)(6) 2019, THE RECEIVER HAD INTERMITTENT AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING INTERMITTENT AUDIO OUTPUT COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8292902 ↗
  • MalfunctionJan 31, 2019Defective Alarm; Defective Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT ON (B)(6) 2019, THE RECEIVER HAD INTERMITTENT VIBRATION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING INTERMITTENT VIBRATION COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8292885 ↗
  • MalfunctionJan 31, 2019Defective Alarm; Defective Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT ON (B)(6) 2019, THE RECEIVER HAD INTERMITTENT AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING INTERMITTENT AUDIO OUTPUT COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8292870 ↗
  • MalfunctionJan 31, 2019Defective Alarm; Defective Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT ON (B)(6) 2019, THE RECEIVER HAD INTERMITTENT VIBRATION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING INTERMITTENT VIBRATION COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8292853 ↗
  • MalfunctionJan 31, 2019Defective Alarm; Defective Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2019 THAT ON (B)(6) 2019, THE RECEIVER HAD INTERMITTENT AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING INTERMITTENT AUDIO OUTPUT COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8292846 ↗
  • MalfunctionJan 30, 2019Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, AN "HWBBT" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8290588 ↗
  • MalfunctionJan 29, 2019Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2019, THAT ON (B)(6) 2019, AN "HWBBT" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8287624 ↗
  • MalfunctionDec 31, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, A "CALL TECH SUPPORT" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8206944 ↗
  • MalfunctionDec 18, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, AN "ERR 68" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8176375 ↗
  • MalfunctionDec 13, 2018Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018 THE RECEIVER INITIALIZED WITHOUT A MANUAL RESTART. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. THE DEVICE HAS BEEN RECEIVED FOR INVESTIGATION. A FOLLOW-UP WILL BE SUBMITTED UPON COMPLETION OF DEVICE INVESTIGATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8161585 ↗
  • MalfunctionDec 12, 2018Unable to Obtain Readings; Unable to Obtain Readings

    DEXCOM WAS MADE AWARE ON (B)(4) 2018, THAT ON (B)(6) 2018, AN "HWBBTT" ERROR WAS DISPLAYED ON THE RECEIVER. NO PRODUCT OR DATA WAS AVAILABLE FOR EVALUATION. CONFIRMATION OF THE ERROR ICON AND A PROBABLE CAUSE COULD NOT BE DETERMINED. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8158283 ↗
  • MalfunctionDec 11, 2018No Audible Alarm; No Audible Alarm

    DEXCOM WAS MADE AWARE ON (B)(4) 2018 THAT ON (B)(6) 2018, THE RECEIVER HAD NO AUDIO OUTPUT. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE. NO PRODUCT WAS PROVIDED FOR EVALUATION. THE COMPLAINT CONFIRMATION OF THE RECEIVER HAVING NO AUDIO OUTPUT COULD NOT BE DETERMINED. A PROBABLE CAUSE COULD NOT BE DETERMINED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8153710 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P120005. One submission often covers many models, so check the product and lot details in each record.

  • Z-1561-2020 · Feb 3, 2020 · Terminated
    Dexcom G5 Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracki
  • Z-1560-2020 · Feb 3, 2020 · Terminated
    Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends a

Other Sensor, Glucose, Invasive devices

All Sensor, Glucose, Invasive devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.