DELTA XTEND 130738106Recalls & reported problems
Depuy (Ireland) · DELTA Xtend Retentive Humeral PE Cup Dia38 /+6R RET · Class 2 · Implant
All models of DELTA XTEND (194)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 21
- Deaths reported
- 0
- Injuries reported
- 19
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 21 adverse event reports that list this exact device (DI 10603295027355), received between Oct 11, 2017 and Oct 15, 2025 — classified as 19 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (9 reports).
10 other recalls cover devices cleared under the same FDA submission (K062250, K081620, K120174, K192855); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated9
- Adverse Event Without Identified Device or Use Problem7
- Difficult to Insert1
- Fitting Problem1
- Insufficient Device Problem Information1
- Material Erosion1
- Naturally Worn1
Patient problems
- Joint Dislocation11
- No Code Available5
- Pain5
- Unspecified Infection3
- Joint Laxity2
- No Consequences Or Impact To Patient2
- Loss of Range of Motion1
Recent adverse event reports
The 21 most recent of 21 reports, in the reporter's own words as published by the FDA.
InjuryOct 15, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISLOCATION. DOI: (B)(6) 2024, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 23296107 ↗InjuryJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT HAD C. ACNES INFECTION AS WELL AS PAIN AND JOINT INSTABILITY. THIS IS PATIENT'S SECOND REVISION.
Patient problems: Joint Laxity; Pain; Unspecified Infection
MAUDE report 22125091 ↗InjuryJan 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT HAD INFECTED SHOULDER. ALL IMPLANTS REMOVED AND ANTIBIOTIC SPACER PLACED. NO DELAYS IN SURGERY. DOI: (B)(6) 2023. DOR: (B)(6) 2024. AFFECTED SIDE: LEFT SHOULDER
Patient problems: Unspecified Infection
MAUDE report 21079590 ↗InjuryDec 12, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT GLENOSPHERE (SHORT SCREW) CANNOT BE IMPLANTED. THE GLENOSPHERE SCREW CANNOT BE IN METAGLENE HOLE. SHORT SCREW PREVENTS VIEW AND FEEL OPTION. SOFT TISSUE FOLD IS NOT VISIBLE. THE PATIENT EXPERIENCED LUXATION OF SHOULDER JOINT SAME DAY POST OP AND REQUIRED REVISION SURGERY.
Patient problems: Joint Dislocation
MAUDE report 20914188 ↗InjuryMay 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INFECTION. EVERYTHING WAS EXPLANTED AND A SPACED WAS PUT IN. DOI: (B)(6) 2023. DOR: (B)(6) 2024. AFFECTED SIDE: RIGHT KNEE.
Patient problems: Unspecified Infection
MAUDE report 19384346 ↗InjuryJan 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO INSTABILITY. DOI: UNKNOWN. DOR: (B)(6) 2024. AFFECTED SIDE: UNKNOWN SHOULDER.
Patient problems: Joint Laxity
MAUDE report 18483550 ↗InjurySep 14, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE SURGEON BELIEVED THE PATIENT HAD A FALL AND DISLOCATED HER SHOULDER. HE DECIDED TO REMOVE THE CURRENT CUP AND REPLACE WITH A THICKER ONE FOR STABILITY. TIME WAS NOT DELAYED AND NOTHING SEEMED TO BE WRONG WITH ANY OF OUR IMPLANTS. DOI: (B)(6) 2023. DOR: (B)(6) 2023. AFFECTED SIDE: RIGHT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 17743680 ↗InjuryAug 18, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS REVISED DUE TO PAIN. IT IS UNKNOWN IF THERE WAS ANY SURGICAL DELAY. DOI: UNKNOWN, DOR: (B)(6) 2023, AFFECTED SIDE: LEFT SHOULDER.
Patient problems: Pain
MAUDE report 17577069 ↗InjuryAug 24, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
ADDITIONAL INFORMATION RECEIVED INDICATED THAT THERE WAS NO CEMENT USED.
Patient problems: Joint Dislocation
MAUDE report 12359744 ↗InjuryJul 6, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PATIENT WAS REVISED TO ADDRESS A RECENT DISLOCATION. THE PATIENT HAD A PRIOR REVERSE CONSTRUCT FOR FRACTURE. DOI: (B)(6) 2021. DOR: (B)(6) 2021. RIGHT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 12121703 ↗InjuryDec 23, 2020Naturally Worn; Naturally Worn
REVISION RIGHT DELTA XTEND TSR. PATIENT UNDERWENT PRIMARY SURGERY IN 2012 AT ANOTHER FACILITY AND DIFFERENT SURGEON. PRESENTED TO CURRENT SURGEON COMPLAINING OF PAIN. SURGEON SUSPECTED CAUSE AS SCAPULAR NOTCHING. SURGERY UNDERTAKEN TO REMOVE POLY INSERT, GLENOSPHERE AND BROKEN INFERIOR METAGLENE SCREW. BROKEN PORTION OF SCREW UNABLE TO BE REMOVED AND LEFT IN-SITU. PREVIOUS 38 ECC GLENOSPHERE REMOVED AND A 42 ECC INSERTED. TRAIL REDUCTION WITH 42, + 6 STD POLY INSERT. SURGEON SATISFIED WITH RANGE OF MOTION, DEFINITIVE 42, + 6 STD POLY IMPACTED AND SHOULDER REDUCED.
Patient problems: Pain
MAUDE report 11065248 ↗MalfunctionJul 21, 2020Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT SURGEON HAD GREAT EXPOSURE AND VISIBILITY. WHEN IMPACTING THE POLY, IT APPEARED TO LOCK IN BUT WAS EASILY LIFTED OUT WITH A FREER ELEVATOR. THE SURGEON REPEATED AND WASN'T COMFORTABLE WITH THE STABILITY OF THE LINER. HE RAN DOWNSTAIRS AND GRABBED ANOTHER LINER. THE SECOND POLY LOCKED IN PLACE LIKE NORMAL. NEED A REPORT BACK TO GIVE THE HOSPITAL AS TO WHETHER THE LOCKING MECHANISM WAS DEFECTIVE OR WAS THERE ANOTHER REASON WHY IT WOULDN'T PROPERLY ENGAGE.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 10300857 ↗InjuryJul 6, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
REVISION OF DELTA XTEND REVERSE SHOULDER REPLACEMENT. PT HAD A FALL AND WAS EXPERIENCING INSTABILITY AND PAIN. X-RAY SHOWED DISPLACEMENT OF STEM. ALL IMPLANTS REMOVED FROM PT, NO IMPLANTS PLACED BACK IN. IF OTHER, DESCRIBE --> REVERSE SHOULDER REPLACEMENT REVISION. DID THE PATIENT EXPERIENCE A POST-OP DEVICE MALFUNCTION? --> NO. DID THE PATIENT EXPERIENCE AN ADVERSE EVENT SUCH AS INFECTION, NON-UNION, ALLERGIC REACTION, OSTEOPOROSIS, OVERLOADING, PAIN, DEGENERATIVE DISEASES, BLEEDING OR OOZING? --> YES, DID THE PATIENT REQUIRE REVISION SURGERY OR HARDWARE REMOVAL? --> YES, FACILITY NAME OF ORIGINAL IMPLANT --> (B)(6) MEMORIAL HOSPITAL , WAS DEVICE EXPLANTED? --> TRUE, HARDWARE/EXPLANT REMOVAL DUE TO: --> PT HAD A FALL, DISPLACED THE STEM, LEAD TO INSTABILITY AND PAIN , DID PATIENT REQUIRE REVISION SURGERY? --> FALSE, PATIENT STATUS/ OUTCOME / CONSEQUENCES --> YES, PATIENT CONSEQUENCE DESCRIPTION/WAS THERE A CLINICAL OUTCOME EXPERIENCED BY THE PATIENT (INFECTION, INFLAMMATION, ETC.)?--> ALL IMPLANTS HAVE BEEN REMOVED FROM PT. NO IMPLANTS WERE PUT BACK IN PLACE - GIRDLESTONE., WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED:
Patient problems: No Code Available; Pain
MAUDE report 10237683 ↗InjuryMar 4, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
THE PATIENT WAS REVISED TO ADDRESS DISLOCATION. DOI: (B)(6) 2019; DOR: (B)(6) 2020; LEFT SHOULDER.
Patient problems: Joint Dislocation; No Code Available
MAUDE report 9787895 ↗InjuryJan 17, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ARTHROPLASTY HAS DISLOCATED THE REVERSE SHOULDER REPLACEMENT AND WILL BE REVISED FOR INSTABILITY.
Patient problems: Joint Dislocation; No Code Available
MAUDE report 8256555 ↗InjuryJan 9, 2019Material Erosion; Material Erosion
PATIENT WAS REVISED TO ADDRESS SHOULDER PAIN AND LACK OF MOBILITY. DURING THE REVISION SURGERY, INSIDE THE CAPSULE THE INFERIOR LOCKING SCREW WAS REMOVED DUE TO IT ERODING THE HUMERAL CUP. NO SCREW WAS USED TO REPLACE IT. A +9 MM SPACER AND A 38 +6 RETENTIVE CUP WERE USED TO GAIN APPROPRIATE STABILITY. DOI: (B)(6) 2018; DOR: (B)(6) 2018; RIGHT SHOULDER.
Patient problems: Loss of Range of Motion; No Code Available; Pain
MAUDE report 8231245 ↗MalfunctionAug 21, 2018Fitting Problem; Fitting Problem
POLLY CUP 38+6R WOULD NOT SEAT IN EPIPHYSIS. PATIENT CONSEQUENCE? NO. ACTION TAKEN FOR PROCEDURE: OPENED ANOTHER IMPLANT AND IT WORKED. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? YES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7801120 ↗InjuryJan 19, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated
PRIMARY (B)(6) 2017, REVISION (B)(6) 2017, REVISION (B)(6) 2017, REVISION (B)(6) 2017. RECURRENT DISLOCATION POST RTSJR. THIS IS THE THIRD REVISION OF THIS JOINT REPLACEMENT. SECOND REVISION INCLUDED USE OF A RETENTIVE SPACER. THERE WAS NO ISSUE WITH VERSION OF THE CEMENTED STEM IN SITU (AS PER SURGEON). THIS WAS LEFT IN SITU AND THE LINER AND GLENOSPHERE WERE UPSIZED. IT WAS SUGGESTED THAT THE DISLOCATIONS WERE POSSIBLY ANTERIOR. THE SURGEON POSITIONED AN ECCENTRIC GLENOSHERE MORE ANTERIORALLY DESPITE BEING INFORMED THEY WERE NOT MEANT TO BE USED IN THIS WAY AND A TRIAL WOULD NOT BE POSSIBLE. THE SHOULDER WAS REDUCED SUCCESSFULLY AND THE SURGEON REPORTED STABILITY BY THE FINISH OF THE PROCEDURE.
Patient problems: Joint Dislocation
MAUDE report 7206483 ↗InjuryDec 20, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
THE PATIENT WAS REVISED TO ADDRESS TRAUMATIC DISLOCATION AND STEM LOOSENING. LOOSENING OCCURRED AT THE CEMENT TO IMPLANT INTERFACE. CEMENT MANUFACTURER IS UNKNOWN. IT WAS ALSO INDICATED THAT THE STEM WAS RE-CEMENTED IN THE CANAL AND POLY WAS EXCHANGED. DOI: (B)(6) 2017; (B)(6) 2017; LEFT SHOULDER.
Patient problems: Joint Dislocation
MAUDE report 7135338 ↗InjuryOct 25, 2017Device Dislodged or Dislocated; Device Dislodged or Dislocated
DER STATES THAT THE PATIENT HAD A DEPUY DELTA EXTEND IMPLANTED. PREVIOUSLY DISLOCATED AND WAS REVISED. PATIENT DISLOCATED AGAIN AND WAS REFERRED TO CURRENT SURGEON. THE SURGEON THEN REVISED THE HUMERAL STEM, CUP AND GLENOSPHERE TO ACHIEVE STABLE CONSTRUCT.
Patient problems: Joint Dislocation
MAUDE report 6975737 ↗InjuryOct 11, 2017Device Dislodged or Dislocated; Device Dislodged or Dislocated
DER STATES THAT THE PATIENT DISLOCATED AND IT WAS ALSO REPORTED THAT THE PATIENT PRESENTED AGAIN WITH DISLOCATION. SURGEON SHARED DURING PROP THAT THE PATIENT HAS LITTLE TO NO SOFT TISSUE SUPPORT. THEY REMOVED THE BELOW COMPONENTS AND IMPLANTED A DELTA XTEND CTA.
Patient problems: Joint Dislocation
MAUDE report 6940595 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K062250, K081620, K120174, K192855. One submission often covers many models, so check the product and lot details in each record.
- Z-0834-2018 · Aug 31, 2017 · TerminatedDELTA XTEND" Reverse Shoulder System Modular Centered and Eccentric Epiphysis Implants intended for use in total shoulder or hemi-shoulder replacement procedures in patients with n
- Z-0833-2018 · Aug 31, 2017 · TerminatedDELTA XTEND" Reverse Shoulder System Modular Centered and Eccentric Epiphysis Implants are intended for use in total shoulder or hemi-shoulder replacement procedures in patients wi
- Z-0534-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 38/+9 STD, sterile, Depuy France; REF 130738209. The device is used as a shoulder implant.
- Z-0537-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+6 STD, sterile, Depuy France; REF 130742206. The device is used as a shoulder implant.
- Z-0538-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+9 STD, sterile, Depuy France; REF 130742209. The device is used as a shoulder implant.
- Z-0535-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 38/+6R RET, sterile, Depuy France; REF 130738106. The device is used as a shoulder implant.
- Z-0539-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42+6R RET, sterile, Depuy France; REF 130742106. The device is used as a shoulder implant.
- Z-0536-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humeral PE Cup, Dia 42/+3 STD, sterile, Depuy France; REF 130742203. The device is used as a shoulder implant.
- Z-0532-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard humeral PE Cup Dia 38/+3 STD, sterile, Depuy France; REF 130738203. The device is used as a shoulder implant.
- Z-0533-2009 · Jul 10, 2008 · TerminatedDepuy DELTA Xtend, Standard Humerus PE Cup, Dia 38/+6 STD, sterile, Depuy France; REF 130738206. The device is used as a shoulder implant.
Other PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED devices
- DJO SURGICAL 509-02-032Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED364 reports
- DELTA XTEND 130790036Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED171 reports
- DELTA XTEND 130770018Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED159 reports
- TRIGEN 71118200Smith & Nephew, Inc. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED123 reports
- DJO SURGICAL 509-02-432Encore Medical, L.P. · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED119 reports
- DELTA XTEND 130738206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED116 reports
- COMPREHENSIVE SHOULDER SYSTEM 113633Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED2 recalls4 reports
- DELTA XTEND 130742206Depuy (Ireland) · PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page